MR Elastography is beneficial in measuring the
stiffness values of tissues which can change in disease. Nevertheless, cardiac MR
Elastography is not only challenging from the acquisition side but also from
the reconstruction side because of the higher stiffness values expected in the
cardiac muscle compared to the other organs in which MR Elastography is commonly
applied. Here, we present data in healthy subjects and a patient acquired with
the gravitational transducer with a novel synchronising strategy in combination
with a single-shot SE-EPI-MRE sequence employing second-order motion
compensated motion encoding gradients with and without ZOnal Oblique Multislice
Imaging.
Hardware: The MRE gravitational transducer concept2 was used to produce shear waves where the transducer runs continuously to maintain the periodic steady-state of the waves. The previous need for external hardware to achieve the synchronization3 has been mostly eliminated by pulse sequence modifications to measure the exact ECG detection time point and delay the excitation accordingly to synchronise the sequence with the transducer. An Arduino UNO is used to send periodic triggers to the transducer to ensure its correct position.
MR Sequence: A cardiac-triggered single-shot SE-EPI sequence, similar to the study4 but with flow and acceleration compensated MEGs, was implemented as shown in Fig.1 with and without ZOOM imaging options. In addition to the delay in the sequence providing synchronization with the transducer, gradually increasing delays were used to achieve different mechanical wave phases.
MR Measurements: Measurements were performed on a 3T Achieva MR scanner (Philips Healthcare, The Netherlands) in breath-hold with imaging parameters: TR=RR; TE = 60ms/76ms with/without ZOOM for a 2x2x4mm3 voxel; SENSE factor = 2; ETL = 31/63 with/without ZOOM. 3 slices were acquired from 3 healthy volunteers (1 female) and a representative patient case at 80Hz MRE vibration frequency where one of the healthy volunteers was scanned for reproducibility on different days with both options of the sequence. Data were acquired with MEGs applied in slice selection, phase encoding, and readout directions including a reference scan without MEGs.
MRE Reconstruction: Measurements from cardiac MRE present challenges for traditional reconstruction algorithms due to the data quality, limited number of available pixels transmurally, and long wavelengths. To overcome these challenges, we introduce a new method which estimates speed by computing the time to max displacement and performing an inversion (see Figure 2, top). To validate the method, we compared in simulated viscoelastic cylinders with varying frequency and stiffnesses, demonstrating efficacy of the estimate for sufficiently high vibration frequency (see Figure 2, bottom).
1. Feinberg DA, Hoenninger JC, Crooks LE, Kaufman L, Watts JC, Arakawa M. Inner Volume MR Imaging: Technical Concepts and Their Application. Radiology. 1985; 156:743-747.
2. Runge J, Hoelzl S, Sudakova J, Dokumaci AS, Nelissen JL, Lee J, Stoker J, Nederveen AJ, Nordsletten D, Sinkus R. A Novel MR Elastography Transducer Concept Based on a Rotational Eccentric Mass: The Gravitational Transducer. In: Intl. Soc. Mag. Reson. Med. 25. Vol 1369. Honolulu; 2017.
3. Dokumaci AS, Schneider T, Hadjicharalambous M, Hoelzl S, Sudakova J, Hollands D, Sinkus R, Nordsletten DA. In Vivo Cardiac MR Elastography Using a Gravitational Transducer. In: Intl. Soc. Mag. Reson. Med. 26. Vol 5585. Paris; 2018.
4. Arani A, Arunachalam SP, Chang ICY, Baffour F, Rossmann PJ, Glaser KJ, Trzasko JD, McGee KP, Manduca A, Grogan M, Dispenzieri A, Ehman RL, Araoz PA. Cardiac MR Elastography for Quantitative Assessment of Elevated Myocardial Stiffness in Cardiac Amyloidosis. J Magn Reson Imaging. 2017;46(5):1361-1367.