Antonia Barghoorn1, Bruno Riemenschneider1, Fei Wang1, Juergen Hennig1, and Pierre LeVan1
1Dept. of Radiology, Medical Physics, University Medical Center Freiburg, Freiburg, Germany
Synopsis
Spin-echo fMRI can be highly advantageous compared to gradient-echo fMRI with
respect to magnetic field inhomogeneity artefacts, but is seldomly used for
BOLD imaging due to its inferior sensitivity. The presented spin-echo implementation
of a highly accelerated gradient-echo fMRI pulse sequence therefore aims to improve
the sensitivity of spin-echo fMRI while profiting from a reduction of
inhomogeneity-induced artefacts. Preliminary measurements using a temporal
resolution of 134 ms for whole brain imaging show less signal dropout when compared
to the gradient-echo counterpart, as well as an increase in the number of
significantly activated voxels compared to the unaccelerated readout.
Introduction
Spin-echo (SE) fMRI can be highly advantageous
compared to gradient-echo (GE) fMRI with respect to magnetic field
inhomogeneity artefacts such as signal dropouts and geometric distortions [1]. However, at clinical field strengths, the
majority of BOLD fMRI experiments are performed using T2* weighted
gradient-echo sequences due to their superior sensitivity compared
to SE fMRI [2]. In
recent years, ultra-fast GE acquisitions using sequences such as simultaneous
multi-slice (SMS) or MR-Encephalography (MREG) have been developed that offer
the possibility of whole brain imaging in a few hundred ms [3,4]. At such
temporal resolutions, benefits include direct filtering of physiological artefacts
and improved sensitivities in activation and network connectivity studies. This
study therefore aims at the design of a highly accelerated spin-echo-MREG pulse
sequence to improve the sensitivity of SE-fMRI while benefiting
from a reduction in magnetic field inhomogeneity artefacts.Methods
In order to enable SE-MREG with low flip angle
excitation RF pulses (α=25°) [5], two refocusing pulses were implemented into the
pulse sequence. This guarantees maximum SNR since a large
percentage of the longitudinal magnetization remains along the B0-direction
for every excitation. Sample
measurements of two healthy volunteers were performed at 3.0 T (Siemens Prisma,
Erlangen, Germany) using a 64 RX head coil array. Additionally,
GE-MREG, GE-EPI and SE-EPI measurements were performed for comparison of BOLD
sensitivity and magnetic field inhomogeneity artefacts. Measurement
parameters for SE-/GE-MREG were TR=134ms/TR=100ms, TE=78ms/TE=37ms, matrix
size=192×192x150, and spatial resolution=(3×3×3)mm3. A single-shot
undersampled non-Cartesian 3D stack of spirals trajectory was used for readout,
and data was reconstructed with MATLAB(Natick, Massachusetts, USA) using a
conjugate gradient method with Tikhonov regularization [6]. For SE-EPI and GE-EPI,
parameters were TR=3980ms/TR=2620ms, TE=75ms/TE=30ms, matrix size
192x192x120/192x192x126, and spatial resolution (3x3x3)mm³. Visual stimulation
was performed for one volunteer with alternating 10s flickering checkerboard/20s
rest, which was repeated five times. A similar paradigm was used for the other
volunteer, who was additionally asked to perform finger tapping at the time of
visual stimulation. Following motion
correction and smoothing by a 6 mm Gaussian kernel with FSL [7], activation maps were computed using FMRISTAT [8] modelling noise as a 10th-order
autoregressive process.Results
In both
fMRI experiments the BOLD sensitivity of SE-MREG was significantly higher than that of
SE-EPI, with a 358% and 244% increase in
activated voxels above a statistical t-value threshold of t=5 (p<0.05, corrected), respectively
(Figure 1). When only considering strong
activations at a threshold of t=7, an increase of 4750%/710% was observed.
Compared to GE-MREG and GE-EPI, it is more robust to signal dropout in the area
of the sinuses (Fig.2 and Fig.3). Blurring
artefacts remain due to the long readout of 76ms which enables trajectory
distortions before and after the spin echo. The activation maps furthermore
revealed a higher sensitivity of GE-MREG compared to GE-EPI in both
experiments, with a 43%/14% increase in activated voxels above a threshold of
t=7. Although the number of activated voxels of SE-MREG is lower than for both
gradient echo pulse sequences, the peak activation value for SE-MREG was higher
than that of GE-EPI for the measurement with visual activation only (t=19.31
compared to t=13.24).Discussion/Conclusion
SE-MREG
demonstrated a superior sensitivity to SE-EPI in both sample fMRI experiments
performed at 3.0 T while additionally benefiting from a high temporal
resolution of TR=134ms, suggesting that accelerated acquisitions could constitute
a practical approach for SE-fMRI at such magnetic field strengths. Although the overall
number of activated voxels was lower than that of both GE-based pulse sequences, likely due to a reduction of intravascular signal in SE fMRI, the local peak
activation value was even higher compared to GE-EPI in one of the two
measurements. Furthermore, susceptibility-induced signal dropout in the area
around the sinuses could be reduced. This offers the opportunity for ultra-fast
fMRI in areas strongly affected by magnetic field inhomogeneity artefacts, as
for example the inferior temporal lobes [1]. In the future, the present study will
focus on implementing an adapted off-resonance correction for SE-MREG. Additionally,
sequence parameters such as TR and acquisition time will be adjusted to balance
temporal resolution and SNR for optimal BOLD sensitivity.Acknowledgements
This work was supported by grant EXC-1086 „BrainLinks-BrainTools“ from the German
Research Foundation (DFG) and grant 13GW0230A from the Federal Ministry of
Education and Research of Germany (BMBF).References
1.Halai,
A. D., Welbourne, S. R., Embleton, K. & Parkes, L. M. A comparison of dual
gradient-echo and spin-echo fMRI of the inferior temporal lobe. Human Brain
Mapping 35, 4118–4128 (2014).
2. Norris, D. G. Spin-echo fMRI: The
poor relation? NeuroImage 62, 1109–1115 (2012).
3. Feinberg, D. A. & Setsompop, K.
Ultra-fast MRI of the human brain with simultaneous multi-slice imaging. Journal
of Magnetic Resonance 229, 90–100 (2013).
4. Zahneisen, B. et al.
Three-dimensional MR-encephalography: Fast volumetric brain imaging using
rosette trajectories. Magnetic Resonance in Medicine 65,
1260–1268 (2011).
5. Tkach, J. A. & Haacke, E. M. A
comparison of fast spin echo and gradient field echo sequences. Magnetic
Resonance Imaging 6, 373–389 (1988).
6. Assländer, J. et al. Single
shot whole brain imaging using spherical stack of spirals trajectories. NeuroImage
73, 59–70 (2013).
7. Jenkinson, M., Bannister, P., Brady,
M. & Smith, S. Improved Optimization for the Robust and Accurate Linear
Registration and Motion Correction of Brain Images. NeuroImage 17,
825–841 (2002).
8. Worsley, K. J. et al. A
General Statistical Analysis for fMRI Data. NeuroImage 15,
1–15 (2002).