Xuesheng Li1, Haibo Qu1, Fenglin Jia1, Gang Ning1, Juncheng Zhu1, Zhijun Ye1, Shaoyu Wang2, Huapeng Zhang2, and Zhitao Zhang2
1West China Second University Hospital,University, Sichuan, China; Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Sichuan, China, Chengdu, China, 2Siemens Healthineers, China, Shanghai, China
Synopsis
Fast
T2-weighted imaging is a commonly used diagnostic sequence for fetal MR.
Because of the limitations of fetal development, the imaging effect of T1
nervous system imaging is difficult to achieve like the post-natal image
effect. The 3D StarVIBE sequence is used as a new sequence under 3T MR, using a radial K-space filling,
not sensitive to motion, can be applied to the fetal nervous system
examination. Recent reports suggest that the TurboFLASH(TFL) sequence with an
optimized TI time of 2500 ms is used as an advantage of excellent contrast and
overcoming motion artifacts in 2D gradient echo sequences. Therefore, this
study will use the 3D StarVIBE sequence and the TFL sequence as a reference to
judge its imaging effect.
Purpose
T1
weighted imaging as a basic sequence is critical in diagnosing fetal brain diseases,
while currently its research application at 3T is limited. This study aims to
improve tissue contrast by applying 3D StarVIBE to fetal neurological imaging
studies at 3T scanner.
Material and methods
In
this study, 32 cases of mid-late pregnancy (25 weeks of pregnancy - 38+1 weeks
of pregnancy) were enrolled and underwent fetal-specific cranial examination
using a 3T MR Scanner (MAGNETOM Skyra, Siemens healthcare, Erlangen, Germany).
All fetal brain T1WI sequences were applied with 3D StarVIBE and TFL sequences.
The basic scanning parameters of 3D StarVIBE were: TR/TE=3.86/1.65, FOV=350mm2,
and total acquisition time=1:05min; the basic scanning parameters of TurboFLASH
(TFL)sequence were: TR/TE=2900/2.54, FOV=380mm2, TI=2500ms, and total
acquisition time=1:03min. This study covered following evaluations : 1) Multiple
flip angles were set for 3D StarVIBE , namely 5.5 degrees, 9 degrees, 12
degrees, and 15 degrees, to determine the optimized flip angle with best
contrast ; 2) StarVIBE sequence was compared with TFL sequence for their
diagnosis performance; 3) The StarVIBE images with optimized flip angle were analysis
in different gestational week, and confirm whether the gestational week impact the
image quality; 4) The image were pointed scores in imaging quality and image
artifact, including contrast and display of important anatomical structures
such as basal ganglia, central groove, brain myelination, etc. The scores of images
quality divided into five grades from 0 to 4. The scores of artifacts are
divided into 4 grades from 0 to 3 scores. Results
By
optimizing the T1WI 3D StarVIBE imaging of 3T scanner, the imaging time is
comparable to that of TFL sequence. The results of image analysis are as
follows: 1) flip angle significantly influenced the contrast and image quality
of the T1 weighted image. A better image was obtained with a flip angle of 9
degrees: the contrast of gray matter was better than the flip angle of 15
degrees (P=0.0326) and the TFL
sequence (P=0.0002); the contrast of
myelination was better than the TFL sequence (P<0.0001). The showing of central sulcus was better than the
flip angle of 15 degrees (P=0.032)
and the TFL sequence (P=0.01); the showing
of basal ganglia structure was better than the flip angle of 15 degrees (P=0.021) and the TFL sequence (P=0.0012); It was also found that
the showing of myelination was better than the TFL sequence at all small angle
flip angles of the StarVIBE e sequence (P<0.05);
2)The showing of anatomical
structure and the image quality of the image were not gradually improved with
the gestational age growth. 3)The imaging noise and
artifacts of the 3D StarVIBE sequence at different flip angles were compared
with the T1 weighted images acquired by the TFL sequence. It was found that the
3D StarVIBE sequence was better than the TFL sequence in artifact generation at
9 degree, 12 degree and 15 degree flip angles (P < 0.05); the difference in image
between the StarVIBE sequence and the TFL sequence was not statistically
significant in terms of noise generation.Discussion
The
3D StarVIBE T1 weighted imaging
sequence g using 3TMR can obtain excellent T1-weighted images in a free
breathing state, overcoming fetal motion and maternal breathing interference,
and small angle excitation can produce better image quality. The optimal flip
angle is the analysis and evaluation of the image quality after imaging by
multiple angle settings, and finally statistical analysis is performed on the
observation imaging capability level of multiple brain tissues, and the ability
to overcome fetal and maternal artifacts is evaluated. The small flip angle 3D
StarVIBE 's ability to overcome motion artifacts has also been confirmed, so it
can provide a stable and effective T1-weighted image for 3T fetal nervous
system examination, with a small flip angle of 9 degrees imaging as a
preference, but its diagnostic performance is not obvious increase with the
gestational age of the fetus.Acknowledgements
No acknowledgement found.References
1. Tee
LM, Kan EY, Cheung JC, Leung WC. Magnetic resonance imaging of
the fetal brain..Hong Kong Med J. 2016 Jun;22(3):270-8.
2.
Malamateniou C, McGuinness AK, Allsop JM, O'Regan
DP, Rutherford MA, Hajnal JV. Snapshot inversion recovery: an
optimized single-shot T1-weighted inversion-recovery sequence for improved
fetal brain anatomic delineation. Radiology. 2011 Jan;258(1):229-35.
3. Ferrazzi
G, Price AN, Teixeira RPAG, Cordero-Grande L, Hutter
J, Gomes A, Padormo F, Hughes E, Schneider
T, Rutherford M, Kuklisova Murgasova M, Hajnal JV. An efficient
sequence for fetal brain imaging at 3T with enhanced T1 contrast and
motion robustness.Magn Reson Med. 2018 Jul;80(1):137-146.
4.
Rousseau F1, Kim K, Studholme C, Koob M, Dietemann JL. On
super-resolution for fetal brain MRI.Med Image Comput Comput Assist
Interv. 2010;13(Pt 2):355-62.