Kosuke Morita1, Masami Yoneyama2, Takeshi Nakaura3, Seitaro Oda3, Masahiro Hatemura1, and Yasuyuki Yamashita3
1Radiology, Kumamoto University, Kumamoto-shi, Japan, 2Philips Japan, Tokyo, Japan, 3Diagnostic Radiology, Kumamoto University, Kumamoto-shi, Japan
Synopsis
We
attempted to obtain brain high-resolution pseudo-3D (2D multi-slice acquisition
with very thin slice thickness) diffusion-weighted echo planar imaging (DW-EPI)
using a hybrid compressed sensing and sensitivity encoding (Compressed SENSE) framework
(EPICS). pseudo-3D-DWI with EPICS achieved
high-resolution (1.15 mm3) isotropic DWI within clinically feasible
scan time. Furthermore, EPICS clearly improved the accuracy and robustness of ADC values in high
b-value brain DWI with pseudo-3D acquisition without any penalty for scan
parameters.
PURPOSE
Typical
clinical brain diffusion-weighted imaging based on Echo Planar Imaging (DW-EPI)
images have limited spatial resolution compared to other imaging due to its high
sensitivity to B0 inhomogeneities. DWI with smaller voxel size
causes further image distortion. DW-EPI with sensitivity encoding (SENSE) helps
to reduce the voxel size without increasing of image distortion, but it often
suffers from increased noise-like artifacts on the center of the images due to
the high geometry factor [1, 2].
Recently, compressed
sensing (CS) or Compressed SENSE (C-SENSE) has emerged and it demonstrated that
the images from highly undersampled measurements can be reconstructed
accurately by using the sparsity of the MR images. We hypothesize that the CS
reconstruction can similarly improve image quality drastically for EPI based
DWI without further optimization of EPI sampling scheme. It has been shown
before that wavelet based denoising is an effective tool for image quality
improvement in high b-value DWI images [3, 4], in this work it is integrated
with SENSE parallel imaging in an iterative implementation.
In
this our study, we attempt to obtain brain high-resolution pseudo-3D (p3D, it
means 2D multi-slice acquisition with very thin slice thickness) DW-EPI using
CS framework (EPICS) and to demonstrate the feasibility of EPICS p3D-DWI.
MATERIALS AND METHODS
Experimental data was collected from 5 healthy
volunteers. Written informed consent was obtained from each volunteer and the
protocol was approved by the ethics committee. All studies were performed with
clinical 3.0T MR scanner (Philips, Ingenia 3.0T CX) and 32-channel dS-head
coil. EPICS is based single-shot DWI-EPI. We did not modify its sampling
pattern of single-shot EPI and we applied it into the C-SENSE framework. Scan
parameters of DWI-EPI were as follows: TR / TE = 15174 / 90 ms, field-of-view =
25.6 × 25.6 cm2, acquisition matrix = 224 × 224 (reconstruction
matrix = 512 × 512), slices thickness = 1.15 mm, voxel size = 1.14 × 1.14 × 1.15 mm, number of slices = 80,
NSA = 6, EPI factor = 55, Band width = 2230.4 Hz, CS factor or SENSE factor =
4.0, b-values = 0 and 1000 sec/mm2, Acquisition time = 6:19-6:25, Transverse
plane acquisition. EPICS p3D DWI were compared to conventional SENSE images for
image quality, especially for the reduction of image noise. ROIs were placed on
left and right white matter (WM). The average signal intensities (SI) of WM on
b0 and b1000 image and ADCs of WM were used for comparison between EPICS and
SENSE. SIs and ADCs of WM were assessed by using paired t-test. RESULTS
Figure
1 shows results of SIs and ADC value between EPICS and SENSE. There was no
significant difference in SI (b0 image). SI (b1000 image) of EPICS and SENSE
were 138.03±14.50 and 168.04±30.32 (p < 0.05). ADC
values of EPICS (0.72±0.04) was significantly
higher than that of SENSE (0.48±0.13) (p < 0.05). The
cause of lower ADC values of SENSE was most likely due to the presence of
severe noise over the brain. On the other hand, there were no significant differences
between ADC values of EPICS and those of reference. Note
that the reference ADC values are obtained by conventional clinical DW-EPI
sequence we are routinely using in our institute. It
indicated that EPICS can provide more accurate ADC values. Figure 2 shows
representative b1000 images and ADC maps using EPICS and SENSE in a volunteer.
EPICS clearly reduced the noise which exists in the center of the SENSE images.
It indicated that EPICS can provide more accurate ADC values with high
reproducibility and robustness. Figure 3 shows representative 3D MPR images
obtained by p3D-DWI. The p3D-DWI with EPICS achieved high-resolution (1.15 mm3)
isotropic DWI within clinically feasible scan time. CONCLUSION
EPICS
clearly reduces noise-like artifacts and significantly improves the accuracy
and robustness of ADC values in high b-value brain DWI with pseudo-3D acquisition
compared with conventional SENSE DW-EPI, without any penalty for scan parameters.
This technique may be helpful to further assess the many brain diseases.Acknowledgements
No
acknowledgement found.References
[1] Patricia
N, et al. Parallel Imaging Artifacts in Body Magnetic Resonance Imaging. Can
Assoc Radiol J. 2009;60: 91–98.
[2] Yanasak
NE, et al. MR imaging artifacts and parallel imaging techniques with
calibration scanning: a new twist on old problems. Radiographics.
2014;34:532-48.
[3] Wirestam R, et al. Denoising of Complex MRI Data by Wavelet-Domain
Filtering: Application to High-b-Value Diffusion-Weighted Imaging. Magn Reson
Med. 2006;56:1114-20.
[4] Yang
X, et al. A wavelet multiscale denoising algorithm for magnetic resonance (MR)
images. Meas Sci Technol. 2011;22:025803.