Keita Fukushima1, Katsuhiro Sano2, Haruhiko Machida3, Toshiya Kariyasu3, Isao Miyazaki1, Tatsuya Yoshioka1, Sanae Takahashi1, Saori Yuda1, Yuta Shimizu1, Takayuki Yonaha1, Akihito Nakanishi1, Hiroshi Kusahara4, Youhei Matsuoka4, Miho Kitamura4, Takao Yamamoto4, and Kenichi Yokoyama3
1Kyorin University Hospital, Tokyo, Japan, 2Diagnostic Imaging, Saitama medical University International medical Center, Saitama, Japan, 3Kyorin University School of Medicine, Tokyo, Japan, 4CANON MEDICAL SYSTEMS CORPORATION, Otawara, Japan
Synopsis
Diffusion-weighted imaging (DWI) with an ultrahigh
b value is expected to improve assessment of tumor cellularity and fluid
viscosity in the liver but can decrease signal-to-noise ratio (SNR) of the
hepatic parenchyma. A state-of-the-art 3T MR scanner with the maximal gradient
magnetic field of 100 mT/m can achieve sufficient SNR on liver DWI even at
ultrahigh b value of 3000 with use of short TE. The present study using our
original phantom and healthy volunteers shows that use of shorter TE
significantly increased the SNR with preserved ADC value on DWI even at
ultrahigh b value of 3000.
Introduction
Diffusion-weighted imaging (DWI) with a high b
value of around 1000 sec/mm2 is commonly acquired to assess tumor
cellularity and fluid viscosity in 3T MR examinations of the liver. Use of a higher
b value is expected to improve this assessment but can degrade the image
quality and interpretability with decreased signal-to-noise ratio (SNR) of the
hepatic parenchyma. Recently, a state-of-the-art 3T MR scanner (Vantage Galan
3T/ZGO; Canon, Tochigi, Japan) with the maximal gradient magnetic field of 100
mT/m has been clinically introduced. This high gradient magnetic field allows
the selection of an ultrahigh b value (e.g., b = 3000) without the need to
increase echo time (TE) on liver DWI with preserving the SNR. The purpose of
the present study was to assess impacts of different b and TE values on quality
of liver DWI and to prove feasibility of the ultrahigh-b-value DWI using this
MR scanner.Methods
We originally generated a phantom consisting of
seven tubes of 3-cm diameter containing gelatin/distilled water with various
weight percentages (0, 10, 20, 30, 40, 45, and 50%), one at the center and six
symmetrically placed around it, in a polyethylene container of 20-cm diameter
filled with superabsorbent polymer (Figure 1). With this MR scanner, we scanned this
phantom with various T2 (31.5-868.6 msec) and apparent diffusion coefficient
(ADC) values (0.45-2.01 x 10-3 mm2/sec) using body (Atlas
SPEEDER body) and spine coils (Atlas SPEEDER spine) and acquired DWI using a
spin-echo echo planner imaging sequence with the following scan parameters:
repetition time (TR), 7000 msec; flip angle, 90/180 degrees; field of view, 33
x 36 cm; matrix size, 160 x 160; b value, 1000 and 3000 sec/mm2; TE,
44 (b = 1000), 53 (b = 3000), and 70 msec (b = 1000 and 3000). We placed
regions of interest (ROIs) in the cross-section image of each tube phantom and
background to measure the SNR and ADC values. We also scanned 10 volunteers (5
men and 5 woman; age, 69.1±19.0 years) and 11 patients who had hepatic masses (5
men and 6 women; age, 72.8±9.7 years) including hepatocellular carcinoma in one
patient; metastatic tumor, one; focal nodular hyperplasia, one; hemangioma,
two; complicated cyst, one; and simple cyst, five. We placed four ROIs in the
both hepatic lobes in each volunteer and one ROI in a hepatic mass in each
patient. We calculated SNR and ADC values at different b and TE values. We used
paired t test to compare the SNR and ADC values of the hepatic parenchyma and
masses between different TE values (TE 70 vs. TE 44 or 53) at the same b value.Results
In the phantom study (Figure 1),
the SNR at TE 44 was higher than that at TE 70 at b value of 1000. The
SNR at TE 53 was higher than that at TE 70 at b value of 3000. The
ADC was comparable between different TE values at the both b values. The mean
SNR of the hepatic parenchyma in volunteers was significantly higher at TE 44
(94.7±15.8) than at TE 70 (35.9±7.1) at b value of 1000; and at TE 53
(18.9±4.4) than at TE 70 (12.4±3.2) at b value of 3000 (p < 0.05 for both) (Figure 2). The mean ADC was comparable between different TE
values (p > 0.05 for both) (Figure 3). The mean
SNR of the hepatic masses was also significantly higher at TE 44 (122.1±61.7)
than at TE 70 (86.5±52.0) at b value of 1000; and at TE 53 (25.7±17.4) than at
TE 70 (19.4±11.8) at b value of 3000 (p < 0.05 for both) (Figure 2). The mean ADC was comparable between different TE
values (p > 0.05 for both) (Figure 3).Discussion
Use of short TE values offers higher SNR on DWI
than at TE 70, commonly used for liver DWI, without affecting ADC values;
actually, can reveal the hepatic parenchyma even at ultrahigh b value of 3000 (Figure 4) and more accurately differentiate malignant (e.g.,
metastatic tumor) from benign hepatic masses (e.g., hemangioma) at the b value
(Figure 5).Conclusion
Use of short TE values can achieve sufficient SNR
on liver DWI even at ultrahigh b value of 3000 with this MR scanner.Acknowledgements
No acknowledgement found.References
No reference found.