Nicole Vike1, Xin Li2, Kelsey Hopkins2, Luis Solorio2, and Joseph Rispoli2,3
1Basic Medical Sciences, Purdue University, West Lafayette, IN, United States, 2Biomedical Engineering, Purdue University, West Lafayette, IN, United States, 3Electrical and Computer Engineering, Purdue University, West Lafayette, IN, United States
Synopsis
Medication effectiveness relies on patient adherence to a
given treatment regimen. Often, patients do not adhere to the temporal
guidelines set by their physicians and treatments therefore remain less
effective. In situ forming implants (ISFIs)
eliminate the need for patient adherence and release an effective dose of drug
overtime. However, no methods exist to noninvasively and temporally validate
drug release in vivo. We conducted in vivo experiments that validated the
use of DWI to monitor ISFI diffusivity overtime. This has enormous implications
in pharmaceutical research as this method can robustly quantify diffusivity in
ISFIs post-implantation to ensure effective drug release.
Introduction
Despite
their popularity, oral dosage forms can lead to patient adherence problems.1,2 Oral medications often require repeated administration
over time to maintain a therapeutic window. Because of frequent dosing
requirements, many patients fail to maintain their dosing regimen, thereby
decreasing treatment efficacy. This can lead to undesirable side effects and/or
disease persistence.1,2 In situ
forming implants (ISFIs) are an alternative drug delivery system which can
eliminate the patient adherence problem.3 When administered using standard needle injection, ISFIs
form a small drug-containing implant in tissue. Over time, drug diffuses across
the shell of the implant via exchange with water in surrounding tissue.4 ISFI synthesis can be adjusted so medications are
released at varying rates. However, no noninvasive and quantitative methods exist to image ISFIs
rigorously and continuously over time, in vivo. By using DWI, we can
characterize the diffusivity in ISFIs in
vivo over time. Results from these experiments can be used to validate
dosing effectiveness. This is critical since tissue environment can drastically
affect dosing effectiveness.5Methods
Animal studies were performed following protocols approved by the
Purdue Animal Care and Use Committee. Four four-week old male C57Bl/6 WT mice
were used. 52 kDa poly(lactic-co-glycolic) acid (PLGA),
N-methyl-2-pyrrolidone (NMP), and fluorescein were combined in a 39:60:1 ratio.
Mice were anesthetized using 1.5% isoflurane with an oxygen flow rate of 2
L/min. 100 L of
solution was injected subcutaneously over the right and left flank using a 23-gauge
needle. For imaging, mice were anesthetized with 2.5% isoflurane and an oxygen
flow rate of 250 mL/min. Standard DWI imaging was performed (TE=17.5 ms,
TR=2500 ms, FOV=30x30 mm2, slice thickness=0.80 mm, b=0,1000 s/mm2) at set time points (1h, 6h, 24h, 72h, 120h post-injection) using a Bruker
BioSpec 70/30 USR 7T Preclinical MRI system and a RF RES 300 1H
075/040 QSN TR volume coil. Trigger was used to gate respirations. Using a
custom Matlab code, two users manually selected ROIs from each slice containing
the implant to calculate MD; this was performed in triplicate. MD was averaged across
slices and between users; the average value was used for further analyses. After
imaging at 120h, mice were euthanized and implants were dissected out for
scanning electron microscopy (SEM). Implants were freeze-fractured over dry ice
and lyophilized for 4-5 days. They were then mounted on aluminum stubs and
sputtercoated with palladium. SEM was completed using a NovaNanoSEM and Quanta
3D FEG SEM. ANOVA test with Tukey multiple comparisons
was used to test for statistical significance between groups. Values are
reported as mean±standard
deviation. Results
Previous phantom results
were presented at ISMRM 2018.6 These in vitro
data are shown in Figure 1 which displays ADC maps for 52 kDa implants over the
full time course. After 14d, the implant was fragmented because of degradation.
Figure 2 shows representative ADC maps from one mouse over the course of five
days. For these images, the slice containing the largest implant area was
selected. Figure 3 compares MD between in
vitro and in vivo experiments. In
general, a similar trend in MD was observed for both in vitro and in vivo
analyses. However, at timepoints 6h and 3d, MD was significantly different
between in vitro and in vivo data. SEM images from in vitro and in vivo experiments are shown in Figure 4. In contrast with in vitro analyses, the in vivo implants were more misshapen due
to constraints of the subcutaneous space and it was more difficult to observe
the distinction between the core and shell of the implant. Discussion
DWI is a noninvasive and
sensitive technique shown to provide valuable insight into drug release
mechanisms of ISFIs. Specifically, MD is a useful metric for the quantification
implant diffusivity in vivo over
time. In general, MD showed similar trends between in vitro and in vivo
analyses. Differences can be attributed to the dynamic environment of the
tissue. Increased pressure from surrounding tissue in vivo can explain the significantly larger MD in vivo versus in vitro at 3d. Conclusion
ISFIs are a promising alternative drug
delivery depot as they eliminate the patient adherence problem. However,
methods to noninvasively and quantitatively assess ISFIs in vivo remain elusive. Because tissue environments likely alter
drug release profiles post-implantation, it is critical observe these changes. Based
on our results, DWI proves to be a sensitive tool to evaluate ISFI diffusivity in vivo.
DWI provides robust information regarding the
time course and magnitude of diffusion in ISFIs. These metrics can be used to
improve ISFI design and ultimately, treatment efficacy. Acknowledgements
We would like to thank Dr. Gregory Tamer for
continued maintenance and assistance with the 7T system. We would also like to thank Dr. Sarah Calve for the mice used in these experiments. References
1. Bose B, Jose
J. Patient Medication Adherence: Measures in Daily Practice. Oman Med J.
2011;26(3):155-159.
2. Brown M, Bussell J. Medication Adherence:
WHO Cares? Mayo Clin Proc. 2011;86(4):304-314.
3. Wertheimer A, Santella T, Finestone A,
Levy R. Clinical and Economic Advantages of Modern Dosage Forms: Improving
Medication Adherence.
https://www.npcnow.org/system/files/research/download/Clinical-and-Economic-Advantages-of-Modern-Dosage-Forms-Improving-Medication-Adherence.pdf.
Published 2006. Accessed Novemeber 6, 2018.
4. Solorio L, Olear AM, Hamilton JI, et al.
Noninvasive Characterization of the Effect of Varying PLGA Molecular Weight bBends
on In Situ Forming Implant Behavior Using Ultrasound Imaging. Theranostics.
2012;2(11):1064-1077.
5. Solorio L, Exner AA. Effect of the
Subcutaneous Environment on Phase-Sensitive in Situ-Forming Implant Drug
Release, Degradation, and Microstructure. J Pharm Sci.
2015;104(12):4322-4328.
6. Vike N, Li X, Hopkins K, Solorio L,
Rispoli J. A Pilot Study of Correlations Between Mean Diffusivity and Drug
Release Over Time For an Implantable Drug Delivery System. In: ISMRM. 1036;
2018.