Maria R. Stefanescu1, Jan-Peter Grunz1,2, David Lohr1, Stefan Herz1,2, Aleksander Kosmala1,2, and Laura Maria Schreiber1
1Comprehensive Heart Failure Center, University Hospital Wuerzburg, Wuerzburg, Germany, 2Department of Radiology, University Hospital Wuerzburg, Wuerzburg, Germany
Synopsis
2D phase contrast MRI offers a fast method for
blood flow evaluation in the ascending aorta which can be used for estimation
of left ventricular function. With proper ECG or acoustic triggering, forward
flow in the aortic root (FFAo) should resemble left ventricular stroke volume
(LVSV) in absence of valve disease. However, for ultra-high field strengths (≥7.0T) flow parameters in the ascending aorta derived from phase contrast
have not yet been validated. Our results suggest that accurate gating reduces
cardiac motion artifacts for 7.0T to an extent where LVSV can be estimated
reliably based on FFAo in phase contrast sequences.
Introduction
Cardiovascular magnetic resonance imaging (CMR)
is established as a non-invasive method for quantifying ventricular function
and flow measurement in the aortic root at clinical field strenghts.1,2,3
In order to provide optimal image quality and reduce motion artifacts, data
acquisition has to be synchronized with the beat of heart using a gating method.4
Due to its potential of increased signal-to-noise ratio (SNR), which
facilitates higher resolution and acquisition speed as well as possibly new
diagnostic information, there is growing demand for the application of
ultra-high field scanners in CMR. Technical challenges increase with the
magnetic field strength though.5 We report results of 2D phase
contrast flow measurement in the ascending aorta by quantifying the influence
of using ECG and acoustic (AC) gating, left ventricular stroke volume (LVSV) and
left ventricular ejection fraction (LVEF) with a 7T Siemens Magnetom™ Terra whole-body MRI system. The purpose of this work
was to compare the
impact of proper ECG and external acoustic triggering on flow parameters in the
aortic root of healthy subjects derived from 2D phase contrast imaging at 7.0T
CMR.Methods
We
examined N = 30 healthy volunteers with local ethic committee approval using a
1TX/16RX thorax coil (MRI.Tools, Berlin, Germany). 28 subjects (14 male, 14
female) were included into analysis after two subjects (1 male, 1 female) had
to be excluded due to strong artifacts. Phase contrast sequences with ECG and acoustic triggering (MRI.Tools,
Berlin, Germany) were acquired
(TE = 98.9 ms, TR = 3.2
ms, voxel size 1.5 x 1.5 x 6.0 mm, segments 7 – 9, phases 20 – 25, VENC = 150 m/s).
We used ECG
and acoustic gating in 25 and 19 participants, respectively. Both gating
methods were applied in 16 participants. Additionally, we performed short-axis
GRE CINE imaging (TE
= 3.57 ms, TR = 30.6 – 88.3 ms, voxel size 0.6 x 0.6 x 6.0 mm, segments 6 – 11,
phases 20 – 35, slices 14 – 17) for all subjects during multiple breath-holds
to reduce respiratory motion artifacts. Volumetric
and flow parameters were analyzed using integrated software Argus Syngo MR
(Siemens Medical Systems, Germany) scanner-side.Results
Mean forward flow in the aortic root (FFAo) based
on phase contrast sequences was 82.1 ml (SD 19.3 ml) for ECG and 79.9 ml (SD
23.3 ml) for external gating. For LVSV based on GRE CINE imaging results showed
a mean of 81.3 ml (SD 17.4 ml) and for LVEF of 62.6 % (SD 5.6 %). FFAo with ECG and acoustic triggering
correlated with r = 0.88 (p < 0.01). The mean difference in Bland-Altman
analysis was 6.7 ml with a standard error of estimation (SEE) of 12.2 ml. LVSV
and FFAo (ECG) correlated with r = 0.93 (p < 0.01) and a mean difference of 0.1
ml (SEE 7.2 ml). The correlation between LVSV and FFAo (AC) was slightly weaker
with r = 0.88 (p < 0.01) and a mean difference of 4.7 (SEE 14.3).Conclusion
2D phase contrast flow measurement in the ascending
aorta allows for accurate and robust quantification of left ventricular stroke
volume and showed potential for assessment of valve disease and heart failure
in 7.0T CMR. Both ECG and acoustic gating proved to be reliable methods for
reducing the effect of cardiac motion on PC imaging; however, using ECG
triggering for synchronization provides a slightly more exact estimation of
LVSV than acoustic gating. Synchronization of image acquisition and the cardiac
cycle was more reliable using the ECG signal despite magnetohydrodynamic effects.Acknowledgements
German
Ministry of Education and Research (BMBF) grants: 01EO1004, 01EO1504.References
-
Gatehouse PD, Keegan J,
Crowe LA et al. Applications of phase-contrast flow and velocity imaging in
cardiovascular MRI. Eur Radiol. 2005;15:2172-84.
- Firmin DN, Nayler GL, Klipstein RH et al. In vivo validation of MR velocity imaging. J comput Assist Tomogr. 1987;11(5):751-6.
- Nayak KS, Nielsen JF, Bernstein MA, et al. Cardiovascular
magnetic resonance phase contrast imaging. J Cardiovasc Magn Reson.
2015;17(1):71.
- Jeltsch M, Ranft S,
Klass O et al. Evaluation of Accordance of Magnetic Resonance Volumetric and Flow
Measurements in Determining Ventricular Stroke Volume in Cardiac Patients, Acta Radiol. 2008;49(5):530-9.
- Karamat MI, Darvish-Molla S, Santos-Diaz A. Opportunities and Challenges of 7 Tesla Magnetic Resonance Imaging: A
Review. Crit Rev Biomed Eng.
2016;44(1-2):73-89.