Joshua S. Greer1,2, Vasu Gooty1, Animesh Tandon1,2, Gerald F. Greil1,2,3, Tarique Hussain1,2,3, and Ananth J. Madhuranthakam2,3
1Pediatrics, UT Southwestern Medical Center, Dallas, TX, United States, 2Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Congenital
heart disease (CHD) is associated with anatomical abnormalities in the
pulmonary arteries (PAs) which can result in asymmetric flow to the lungs, as
well as an overall reduction in lung perfusion. Non-contrast pulmonary perfusion
imaging using arterial spin labeling has the potential to provide both
quantitative perfusion maps as well as a measure of asymmetric pulmonary blood flow. In
this study, multi-slice, free-breathing pulmonary perfusion imaging is
demonstrated using the FAIR technique for full coverage of the lungs, and flow
measurements are compared with the standard phase contrast measurement approach.
Introduction
Congenital
heart disease (CHD) is associated with anatomical abnormalities in the
pulmonary arteries (PAs) which can result in asymmetric flow to the lungs, as
well as an overall reduction in lung perfusion. Clinical evaluation of patients
with the potential for asymmetric pulmonary flow is currently performed using
phase contrast MRI. However, pulmonary perfusion imaging has the potential to provide
a measure of flow distribution, as well as regional maps of quantitative
perfusion. Arterial spin labeling (ASL) is a non-contrast perfusion MRI
technique that can be used to generate quantitative perfusion maps, making it
ideal for CHD patients needing longitudinal monitoring and for those with
contraindications to contrast administration. Although ASL has been widely applied
for pulmonary perfusion imaging in adults [1-3],
studies in pediatric populations are limited [4].
The purpose of this study was: 1) to demonstrate the feasibility of whole-lung
perfusion imaging using ASL in pediatric populations with CHD, 2) to compare
the resulting perfusion measurements with the established phase-contrast
technique, and 3) to assess the ability of optimized background suppression
(BGS) to reduce motion artifacts to enable robust quantitative perfusion
mapping during free-breathing.Methods
Nine patients with CHD (11 ± 7
years old; Figure 1), were scanned on a 1.5T Philips Ingenia with IRB approval and
informed assent. Sagittal multi-slice flow alternating inversion recovery (MS-FAIR)
[2] perfusion images were acquired
across both lungs in each patient. The sequence began with saturation of the
imaging region, followed by labeling with a pair of selective and non-selective
inversions using a hyperbolic secant pulse. A post-labeling delay of one
cardiac cycle (R-R interval) [1] was used to allow labeled blood to perfuse the lungs.
A SShTSE acquisition was used to minimize the susceptibility artifacts due to
B0 inhomogeneities in the lungs. The sequence was ECG-triggered to place the
acquisition during the diastolic phase of the following cardiac cycle. BGS was
applied using four non-selective inversion pulses during the inversion time [5]. FOCI inversion pulses were used
for improved suppression in regions of B0 and B1 inhomogeneities [3,
6]. Although FAIR perfusion images are regularly
acquired during guided breathing or an extended breathhold, these approaches
can be difficult for some pediatric patients. Because BGS has been shown to
significantly reduce respiratory motion artifacts near the diaphragm [3], a free breathing approach was
used in this study for patient comfort, and to evaluate the resulting image
quality in the presence of uncontrolled respiratory motion. 5-8 slices were
acquired to cover an entire lung with 3 signal averages over 1.5 minutes of
free breathing. The slice thickness was 15mm and resolution = 3x3 mm2.
A separate M0 image was acquired for quantification [2,
7]. To measure true tissue perfusion, pooled blood in
the major pulmonary vasculature was masked prior to measuring average perfusion
across all slices [2]. Images were reconstructed and
segmented offline in MATLAB. As part of the clinical protocol, phase contrast
measurements were made in the branch PAs to assess differential pulmonary blood
flow [8]. Both the flow ratio between
lungs, as well as flow rate in the PAs were correlated with FAIR measurements
in each lung.Results
Figure 2 shows representative
MS-FAIR perfusion images across both lungs in one patient. BGS enabled the
patients to breathe freely throughout all perfusion scanning without
introducing misregistration artifacts at the chest wall or diaphragm. Figure 3
(A) shows excellent agreement between quantified perfusion and the flow rate through
the PAs to each lung, and (B,C) show excellent agreement between the left/right
flow ratio between phase contrast and ASL. Figure 4 shows quantified perfusion
maps across the lungs in two patients, showing good agreement with the flow per
heartbeat in the right and left PAs. The average perfusion across all subjects
was 183±72 and 168±42 mL/100g/min in the right and left lungs. Figure 5 shows
one patient that also had pulmonary hypertension. The FAIR technique was able
to detect patches of perfusion deficits that are characteristic of pulmonary
hypertension. A contrast-enhanced angiography (Fig. 5D) scan also showed the
expected peripheral pulmonary vascular pruning that occurs in these patients,
leading to reduced pulmonary perfusion.Discussion
MS-FAIR with BGS enabled whole-lung
coverage during free-breathing in a clinically acceptable scan time. Quantified
perfusion measurements in the lungs agreed well with the established phase
contrast approach, although the average perfusion was markedly lower than the
expected perfusion in adults (400-600 mL/100g/min) [1, 2]. Future improvements to this technique will include reducing
bright signal in the pulmonary vasculature to enable evaluation of the
underlying perfusion using inflow saturation during the end of the inversion
time [3].Acknowledgements
No acknowledgement found.References
[1] D.
S. Bolar et al., "Quantification
of regional pulmonary blood flow using ASL-FAIRER," Magn Reson Med, vol. 55, no. 6, pp. 1308-17, Jun 2006.
[2] J. S. Greer et al., "Non-contrast quantitative pulmonary perfusion using
flow alternating inversion recovery at 3 T: A preliminary study," Magnetic resonance imaging, vol. 46, pp.
106-113, 2018.
[3] J. S. Greer, X. Wang, and A. J. Madhuranthakam,
"Robust non-contrast perfusion imaging of whole-lungs using multi-slice
FAIR at 3T," in International
Society for Magnetic Resonance in Medicine, Paris, France, 2018, p. 2178.
[4] M. C. Liszewski et al., "Magnetic resonance imaging of pediatric lung
parenchyma, airways, vasculature, ventilation, and perfusion: state of the
art," Radiologic Clinics, vol.
51, no. 4, pp. 555-582, 2013.
[5] N. Maleki, W. Dai, and D. C. Alsop,
"Optimization of background suppression for arterial spin labeling perfusion
imaging," MAGMA, vol. 25, no. 2,
pp. 127-33, Apr 2012.
[6] X. Wang et al., "Frequency Offset Corrected Inversion Pulse for B0
and B1 Insensitive Fat Suppression at 3T: Application to MR Neurography of
Brachial Plexus," Journal of
Magnetic Resonance Imaging, vol. 48, no. 4, pp. 1104-1111, 2018.
[7] R. B. Buxton et al., "A general kinetic model for quantitative perfusion
imaging with arterial spin labeling," Magnetic
resonance in medicine, vol. 40, no. 3, pp. 383-396, 1998.
[8] S. Sridharan, G. Derrick, J.
Deanfield, and A. Taylor, "Assessment of differential branch pulmonary
blood flow: a comparative study of phase contrast magnetic resonance imaging
and radionuclide lung perfusion imaging," Heart, vol. 92, no. 7, pp. 963-968, 2006.