Sylvana García-Rodríguez1, Philip A. Corrado2, Alejandro Roldán-Alzate1,3, and Christopher J. Francois1
1Department of Radiology, University of Wisconsin - Madison, Madison, WI, United States, 2Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 3Department of Mechanical Engineering, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
False
lumen hemodynamics is an important factor in aortic dissection progression. As
a methodology to further characterize false lumen velocities, two patient-specific
3D printed models underwent 4D Flow MRI, from which histograms of velocity
components were generated at several locations along the lumen. Two VENC
settings were used and the data was grouped in diastole and systole. Histograms
of normal and tangential components serve as descriptors of flow regimes and
offer the possibility to correlate with thrombus formation and clinical
progression. VENC is important especially for the assessment of tangential
components.
Purpose
To analyze
3D printed aortic dissection false lumen velocity component distribution per
cardiac phase, acquired with 4D Flow MRI using two VENCs.Introduction
Aortic
dissection (AD) results in a false lumen, leading to complications including
aneurysm growth, rupture, end-organ malperfusion and hypertension.1 Several
studies highlight the importance of false lumen status - patent, partially, or
completely thrombosed - as a disease progression factor;2 few
studies have analyzed underlying hemodynamics. AD 4D Flow MRI is complex due to
the wide velocity range and the propensity for chaotic flow regimes. Previous
studies3,4 have acquired data with a VENC common for aorta imaging,
but not necessarily for false lumen assessment, where low velocities often
predominate. In this study, we analyze velocity component distributions during
each cardiac phase within the false lumen of two 3D printed aortic dissection
models at two VENC settings.Methods
Following IRB-approved and HIPAA-compliant protocols,
magnetic resonance angiography (MRA) and computed tomography angiography (CTA) data
from two patients (55 year-old female, MRA, and 40 year-old male, CTA) with acute
descending thoracic AD were used to generate models for 3D printing.
Patient-Specific Models: In vitro anatomical models were segmented
from CTA images (Mimics, Materialise; Leuven, Belgium). Surfaces were corrected
and smoothed (3-matic, Materialise) (Figure 1A); the geometries were hollowed
and tubing connections were added at inlets (ascending aorta) and outlets (descending
aorta and main arteries). The two geometries were exported in STL format for selective
laser sintering 3D printing to scale.
In Vitro MRI: Each physical model (Figure 1B) was
connected to a pulsatile pump (BDC PD-1100, BDC Laboratories, Wheat Ridge, CO) that
circulated water at 4 L/min as the maximum input flow of a physiological flow
waveform (35% systole, 60 bpm). In vitro 4D Flow MRI5 was performed
on a 1.5T scanner (MR750, GE Healthcare, Waukesha, WI) with parameters:
320 x 320 x 320 mm FOV, 1.25 mm isotropic spatial
resolution, TR/TE = 6.72/2.82 ms, FA = 8, scan time 10 minutes. For
each model, two VENCs were used: 70 and 150 cm/s.
Velocity
Distribution: The
false lumen of each 3D printed model was segmented from magnitude images (Mimics).
The resulting mask defined the velocity analysis domain (EnSight, CEI Inc.; Apex, NC). Several perpendicular planes were placed along
the lumen, for which three-directional velocities were exported. Plane data was
imported into a custom-developed Matlab tool. Histograms of normal and
tangential velocity components for each cardiac phase were computed for each plane.
An average histogram of all the planes was calculated for each VENC. Three
individual planes were also analyzed.
Results
Velocity
streamlines in Figures 2-3 demonstrate low velocities and small eddies, as well
as average histograms for systole. In general, low VENC scans incorporate lower
velocities when compared to high VENC scans, more so on tangential components. The
average percent difference of mode, mean and median is 21.2%, 16.7% and 13.8%,
respectively, when comparing low to high VENC data. In terms of cardiac phase,
the average percent difference of mode, mean and median when comparing systole
versus diastole, was 15.4%, 19.6% and 18.6%, respectively. Figures 4-5 show
velocity distributions at sample planes.Discussion
Different
false lumen velocity aspects can be inferred from resulting histograms (Figures
4-5). Model 1: Positive values (cephalocaudal
direction) on the normal component average histograms, might indicate primarily
forward flow. Plane analysis shows equally relevant normal and tangential
components from angled velocity vectors (Plane 1); predominant low magnitude
tangential and normal components, represent recirculation (Plane 2); high
magnitudes of normal components show mainly forward flow (Plane 3). Model 2: In average, normal components
revolve around zero along with positive and negative velocities, indicating important
recirculation. Plane 1, with low velocity, shows low magnitude of both normal
and tangential components; the double peaks in opposite directions in the normal
component histogram, along with low tangential velocities, shows markedly
recirculating flow (Plane 2); Plane 3 shows high normal and tangential
velocities, as still part of the recirculating pattern. The small effect of cardiac
phase is suspected to be due to the absence of an aortic valve.Conclusions
AD
false lumen velocity magnitude and directionality information can be extracted
from 4D Flow MRI data. Assessment of normal and tangential component histograms
along the lumen might be a valuable tool for false lumen hemodynamics characterization,
bringing insight into thrombus formation and disease progression. Venc
selection is important in false lumen flow assessment, especially of tangential
components, and might have implications on further hemodynamic studies.Acknowledgements
David
Rutkowski (UW Mechanical Engineering); UW Radiology R&D 2015; GE
Healthcare.References
1. Khan IA, Nair,
CK. Clinical, diagnostic, and management perspectives of aortic dissection.
Chest 2002;122:311-328.
2. Tsai TT,
Evangelista A, Nienaber CA, et al. Partial thrombosis of the false lumen in
patients with acute Type B aortic dissection. N Engl J Med 2007;357:349-59.
3. François CJ, Markl, M, Schiebler, ML, et
al. Four-dimensional, flow-sensitive magnetic resonance imaging of blood flow
patterns in thoracic aortic dissections. J Thorac Cardiov Sur 2013;145:1359-66.
4. Clough RE, Waltham M, Giese D, et al. A new imaging method for
assessment of aortic dissection using four-dimensional phase contrast magnetic
resonance imaging. J Vasc Surg 2012; 55:914-923.
5. Gu T, Korosec, FR, Block, WF, et al. PC VIPR: a
high-speed 3D phase-contrast method for flow quantification and high-resolution
angiography. AJNR Am J Neuroradiol 2005;26:743-749.