Kai Dolde1,2,3, Christian Dávid3,4, Gernot Echner1,2, Ralf Floca2,5, Clemens Hentschke2,5, Nina Niebuhr1,2,3, Kai Ohmstedt1,2,6, Nami Saito7, Merkur Alimusaj8, Beate Flügel8, and Asja Pfaffenberger1,2
1Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Heidelberg Institute for Radiooncology (HIRO), National Center for Radiation Research in Oncology (NCRO), Heidelberg, Germany, 3Department of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 4X-Ray Imaging and Computed Tomography, German Cancer Research Center (DKFZ), Heidelberg, Germany, 5Medical Image Computing, German Cancer Research Center (DKFZ), Heidelberg, Germany, 6Department of Medicine, Heidelberg University, Heidelberg, Germany, 7Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany, 8Center for Orthopedic and Trauma Surgery, Heidelberg University Hospital, Heidelberg, Germany
Synopsis
Abdominal organ motion
provides challenges for radiotherapy treatments, leading to inhomogeneous dose
distributions with over- and underdosage regions in the target volume. Repeated
4D-MRI acquisitions, allow to analyze inter- and intrafractional spatial motion.
The aim of this study was to investigate the impact of abdominal corsets for
motion reduction purposes, based on repeated 4D-MRI data sets. We found pronounced
reductions in cranio-caudal and anterior-posterior direction using corsets,
which additionally lead to more reproducible motion amplitudes. Lower
amplitudes and better reproducibility are beneficial for radiotherapy and could
lead to smaller irradiation margins and dose reductions to healthy tissue.
Introduction
In radiotherapy (RT) treatments
of abdominal tumors, organ and tumor motion provide challenges to the treatment
delivery. This is especially the case for particle therapy (PT) treatments,
where due to the inverse dose profile and the limited range of protons or ions,
over- or undershooting effects at the distal edge of the tumor may occur1,2.
Moreover, in PT of the pancreas, interplay effects between the scanning pencil
beam and intrafractional abdominal organ motion occur3,4. The resulting
over- and underdosage depends on the underlying abdominal motion amplitudes.
Therefore, motion
mitigation techniques, like the usage of abdominal corsets, are a required
reduce abdominal motion during delivery. This study investigates the motion and
deformation reduction of the pancreas by means of abdominal corsets, based on 10
repeated 4D-MRI measurements of 5 volunteers. Moreover, the important questions
of the impact of corsets on the reproducibility of pancreatic motion is
investigated. 4D-MRI is a suitable imaging modality for this purpose, since it provides
high soft-tissue contrast in the abdomen without exposing the subjects to any
imaging dose.Methods
For 5 healthy
volunteers each, 10 4D-MRI imaging sessions were acquired each within 2-4 weeks
to simulate repeated imaging during fractionated RT. In each session, 4D-MR
images were taken with and without a personalized MR-compatible abdominal
corset, made of homogeneous polyethylene, which was fitted to the abdominal
region of the respective volunteer. The MR images were acquired by means of a
T1-weighted gradient echo MR sequence with radial sampling and golden angle
spacing (field of view = 400x400 mm2
voxel size 1.5x1.5x3 mm3, spokes per partition = 2100,
bandwidth = 610 Hz, TE = 1.5 ms, TR = 3.3 ms, α=12°). The measurements were performed under free breathing on a 1.5 T
MR scanner (MAGNETOM Aera, Siemens Healthineers,
Erlangen, Germany) with a measurement time of 8 minutes. The data were offline
reconstructed by means of an iterative motion-compensated reconstruction
algorithm, using a k-space-center-based self-gating signal5. For each 4DMRI data set, 20 overlapping
breathing phases were obtained. The pancreas was manually delineated on each
end-exhale image. Based on deformable image registration with a Demons
algorithm, the motion amplitudes of all voxels within the pancreas delineation
were extracted between the breathing phases, see figure 1.
The mean motion amplitudes of the resulting
motion-volume-histograms (MVHs) in cranio-caudal (CC) and anterior-posterior
(AP) direction were used to quantify the motion reduction by the abdominal
corset. The standard deviations of the mean motion amplitudes among different
imaging sessions were used as a surrogate to quantify the reproducibility of
motion amplitudes. Similarly, pancreatic deformations were characterized by the
width (standard deviations) of the respective MVHs.Results
Highly variable pancreatic motion amplitudes were observed with mean
cranio-caudal (CC) motion amplitudes of up to 28.5/9.3 mm without/with
abdominal corset, respectively, resulting in an averaged CC motion reduction of
48% by the corset. Similarly, a mean anterior-posterior (AP) corset-based motion
reduction by 69% was observed among the subjects, as illustrated in figure 2.
No differences were found in left-right direction.
The corsets further enabled more reproducible motion amplitudes with
smaller day-to-day fluctuations, reduced by 62%/130% in CC/AP direction. With
respect to pancreatic deformations, for 4 out of 5 subjects, the corset
measurements revealed a more rigid pancreatic motion compared to the pronounced
observed deformations without corset.
Discussion
In particle therapy
treatments of the pancreas, abdominal organ motion is a major challenge. However,
repeated 4D-MRI measurements offer a viable opportunity to investigate intra-
and interfractional organ motion and deformation without applying any imaging
dose to the patients and provides high abdominal contrast. 4D-MRI can be
further used as a viable basis of 4D dose evaluation studies6. This
corset study indicates the advantage of using abdominal corsets when aiming at
low motion amplitudes and small day-to-day motion variations. The motion
reduction in CC/AP direction could contribute to avoid overshooting of the
target, which may lead to a reduced dose deposition in adjacent organs at risk.
Moreover, smaller safety margins for irradiation may be applicable by means of
smaller motion amplitudes and better reproducibility of the motion patterns.
Nevertheless, more subject data is necessary to conclude on the significance of
the corset benefits, and patient comfort and fitness need
to be investigated in further studies.Conclusion
4D-MRI was applied for
the investigations of intra- and interfractional motion for radiotherapy
purposes. This 4D-MRI-based study showed that the usage of corsets leads to
smaller pancreatic motion and deformation and ensures less day-to-day motion
variations.Acknowledgements
The authors would like to thank Joao Seco for fruitful discussions.
References
1Kumagai M et al. Impact of
intrafractional bowel gas movement on carbon ion beam dose distribution in
pancreatic radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2009; 15:1276–81.
2 Durante M et al. Charged particles in radiation oncology Nat. Rev.
Clin. Oncol. 2010; 37:37–43.
3 Dolde K et al. 4D dose calculation for pencil beam scanning proton
therapy of pancreatic cancer using repeated 4DMRI datasets. Phys. Med. Biol.
2018; 63:165005 (10pp)
4 Batista V et al. Significance of intra-fractional motion for
pancreatic patients treated with charged particles. Radiat. Oncol. 2018; 13:120.
5 Rank C et al. 4D respiratory motion-compensated
image reconstruction of free-breathing radial MR data with very high
undersampling. Magn. Res. Med. 2017; 77(3): 1170-1183.