4D flow MRI with six-directional flow encoding has enabled the assessment of turbulent flows, including mapping of incoherent flow variance. Using such, non-invasive estimation of turbulence-driven pressure drops can be computed. Here, we present an extension of the virtual-Work-Energy-Relative-Pressure method8 for the assessment of turbulence-driven pressure drops. Using the concept of virtual fields, the method accurately assesses pressure drops over a range of stenotic valve phantoms, being validated against catheter-based measurements. With virtual probing enabling the assessment of pressure drops through complex, narrow vasculatures, the incorporation of turbulence enhances the utility of the method, enabling for refined clinical hemodynamic analysis.
The proposed method originates from a work-energy formulation (vWERP) assessing pressure drops from a flow field $$$v$$$7 using an additional virtual field $$$w$$$ to probe any vascular segment within the imaged domain8. By doing so, we have shown that the influence of measurement inaccuracies, low flow magnitudes and flow bifurcations can be minimized8.
To extend the method to handle turbulent flow, flow inconsistencies are introduced in the original formulation6, where such is given by the variance in the flow assessed by ICOSA6 4D flow MRI. With such, and following our previously proposed derivation of virtual energy7, the pressure drop $$$\Delta p$$$ over any given vascular segment can be evaluated by:
$$\Delta p=-\frac{1}{Q_w} (\frac{\partial}{\partial t}K_w+A_w+V_w+T_w)$$
with $$$K_w$$$, $$$A_w$$$ and $$$V_w$$$ being the kinetic, advective, and viscous virtual energy components when using $$$w$$$ as the weighting velocity field, and $$$Q_w$$$ being the flow of $$$w$$$ over the domain inlet (for details see6,7). Additionally, $$$T_w$$$ is the turbulent energy component of $$$v$$$ with $$$w$$$ as the weighting velocity field, with the term originating from the assessed variance of $$$v$$$ by:
$$T_w=-\rho \int_{\Omega} Cov\left[v,v\right] \cdot \triangledown w \cdot dx$$
with $$$\rho$$$ being the fluid density, $$$\Omega$$$ the domain over which the flow is being assessed, and $$$Cov\left[v,v\right]$$$ the covariance matrix of the observed flow.
To evaluate method performance, an in-vitro setup from5 was used. ICOSA6 4D flow MRI (for a full methodolgical description, see5) was acquired for 7 different shaped stenotic valve phantoms at 3 different flow speeds. Irreversible pressure drops were assessed over all valves and compared against invasive catheter measurements. Estimates were also compared against alternative measures for turbulence-driven pressure drop assessment, evaluated in previous work5.
Figure 1 shows the imaged velocity field, estimated flow variance, and virtual field for one of the imaged phantom valves. With the virtual field solved as a Stokes flow boundary value problem, the field is primarily governed by the geometry of the phantom, with maximum velocities in and around the stenosis.
Figure 2 shows summarized results. As seen, all pressure drops were assessed at high accuracy, with only a slight overestimation of 0.6 mmHg (Figure 2b), with the overestimation seemingly increasing with increasing pressure drop magnitude (Figure 2a). The linear regression coefficient of 0.98 (Figure 2a) indicates a well-tuned assessment method, with estimated values coinciding well with the invasively measured pressure drops.
The proposed method using virtual fields also compared well when evaluated against other methods tested on the same phantom dataset (see5). In particular, the method seems to render results with lower variance, indicating increased precision with maintained accuracy compared to previously published methods.
1. Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, et al. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. J. Am. Soc. Echocardiogr. 2008:101-102
2. Vahanian A, Baumgartner H, Bax J, Butchart E, Dion R, et al. Guidelines on the management of valvular heart disease: the task force on the management of valvular heart disease of the european society of cardiology. Eur. Heart J. 2007:28(2):230-68
3. Dyverfeldt P, Hope MD, Tseng EE, Saloner D, Magnetic Resonance measurement of turbulent kinetic energy for the estimation of irreversible pressure loss in aortic stenosis. JACC: Cardiovasc Im. 2013:6(1):64-71
4. Ha H, Lantz J, Ziegler M, et al. Estimating the irreversible pressure drop across a stenosis by quantifying turbulence production using 4D flow MRI. Sci Rep. 2017:7:46618
5. Ha H, Escobar Kvitting J-P, Dyverfeldt , Ebbers T. Validation of pressure drop assessment using 4D flow MRI-based turbulence production in various shapes of aortic stenosis. Magn Res Med. 2018:1-14
6. Gülan U, Binter C, Kozerke S, Holzner M. Shear-scaling-based approach for irreversible enery loss estimation in stenotic aortic flow – an in vitro study. J Biomech. 2017:56:89-96
7. Donati F, Figueroa CA, Smith NP, Lamata P, Nordsletten DA. Non-invasive pressure difference estimation from PC-MRI using the work-energy equation. Med. Image Anal. 2015:26(1):159-72
8. Marlevi D, Ruijsink B, Balmus M, Dillon-Murphy D, Fovargue D et al. Non-invasive pressure estimations by virtual fields – cardiovascular pressure drops from 4D flow MRI. ISMRM 2018, Paris, France.