Dominik Ludwig1,2, Frederik Bernd Laun3, Karel D. Klika4, Peter Bachert1, and Tristan Anselm Kuder1
1Department of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen (FAU), Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany, 4Molecular Structure Analysis, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
Diffusion pore
imaging can be used to retrieve the pore-space function of an arbitrary closed
pore. While a few publications involving pore imaging using the long–narrow
gradient approach exist [1–3], all of them involve the use of sophisticated
acquisition schemes like CPMG or involve an additional readout gradient. In
this study, we were able to show that it is indeed possible to acquire the pore
space function of capillaries with a diameter of 15 µm using a simple modification
of the long–narrow gradient scheme on a 14.1 T NMR spectrometer.
Introduction
Diffusion pore
imaging enables the direct measurement of the shape of arbitrary closed pores
which are filled with an NMR-detectable medium [1-3]. The first implementation
of diffusion pore imaging on an NMR spectrometer [1] made use of the long–narrow
approach [4] in order to retrieve the phase information necessary for
reconstruction of the pore space function. In order to achieve correct
measurements for 20 µm capillaries, a CPMG-train of 180° pulses was used in [1],
where the long gradient with low amplitude was split into many short pules.
Also, the short high-amplitude gradient at the end of the pulse train was spilt
into two parts separated by a 180° pulse. More recently, it was shown that it
is also possible to use a spin-echo approach with a readout gradient in order
to counteract gradient imperfections. The aim of this work was to show the
possibility of acquiring diffusion pore images of glass capillaries with a
diameter of 15 µm by a simple modification of the long–narrow spin-echo approach
using only one 180° pulse without any further adaptations.
Methods
Figure 1 shows the
acquisition sequence used for the experiments. A 180° refocusing pulse was
inserted to realize a spin echo sequence to counteract $$$T_2^*$$$ relaxation. The
gradient pulses were arranged antisymmetrically around the refocusing pulse;
both the long and the short gradient pulse were split into two parts. The
q-value is thus given by (see Fig. 1)$$q=\gamma\delta_LG_L=-\gamma\delta_SG_S.$$ All
measurements were carried out on a Bruker 14.1 T Avance 600 NMR spectrometer
using a 5-mm TBI-probe with xyz-axis gradients with nominal gradient amplitudes
of 0.5 T/m in both the x and y directions. The gradient amplitudes and
durations used for the measurements presented here are shown in Table 1. The
gradient amplitude of the narrow gradient was always set to $$$G_S=500\,$$$mT/m and the
amplitude of the long gradient was
adjusted accordingly. The signal was acquired for 11 q-values with 12 averages
and a repetition time of 10 s. To demonstrate the effect on the measured pore
space function, the duration of the short gradient pulses was varied. The gradient strength was validated using a
free diffusion experiment on a doped water sample employing the known diffusion
coefficient for a given temperature.
Trapezoidal
gradients were used and all ramp times $$$\epsilon$$$ were set to 0.3 ms.
Capillaries with
an inner diameter of 15 µm and an outer diameter of 360 µm were cut into 5 cm
long pieces and submerged in distilled water for several days. For faster degassing,
the submerged capillaries were boiled and placed in a vacuum chamber.
Capillaries were then dried on the outside and vertically stacked in a 5-mm NMR
tube (Figure 2). The water signal of the phantom was manually shimmed to a FWHM
of 6 Hz. All measurements were carried out at 298 K; therefore a free water diffusion
coefficient of D = 2.3 µm2/ms was assumed for the simulations. Simulations were conducted
using a matrix approach to solve the Bloch-Torrey equations [5,6].Results
The measured signals,
as functions of the q-values, and simulations are shown in Figure 3 A–C for
three durations of the short gradient pulse. In all cases, the measurements are
in good agreement with the simulations, with only slight deviations for some of
the low q-values. The reconstructed pore-space functions in Figure 4 A–C
represent the projection of the cylindrical shape onto the gradient direction. The
pore-space functions match well to the expected function of a cylinder with a
diameter of 15 µm for these gradient settings. However, due to the relatively
long duration ($$$\delta_S$$$)
of narrow gradients pulses, a shrinkage of the
reconstructed pore space function is induced due to the edge enhancement effect,
being most severe for the longest duration measurement #3 (Figure 4C).Discussion and Conclusion
The results
prove the feasibility of diffusion pore imaging using a spin-echo version of the long–narrow
approach with antisymmetric arrangement of the gradient pulses without further
modifications. This approach may reduce the influence of confounding effects
such as concomitant fields. The
measurements were limited by the available gradient strength of 0.5 T/m, which
is insufficient to fully resolve pores with a diameter of 15 µm resulting in
considerable edge enhancement.Acknowledgements
Financial
support by the DFG (Grant No. KU 3362/1-1 and LA 2804/6-1) is gratefully
acknowledged.References
[1]: Hertel, Stefan, Mark Hunter, and Petrik Galvosas. "Magnetic resonance pore imaging, a tool for porous media
research." PRE 87.3 (2013): 030802.
[2]: Hertel, Stefan Andreas, et al.
"Magnetic-resonance pore imaging of nonsymmetric microscopic pore
shapes." PRE 92.1 (2015): 012808.
[3]: Bertleff, Marco, et al. "1D and
2D diffusion pore imaging on a preclinical MR system using adaptive rephasing:
Feasibility and pulse sequence comparison." JMR 278 (2017): 39-50.
[4]: Laun, Frederik Bernd, et al. "Determination
of the defining boundary in nuclear magnetic resonance diffusion
experiments." PRL 107.4 (2011): 048102.
[5]: Laun, Frederik Bernd, et al. "NMR-based diffusion pore imaging." PRE 86.2
(2012): 021906.
[6]: Grebenkov, Denis S. "Laplacian eigenfunctions in NMR. I. A
numerical tool." Concepts in Magnetic Resonance Part A: An Educational Journal
32.4 (2008): 277-301.