Philippe Garteiser1, Gwenael Page1, Aimé-Pacifique Manzi2, Valérie Vilgrain2, Ralph Sinkus3,4, and Bernard E Van Beers1,2
1Center for Research on Inflammation, Inserm UMR1149, Paris, France, 2Radiology department, Beaujon hospital, AP-HP, Clichy, France, 3King's College London, London, United Kingdom, 4LVTS, Inserm U1149, Paris, France
Synopsis
Although MR
elastography (MRE) is a preferred diagnostic tool for hepatic fibrosis, MRE-derived
mechanical properties are affected by factors such as portal pressure. Since
portal pressure is modulated across the breathing cycle, we sought to evaluate
the effect of breathing condition by acquiring MRE in end expiration and end
inspiration in 19 healthy volunteers. A gravitational transducer operating at
50Hz was used, which yielded a satisfactory repeatability. Storage modulus
values were significantly higher at end inspiration than at end expiration, and
the amplitude of the difference was significantly correlated to the end
inspiration storage modulus values.
Background
Although
magnetic resonance elastography (MRE) is widely accepted as useful in the
diagnosis of liver fibrosis, the mechanical properties of the liver are known
to be influenced by a variety of factors besides the collagen content of the
tissue. Portal hypertension and chronic hepatic congestion are known to increase the apparent stiffness[1-3].
Post-prandial stiffness has been reported to increase relative to normal state[4]. Finally, the effect of breathing
conditions at MRE is not extensively studied. However, the change in position
of the diaphragm and visceral organs that occurs between the two extreme
positions of end inspiration and end expiration, and more importantly the
corresponding effect on hepatic venous blood flow could potentially modify the
mechanical properties of the liver[5]. In this work, we propose to
evaluate the repeatability of a novel type of mechanical actuation in MRE, and
to apply this mode of actuation for studying the variation in mechanical
properties between inspiration and expiration breath hold mode in volunteers
with healthy livers.Methods
This
investigational clinical study was approved by the appropriate ethics
committee, and informed consent was obtained from each participant. Recruitment
was done on the basis of prescription of an MRI examination with absence of
known hepatic disease. Acquisition was performed on a 3.0T Philips Ingenia
system equipped with 80mT/m gradients. MR elastography data were acquired with
the “eXpresso” sequence, consisting of a fractionally encoded gradient echo
sequence[6]. Mechanical actuation was performed
at a frequency of 50Hz with a gravitational transducer [7]. Acquisitions were carried out at
an isotropic spatial resolution of 4mm. An echo time of 6.91ms (third in phase condition
between water and fat) was chosen to maximize signal and prevent water-fat shift
artifacts. Repetition time was 85ms, and flip angle was 20°. Eight slices were
acquired with 4 temporal steps along the mechanical cycle in a single breath
hold. Encoding was performed sequentially in three orthogonal directions,
followed by an acquisition without motion encoding (unbalanced four-points
encoding)[8], for a total scan duration of 4
breath holds. Storage modulus was calculated by inverting the Helmholtz wave
equation using the rotational of the unwrapped phase images. Patients were
instructed to hold their breath either at end inspiration or end expiration.
The acquisition in inspiration was repeated twice for a repeatability
assessment. Additionally in two patients the acquisition at expiration could
also be acquired twice, in which cases the datasets were also added in the
repeatability study.Results
Seventeen
patients were recruited. An average storage modulus (G’) of 1.80 ± 0.28kPa was obtained during the first repeatability scan, and the
average value was 1.86 ± 0.33 during
the second repeatability scan. No significant difference was found between the
two repeatability scans with the Mann-Whitney test. Bland-Altman analysis of
the repeatability study indicated a 95% limits of agreement interval of [-15% -9%],
with a negligible bias of -2.7 ± 6.3%. Mechanical properties were found to increase
from expiration to inspiration. Indeed, average expiration G’ values were 1.47 ±
0.27kPa, while in inspiration those reached 1.94 ± 0.41 kPa (Mann-Whitney
p=0.0002), amounting to a difference of 0.49 ± 0.35 kPa between expiration and inspiration
(significantly different from zero, p < 0.0001). The extent of the
difference between inspiration and expiration breath hold storage moduli was
found to be significantly correlated with the the storage modulus at inspiration
(Spearman r = 0.64, p = 0.0035).Conclusions
These
results indicate that the storage modulus at MRE can precisely be estimated in
the liver using a gravitational transducer. We observed a significant effect of
respiration on the storage modulus of the liver. Mechanical properties were
higher at end inspiration than at end expiration, and the increase in storage
modulus was higher in individuals with high storage modulus at inspiration.
This effect could potentially be caused by decreased liver congestion in
blocked expiration, similar to the observation during Valsalva maneuver[9]. The use of breath hold mode may be
useful in manipulating the mechanical properties via a modulation in
intrahepatic pressure.Acknowledgements
This project has received funding from the European Union’s
Horizon 2020 research and
Innovation programme under grant agreement No 668039
References
1. Sun, H.Y., et al., Usefulness of MR elastography for predicting esophageal varices in cirrhotic patients. J Magn Reson Imaging, 2014. 39(3): p. 559-66.
2. Gharib, A.M., et al., Magnetic Resonance Elastography Shear Wave Velocity Correlates with Liver Fibrosis and Hepatic Venous Pressure Gradient in Adults with Advanced Liver Disease. Biomed Res Int, 2017. 2017: p. 2067479.
3. Simonetto, D.A., et al., Chronic passive venous congestion drives hepatic fibrogenesis via sinusoidal thrombosis and mechanical forces. Hepatology, 2015. 61(2): p. 648-59.
4. Jajamovich, G.H., et al., Quantitative liver MRI combining phase contrast imaging, elastography, and DWI: assessment of reproducibility and postprandial effect at 3.0 T. PLoS One, 2014. 9(5): p. e97355.
5. Ipek-Ugay, S., et al., Physiologic Reduction of Hepatic Venous Blood Flow by the Valsalva Maneuver Decreases Liver Stiffness. J Ultrasound Med, 2017. 36(7): p. 1305-1311.
6. Garteiser, P., et al., Rapid acquisition of multifrequency, multislice and multidirectional MR elastography data with a fractionally encoded gradient echo sequence. NMR Biomed, 2013. 26(10): p. 1326-35.
7. Runge, J., et al. Abstract 1369: A novel MR Elastography transducer concept based on a rotational eccentric mass: the gravitational transducer. in Proceedings of the 2017 International Society of Magnetic Resonance in Medicine conference. 2017. Honolulu, USA.
8. Guenthner, C., et al., Analysis and improvement of motion encoding in magnetic resonance elastography. NMR Biomed, 2018. 31(5): p. e3908.
9. Bang, D.H., et al., Doppler ultrasonography measurement of hepatic hemodynamics during Valsalva maneuver: healthy volunteer study. Ultrasonography, 2015. 34(1): p. 32-8.