Marylène DELCEY1,2,3,4, Pierre BOUR1,3,4, Valery OZENNE1,3,4, Wadie BENHASSEN2, and Bruno QUESSON1,3,4
1IHU LIRYC, PESSAC, France, 2Siemens Healthcare, Saint-Denis, France, 3Univ. Bordeaux, Centre de recherche Cardio-Thoracique de Bordeaux, Bordeaux, France, 4INSERM U1045, Bordeaux, France
Synopsis
In the context of radiofrequency ablation of
cardiac arrhythmia, a catheter is inserted into the heart for contact electrophysiology
recording and/or ablating the arrhythmogenic substrate. In this context, we propose to exploit
MR-compatible microcoils located near the catheter tip to estimate the local
motion of the heart. For this, a tracking module was interleaved with segments
of a radial acquisition in a FLASH sequence. k-space segments were sorted
retrospectively according to tracking readings before image reconstruction.
Using this approach, artifact free images of a pig in vivo could be
reconstructed.
Introduction
Conventional cardiac
MRI uses ECG and respiratory sensors (belt, navigator) to prospectively/retrospectively
compensate the cardiac and respiratory motion. In the context of radiofrequency
ablation of cardiac arrhythmia, a catheter is inserted into one or several
cardiac chambers for contact electrophysiology recording and/or ablating the
arrhythmogenic substrate [1,2]. These instruments may include position sensors
whose readings may be used to sort k-space data to improve image quality,
particularly in presence of arrhythmia where ECG may be distorted. Moreover,
respiratory compensation using navigator-based techniques may suffer from
different scaling factors between displacement of the diaphragm and actual
displacement of the heart, depending on the slice location (ex atrium vs apex)
[3]. This study presents a method that exploits MR-compatible microcoils
located near the catheter tip to retrospectively sort k-space data and retrieve
a correct image without using conventional physiological sensors or
echo-navigation. The method was evaluated in vivo in pig.Methods
After sedation of the animal and mechanical ventilation at 12
breaths/min (protocol approved by ethic committee), a 9Fr MR-compatible
steerable catheter (Vision-MR, Imricor, Burnsville, MN, USA) was inserted
through the femoral vein and navigated until it reached the right ventricle
(RV) wall close to the apex. Conventional surface ECG were used to record
cardiac rhythm of the animal (85bpm). To perform active catheter tracking
during MRI scans and determine the 3D coordinates of the micro coils, a pulse
sequence integrating a tracking module interleaved with a radial Gradient Echo
(FLASH) imaging sequence was implemented [4]. Radial acquisition was chosen for
its intrinsic robustness to motion. The active tracking sequence comprised
three non-selective projection acquisitions along the respective axis followed
by acquisition of N k-space projections (Figure 1a-b).
Imaging parameters were slice thickness=3mm, TE/TR=4.6 /24.96ms, FA
= 15°, FOV=169mm, matrix=112*112, N=3 segments per TR, radial views = 6000, TA
= 98s), interleaved with tracking data acquisitions (FOV = 450mm, 512 sampled
points, time of tracking = 24.04ms). All MRI scans were performed on a 1.5 T MR
scanner (MAGNETOM, Aera, Siemens Healthcare, Erlangen, Germany).
Following the acquisition, phase of the k-space of the tracking data
for each microcoils was analyzed and converted into relative displacement.
Directions where a dominant modulation due to cardiac and/or respiratory motion
were filtered using a low pass Gaussian filters (sigma = 9) to extract both respiratory
and cardiac motion components. Data were binned into 10 cardiac phases then
each cardiac phase was sorted into 2 respiratory motion phases. Images were then
reconstructed after Cartesian regriding and inverse Fourier transform of the
corresponding k-space data. A cine image was reconstructed by playing out the
10 cardiac phases for a fixed respiratory phase.Results
Figure
2a shows an example of raw signals derived from the most distal microcoil of
the catheter on X and Y axes. From these signals, cardiac and respiratory
motion were extracted through filtering Figure 2b. The resulting respiratory
rate was 12.8 breath/min and ECG 87 bpm, consistent with ventilator frequency
and ECG readings (around 85 bpm). Figure 3 displays the segmentation of signals
in Figure 2b in 2 respiratory and 10 cardiac phases. As a result, between 246
to 492 projections were combined together to create each image. Figure 4 displays each image at different cardiac phase reconstructed from the proposed
method for one respiratory state (Figure 5 is an animated gif of images of Figure
4). A good image quality is obtained with contraction-relaxation of the
myocardium.Discussions and conclusions
This
proposed retrospective gating method exploits intra-cardiac position sensors
integrated into the catheter. The time penalty introduced by the tracking
module appears acceptable (around 20 ms). Although the results presented here
show image quality with 3 segments per TR, the sequence implementation offers
flexibility between tracking update time and the number of segments acquired
between two consecutive tracking modules. Such a method relies on local
measurement of displacement of the catheter during cardiac contraction and
respiration. Thus it is expected to provide alternative means for retrospective
image reconstruction in presence of arrhythmia where conventional ECG and
navigator-based approaches may fail. The method has been implemented on a
radial gradient echo image acquisition for proof of concept but may be extended
to other acquisition sequences.Acknowledgements
References
[1] Grothoff, Matthias, et al.
"Magnetic resonance imaging guided transatrial electrophysiological
studies in swine using active catheter tracking–experience with 14 cases."
European radiology 27.5 (2017): 1954-1962
[2] Mukherjee, Rahul K., et al. "Epicardial
electroanatomical mapping, radiofrequency ablation, and lesion imaging in the
porcine left ventricle under real-time magnetic resonance imaging guidance—an
in vivo feasibility study." EP Europace (2017).
[3] Roujol, Sébastien, et al. "Characterization
of respiratory and cardiac motion from electro-anatomical mapping data for
improved fusion of MRI to left ventricular electrograms." PloS one
8.11 (2013): e78852.
[4] Homagk, Ann‐Kathrin, et al. "An expandable
catheter loop coil for intravascular MRI in larger blood vessels." Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine 63.2 (2010): 517-523.