Lorenz Pfleger1,2, Lukas Hingerl2, Albrecht Ingo Schmid2,3, Philipp Moser2, Wolfgang Bogner2,4, Yvonne Winhofer1, Siegfried Trattnig2,4, and Martin Krššák1,2,4
1Division of Endocrinology and Metabolism, Department of Medicine III, Medical University of Vienna, Vienna, Austria, 2High-field MR Centre, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria, 3Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria, 4Christian Doppler Laboratory for Clinical Molecular MR Imaging, Vienna, Austria
Synopsis
This
study focuses on 31P 3D MR spectroscopic imaging (MRSI) of
the gallbladder and the reduction of voxel bleeding by Non-Cartesian
encoded data sampling. Our results demonstrate on a phantom that the
contamination due to point-spread-function (PSF) can be decreased
compared to conventional Cartesian phase encoding. Qualitative
improvements were investigated by metabolic mapping of biliary
phosphatidylcholine (PtdC) originating from the gallbladder.
Purpose
31P
MR spectroscopic imaging (MRSI) offers a possibility to detect
phosphatidylcholine (PtdC), the major component of the human bile1.
Concentration changes may point towards altered metabolism2
or disturbed PtdC secretion into the bile which indicates possible
hepatobiliary pathologies3. MRSI at ultra-high fields
leads to an enhanced spectral resolution and improved SNR per time
unit4. For a proper quantification of PtdC it is important
that the apparent voxel size of the analyzed volume does not spread
over the border of the gallbladder to adjacent hepatic tissue and
vice versa due to partial volume effects. Thus, narrowing the
point-spread-function (PSF) is advisable. This study focuses on the
PSF analysis and comparison between different k-space sampling
methods such as conventional Cartesian phase encoding (CPE), encoding
using spiral trajectories (SPT)5 and encoding using
concentric ring trajectories (CRT)6 and their feasibility
for in vivo abdominal application. Due to faster sampling using SPT
or CRT sensitivity to motion artifacts could be reduced.Methods
All
measurements were performed on a 7T MR System (Siemens Healthineers,
Erlangen, Germany) using double-tuned 1H/31P
surface coils (31P-loops: 10.5cm and 14cm, Rapid
Biomedical, Rimpar, Germany). The smaller one was used for a phantom
measurement while in vivo the 14cm coil was used due to deeper B1
tissue penetration to localize the gallbladder properly.
The
point-spread-function (PSF) was determined by the following
measurement: to ensure proper shimming and coil loading a large
phantom (20mmol/L 31P, 5L) was placed on top of the coil
and a small cylindrical vial (1mL, triphenylphosphate (TPP) diluted
in chloroform) serving as the point source was positioned slightly
below the coil (Fig.1A). Three different FID-based 3D-MRSI sequences (with the
same settings for TE=0.9ms, TR=1.8s, FOV:
20x20x20cm³, 12x12x12 matrix zero-filled to 16x16x16 placed in the
isocenter, frequency centered on the TPP signal) were used for
acquisition: one with conventional Cartesian phase encoding (CPE,
600µs block pulse, 100V reference voltage, bandwidth 3500Hz), and
two Non-Cartesian encoding sequences using spiral trajectories (SPT,
600µs block pulse, 100V reference voltage, bandwidth 3500Hz) and
concentric ring trajectories (CRT, 600µs sinc pulse, 100V reference
voltage, bandwidth 3846Hz) in-plane and phase encoding in partition
direction. The k-space was sampled spherical with CPE and cylindrical
with both SPT and CRT. To keep the measurement time similar averaging
was applied for the Non-Cartesian sequences in partition direction.
The total measurement time was 16:20min for CPE, 18:10min for SPT (5
averages) and 17:25min for CRT (4 averages). The TPP signal at 0ppm
was fitted with a single peak with jMRUI using AMARES7,8
and its distribution was analyzed to characterize the PSF. The
in-plane diameter of the ‘real’ voxel size was estimated as
full-width-half-maximum (FWHM) of the interpolated signal
distribution along the x-direction.
An
in vivo measurement was performed using the same sequences with a
reference voltage of 300V. The volunteer (male, 25y, BMI: 21.7kg.m-2)
was measured after over-night fasting to ensure the gallbladder
filled with bile. MR-acquisition was performed with the volunteer in
a right lateral position and the gallbladder centered above the
RF-coil (Fig.1B). A slice through the gallbladder was analyzed by fitting
PtdC, the prominent signal of the bile (at ~2.2ppm).
Results
The
signal distribution of TPP in-plane of the three methods visualizing
the PSF is shown in Fig.2. The in-plane FWHM was calculated to be
31mm for CPE, and 26mm for both SPT and CRT and thus leading to a
size reduction (in-plane) of about 30% for the Non-Cartesian encoding
sequences. In vivo results are presented in Fig.3: The PtdC signal
distribution of the slices containing the voxel of most PtdC signal
is mapped on the corresponding anatomical localizer images. The area
with high PtdC signal is less extended when using SPT or CRT compared
to CPE.Discussion/Conclusion
The
PSF analysis shows a great in-plane reduction when using
Non-Cartesian encoding methods. This is beneficial for MRSI
measurements and quantification of small volume organs such as the
gallbladder due to reducing signal bleeding to adjacent voxels as
well as signal contamination spreading from surrounding tissue into
the organ of interest. The in vivo application of the three different
acquisition methods shows that the localization can be achieved more
accurate when encoding with CRT or SPT rather than with CPE. As a
next step a quantitative SNR analysis needs to be performed to
evaluate the characteristics of SPT and CRT in a more detail. This
was not part of the study for various reasons: The measurement time
could not be kept equal, the bandwidth is determined by other
parameters and cannot be chosen freely for Non-Cartesian encoding and
a block pulse for excitation was not yet available in our CRT
sequence.Acknowledgements
The
study was supported by the Austrian National Bank Project #16724.References
-
Chmelík M, Valkovič L, Wolf P, et al. Phosphatidylcholine contributes to in vivo (31)P MRS signal from the human liver. Eur Radiol. 2015
- Khan SA, Cox IJ, Thillainayagam AV, et al. Proton and phosphorus-31 nuclear magnetic resonance spectroscopy of human bile in hepatopancreaticobiliary cancer. Eur J Gastroenterol Hepatol. 2005
-
Valkovič L, Chmelík M, Krššák M. In-vivo 31P-MRS of skeletal muscle and
liver: A way for non-invasive assessment of their metabolism. Anal Biochem
2017
- Chmelík M, Považan M, Krššák M, et al. In vivo (31)P magnetic resonancespectroscopy of the human liver at 7 T: an initial experience. NMRBiomed 2014
- Valkovič L, Chmelík M, Meyerspeer, M, et al. Dynamic 31P–MRSI using spiral spectroscopic imaging can map
mitochondrial capacity in muscles of the human calf during
plantar flexion exercise at 7 T. NMR Biomed 2016
- Hingerl L, Bogner W, Moser P, et al. Proton and phosphorus-31 nuclear magnetic resonance spectroscopy of human bile in hepatopancreaticobiliary cancer. Magn Reson Med. 2018
- Naressi A, Couturier C, Devos JM, et al. Java-based graphical user interface
for the MRUI quantitation package. MAGMA 2001
- Stefan D, Cesare FD, Andrasescu A, et al. Quantitation of magnetic resonance
spectroscopy signals: the jMRUI software package. Meas Sci Technol
2009