Meng Gu1, Ralph E. Hurd1, and Daniel M. Spielman1
1Radiology, Stanford University, Stanford, CA, United States
Synopsis
The SNR advantage of short-TE methods over long-TE
editing-methods provided a compelling motivation for reliable quantification of
glutathione using LCModel. In this study, we improved the accuracy of the
glutathione basis by using an experimental glutathione in our otherwise
synthetic LCModel basis. We further added ascorbate as a complementary measure
of oxidative stress. Validation was made using a brain metabolite titration
phantom with a matrix of glutathione and ascorbate. The near perfect linear regression results
demonstrated that both GSH and ASC at physiological concentrations can be
reliably quantified using LCModel. In vivo study also showed consistent
glutathione quantification.
INTRODUCTION
In a recent comparison of
PRESS and MEGA-PRESS quantification of glutathione (GSH) in human brain at 3T,
Nezhad et al.1 demonstrated a clear need
for phantom validation when using basis-set fitting of metabolites such as GSH in
short TE PRESS. Although MEGA-PRESS was found to have a clear advantage in the
Nezhad study, the SNR advantage of short-TE methods over long-TE editing-methods
provided a compelling motivation for improving, testing, and validating a short-TE
basis for glutathione. Due to the complexity of the GSH resonances and their dependence
on temperature, simulated basis spectra are often inaccurate, resulting in
compromised in vivo quantification.2 In this study, this problem was remedied by using an experimental GSH (pH
7.2, 37C) spectrum in our otherwise synthetic LCModel basis. We further
explored adding experimental ascorbate (ASC) spectrum to our basis set for its
value as a complementary measure of oxidative stress. The impact of ASC was
also important as it partially overlaps with GSH signals.METHODS
Validation was made using a
37C brain metabolite titration phantom with varying GSH and ASC concentrations,
having each metabolite concentration ranging from 0 to 4.0 mM in 0.5 mM steps. Prior
to LCModel fitting3, phantom data were modified with in-vivo lineshapes (0.05 ppm) and SNR
(140:1 for NAA). Linear regressions of both GSH and ASC titrations were also made.
The hybrid LCModel basis set was also tested for repeatability on in vivo
spectra where data were acquired six times from a voxel in the occipital lobe
of a healthy subject in one exam. To avoid partial suppression of resonances
above 3.8 ppm, a GE PROBE-p™ sequence was
used with a reduced water-suppression
RF pulse bandwidth of 75Hz. All spectra were acquired on a GE 750 3T scanner with 2x2x2cm voxel, TE/TR=35ms/2s,
64 averages, 2:56 min total scan time. RESULTS
Spectra
and their LCModel fits for the GSH titration with ASC at 2
mM using the all simulated basis set versus hybrid basis set are shown in
Figure 1a and Figure 1b respectively. Those
for the ASC titration with GSH at 2
mM using the all simulated versus hybrid basis sets are shown in Figure 2a and
Figure 2b respectively. Figure 3 and Figure 4 shows the linear regressions of GSH
and ASC titrations, respectively. Figure 5 shows a representative in vivo
spectrum and its LCModel fit. Estimated from the in vivo repeatability study, GSH/(PCr+Cr)
in the occipital lobe was 0.199+/-0.027 with a mean CRLB of 14% while ASC/(PCr+Cr)
was 0.321+/-0.053 with a mean CRLB of 16%.DISCUSSION
Compared
to using the all simulated basis set, linear regression results showed
clear improvement in the quantification of GSH and ASC using the hybrid basis
set. The
near perfect linear regression results of
both GSH and ASC titrations using the hybrid basis set demonstrated that both
GSH and ASC at physiological concentrations can be reliably quantified using a
short-TE PRESS acquisition in combination with an improved LCModel basis set.
Assuming an 8 mM (PCr+Cr) concentration, the estimated GSH in OCC was 1.59+/-.22
mM, consistent with the study by Nezhad et al. CONCLUSION
With an accurate and
complete basis set, in vivo GSH can be reliably quantified from short TE PRESS
spectra without editing using LCModel. Compared to the editing method with long
TE, short TE measurement utilizes higher SNR and thus requires shorter
acquisition time.Acknowledgements
NIH EB015891, NIH MH110683,
the Lucas foundation, GE Health Care, and the Stanford Center for Cognitive and
Neurobiological ImagingReferences
1. Sanaei Nezhad, F., Anton, A., Parkes,
L. M., Deakin, B., and Williams, S. R. (2017) Quantification of glutathione in
the human brain by MR spectroscopy at 3 Tesla: Comparison of PRESS and
MEGA-PRESS. Magn Reson Med 78, 1257-1266
2. Kaiser,
L. G., Marjanska, M., Matson, G. B., Iltis, I., Bush, S. D., Soher, B. J.,
Mueller, S., and Young, K. (2010) (1)H MRS detection of glycine residue of
reduced glutathione in vivo. J Magn Reson
202, 259-266
3. Provencher,
S. W. (2001) Automatic quantitation of localized in vivo 1H spectra with
LCModel. NMR Biomed 14, 260-264