1975

Voxel-wise comparison of CFD and 4DMR results in the cerebral venous outflow tract of a pulsatile tinnitus patient
Evan I-Tai Kao1, Henrik Haraldsson1, Megan Ballweber1, Matthew R Amans1, and David Saloner1

1Radiology, UCSF, San Francisco, CA, United States

Synopsis

To investigate abnormal hemodynamics in pulsatile tinnitus patients in vivo, we use 4DMR. We also use computational fluid dynamics (CFD) to overcome 4DMR’s limited spatial and temporal resolution. To ensure the simulation actually reflects in vivo blood flow, we systematically adjust the simulation boundary conditions to match the CFD and 4DMR data. This requires downsampling CFD data to 4DMR resolution and a quantitative voxel-wise comparison. Our results suggest 4DMR underestimates velocities in vivo due to its resolution. This effect is confirmed by phantom 4DMR data taken at multiple resolutions.

Introduction

Abnormal patient hemodynamics are often implicated in disease pathology. In the case of pulsatile tinnitus (PT), which is a pulse-synchronous sound of intracranial origin, vessel remodeling can lead to geometries that result in noise-generating flows. 4DMR can be used to capture the full blood velocity field in vivo over the cardiac cycle, but only at a limited spatial and temporal resolution. Computational fluid dynamics (CFD) can be used for patient-specific simulations at much greater resolutions, but is only as accurate as the boundary conditions used, which may be measured from 2DMR or simply derived from literature. To ensure the simulation actually reflects in vivo blood flow, we investigated how to systematically adjust the simulation boundary conditions such that the results match corresponding 4DMR data.

Methods

Patient data acquisition and processing

A patient with PT suspected to originate from a transverse sinus diverticulum was imaged on a 3T Siemens scanner using 4DMR (at 1.1 x 1.1 x 1.4 mm3 resolution) and CE-MRA (at 0.5 x 0.5 x 0.5 mm3 resolution) to obtain the in vivo velocity field and geometry. The 4DMR data was processed using in-house Python tools. As a starting point, the blood flow was measured from 4DMR and used as a boundary inlet condition for the CFD simulation.

CFD downsampling and readjustment

To quantitatively compare CFD and 4DMR, the CFD results were first downsampled to MR resolution such that the velocity ($$$a$$$) measured at a voxel is a gaussian distance ($$$w$$$) and volume-weighted ($$$v$$$) average of the neighboring CFD cells found within the voxel’s sampling volume (Figure 1), to mimic MR physics [1][2]:

$$a_{\textrm{voxel}} = \sum_{i=\textrm{cell}}{w_i\cdot v_i\cdot a_i }$$

$$w_i = w(r_i) = \frac{1}{(2πσ^2)^(\frac{3}{2}}e^{-(\frac{r_i}{2σ})^2}$$

At each point, the difference in velocity magnitude was calculated and a local mean error was obtained:

$$\textrm{error} = \textrm{mean of}\frac{\textrm{difference}}{\textrm{MR velocity}} = \frac{ \overline{ v_{CFD,i} - v_{MR,i} } }{ v_{MR,i} } $$

To avoid noise and partial voluming, the error calculation excludes voxels near or at the vessel wall and voxels below a velocity threshold. This error was then used to scale the inlet boundary condition of the next simulation iteration (n+1) as follows (Figure 2):

$$v^{n+1}_{\textrm{inlet}} = \frac{v^{n}_{\textrm{inlet}}}{1 + \textrm{error}}$$

Phantom data acquisition

The downsampling process is akin to decreasing image resolution. To confirm its effects on the velocity field experimentally, a silicone phantom created from the patient geometry was imaged to obtain 4DMR results at multiple resolutions: 0.7mm (high), 1.2mm (standard), and 1.5mm (low) isotropic.

Results

The effect of adjusting the CFD results based on the voxelwise comparison between CFD and 4DMR data is provided in Figure 2. Qualitatively, the downsampled and adjusted CFD results are closer to the 4DMR results than the downsampled, unadjusted CFD results. Quantitatively, the mean local velocity error calculated for the unadjusted and adjusted simulations are 0.103 and -0.003, respectively.

Figure 3 shows streamlines demonstrating the effects of downsampling the CFD results to MR grids of increasing resolution. Downsampling appears to severely affect near-wall velocity values due to voxel contributions from empty space. Inner streamlines are less affected, but velocities in abnormal geometries (diverticulum, arachnoid granulation) are still underestimated. Increasing the resolution retains more of the velocity magnitude and original flow features.

Figure 4 compares the patient 4DMR data to both the original and downsampled CFD results after the CFD has been readjusted. Overall, the velocity field from the CFD simulation appears to be much higher in magnitude than from 4DMR. The velocities are closer after the CFD data is downsampled but many of the detailed flow features may be lost, particularly in thin vessels.

We experimentally confirmed the effects of downsampling by analyzing the 4DMR results in a flow phantom at different resolutions (Figure 5). The overall velocities and local peak velocity drop as the resolution decreases.

Discussion

The downsampling process results in a smoothing effect that underestimates any peak velocities particularly in small vessels and vessel abnormalities. This suggests that in general, 4DMR-derived velocity fields taken at conventional resolutions may underestimate the true in vivo velocity field. Further evidence for this effect is provided from the 4DMR phantom results, which reveal a loss in velocity magnitude as resolution decreases. This is particularly important as research groups increasingly use 4DMR to derive clinically-relevant hemodynamic parameters, such as wall shear stress.

4DMR and CFD can be complementary diagnostic tools: 4DMR is valuable for both validating and adjusting CFD results, while CFD can capture details that may be missed due to noise and low resolution.

Acknowledgements

No acknowledgement found.

References

[1] Belen Casas, Jonas Lantz, Petter Dyverfeldt, and Tino Ebbers. 4d flow MRI-Based pressure loss estimation in stenotic flows: Evaluation using numerical simulations: Pressure Loss Estimation in Stenotic Flows Using 4d Flow MRI. Magnetic Resonance in Medicine, pages 1808-21, May 2015.

[2] Haraldsson, H., Kefayati, S., Ahn, S., Dyverfeldt, P., Lantz, J., Karlsson, M., Laub, G., Ebbers, T., Saloner, D., (2018), Assessment of Reynolds stress components and turbulent pressure loss using 4D flow MRI with extended motion encoding, Magnetic Resonance in Medicine, 79(4), 1962-1971. https://doi.org/10.1002/mrm.26853

Figures

Figure 1. Downsampling CFD data down to the 4DMR image grid for one slice. The rightmost figure shows the effect of compensating for partially unfilled voxels at the computational wall.

Figure 2. Streamlines exhibiting the effect of rescaling the inlet boundary condition based on the velocity difference error. By quantitatively comparing the velocity field between 4DMR and downsampled CFD data, the inlet boundary condition is rescaled for a second simulation (Adjusted, downsampled) that better resembles 4DMR.

Figure 3. Velocity map at a slice and corresponding streamlines demonstrating effect of MR resolution on retention of flow features and velocity magnitude after downsampling. A) CFD. B) Original MR Resolution (1.1x1.1x1.4 mm3). C) 2.37x MR resolution. D) 8x MR resolution.

Figure 4. Comparison of the velocity field between results from CFD, 4DMR, and CFD downsampled to 4DMR resolution at selected slices. The slices are specified on the geometry by colored lines. (Red) Sigmoid sinus. (Orange) Diverticulum. (Green) An arachnoid granulation compressing the vein. (Blue) Second arachnoid granulation that more severely affects the flow, splitting it into two streams.

Figure 5. Comparison of 4DMR phantom data of decreasing resolution from left to right (displayed in top left of each column). The top row displays streamlines, while the bottom row displays the velocity map at a slice in the image. The location of the slice in the vessel geometry is represented by the white dotted line. The velocity at the voxel at the same approximate location in each acquisition (white outline) is 1.03, 0.93, and 0.84 ms-1, giving a sense of the loss in peak velocity.

Proc. Intl. Soc. Mag. Reson. Med. 27 (2019)
1975