Elizabeth Iffrig1, John Wilson2, W. Robert Taylor2,3, Xiaodong Zhong4, and John Oshinski2
1Department of Medicine, Emory University, Atlanta, GA, United States, 2Emory University, Atlanta, GA, United States, 3Atlanta VA Medical Center, Atlanta, GA, United States, 4Siemens, Atlanta, GA, United States
Synopsis
Herein, we describe a method for implementation of cine DENSE imaging in the abdominal aortic wall. The method uses in house software to post-process and track displacement of intermural pixels and applies a quadrilateral based solution to measure strain. This new method permits the measurement of circumferential heterogeneity of both displacement and strain.
Introduction
Vascular
kinematics are predictors of cardiovascular mortality (1) but the ability to measure relevant parameters
using MRI is insufficient for clinical utilization. Current MRI
techniques used to measure vascular displacement and
strain are unable to resolve circumferential heterogeneity. (2) The ability to quantify regional kinematics in
the aorta would constitute a significant contribution given that heterogeneity is critical in rupture risk stratification for abdominal
aortic aneurysms (AAA) for which rupture location depends on local,
patient-specific biomechanics (3). Displacement-encoding with stimulated echoes
(DENSE) is an MRI sequence capable of quantitatively measuring sub-pixel
displacement and strain in the vessel wall (4). The
objective of this study was to apply a time-resolved, 2D, spiral cine DENSE
imaging sequence to measure displacement of the infrarenal aorta wall to assess baseline heterogeneities in
radial and circumferential displacement, and to use this displacement data to
quantify circumferential strain. We hypothesized that both displacement and
circumferential strain would significantly differ around the circumference of
the vessel.Methods
Seventeen
subjects (n=9 women, n=8 men) with no cardiovascular disease were
recruited for the IRB-approved study. For
each subject, a prospectively ECG-triggered and free-breathing navigator-echo
gated, spiral cine DENSE sequence was used to acquire a slice perpendicular to
the aorta midway between the renal artery origin and the aortic bifurcation
(Figure 1A). The sequence used segmented, spiral k-space sampling with displacement encoding value (ω) of 0.25 cycles/mm ,TE of
1.25 ms, TR of 16 ms, 8 spiral interleaves per image, 1 spiral interleave per
heartbeat, slice thickness of 8 mm, and reconstructed pixel size of 1.8× 1.8mm
(Figure 1B). Sequential acquisitions were performed to encode
displacement in the two directions of the image plane (Figure 1C) and post
processing was performed using a custom DENSE analysis program written in
Matlab.. Using each of the phase
image sets, displacement fields were created for the vessel
wall (Figure 1D) and converted to polar coordinates. Green strain was calculated in the image
coordinate system using an interpolation method based on quadrilateral
elements(5) and converted to a polar
coordinate system. For each volunteer, peak displacement and stain in each pixel was
assigned to one of 8 normalized sectors (Figure 2A) to quantify and visualize regional heterogeneity while minimizing
noise. Comparisons between sectors mean displacement and strain were made with a linear mixed model. An additional phase
contrast cine image was acquired in the same location as the DENSE image and
distensibility was calculated as $$$R^2/r^2-1$$$, where R is the radius at peak systole and r is the radius at end diastole. Distensibility was
compared with homogenized DENSE-derived strain using Bland Altman
analysis.Results
Average
peak displacement was greater in the anterior vessel compared with
the posterior half (1.5 ± 0.7mm vs 0.9±0.6 mm, p<0.05, Figure 2B). Sectors I-III (left anterolateral and
anterior walls) had larger radial than circumferential displacement (p<0.01,
Figure 3A), but sectors IV-VIII (right anterolateral through left
posterolateral wall) demonstrated greater circumferential displacement (p<0.01).
Peak circumferential strain (Figure 2C) in the anterior wall was lower than in each
of the lateral walls (p<0.05 comparing anterior to left and p<0.05
comparing anterior to right, Figure 3B). The mean (i.e., homogenized) circumferential
strain was calculated for all volunteers to be 0.14±0.05. Correlation between
distensibility and homogenized strain was evaluated with a Pearson correlation
coefficient of 0.76 and Bland Altman analysis demonstrated an average bias of 0.005±.0050.Discussion
This work demonstrates that cine DENSE imaging can be used to
determine regional displacement and circumferential strain in the abdominal
aorta. Application of this technique shows that while displacement is greatest
in the anterior wall, circumferential strain is greatest in the lateral walls. These
findings suggest posterior fixation of the aorta (low displacement and low
strain) with displacement of the anterior wall in “bulk” (high
displacement but low strain). Thus, the circumferential stretching of the
aortic wall to is accomplished
by stretch of the lateral walls (moderate displacement, high strain). The value
of the proposed method’s ability to measure displacements with sub-pixel
resolution is important when considering the potential for clinical application.Conclusion
Spiral
cine DENSE imaging is able to resolve heterogeneities in aortic kinematics. We demonstrate that 2D cine
DENSE imaging in the vessel wall is feasible in healthy volunteers
and we report significant differences in the displacement and circumferential strain around the vessel circumference. The kinematics of the aorta are heterogeneous, even in young healthy adults which may have implications in understanding susceptibilities of distinct regions to
the development and progression of pathologies of the infrarenal aorta, such as
AAAs, in whichevolving biomechanical
properties of the extracellular matrix likely play a significant role.Acknowledgements
The work in this abstract was supported by a pre-doctoral fellowship from the American Heart Association anda grant from Siemens Healthineers.References
1. Cavalcante
J, Lima JA, Redheuil A, Al-Mallah MH. Aortic stiffness: current understanding
and future directions. Journal of American College of Cardiologists
2011;5(57):1511-1522.
2. Suever J, Oshinski JN, Rojaas-Campos E,
Huneycutt D, Cardarelli F, Stillman AE, Raggi P. Reproducibility of pulse wave
velocity measurements with phase contrast magnetic resonance and applanation
tonometry. Internation Journal of Cardiovascular Imaging 2011.
3. Martufi G, Christian Gasser T. Review:
the role of biomechanical modeling in the rupture risk assessment for abdominal
aortic aneurysms. Journal of Biomechanical Engineering 2013;135(2).
4. Aletras A, Ding S, Balaban RS, Wen H.
DENSE: Displacement Encoding with Stimulated Echoes in Cardiac Functional MRI.
Journal of Magnetic Resonance 1999;137:247-252.
5. JD H. Cardiovascular Solid Mechanics:
Cells, Tissues, and Organs. New York, NY: Springer: 2002.