Yajing Zhang1, Jiazheng Wang2, and Chenguang Zhao1
1Philips Healthcare, Suzhou, China, 2Philips Healthcare Greater China, Beijing, China
Synopsis
We
have developed a novel 3D radial sequence for motion insensitive MRI, which
replaces the frequency encodings in the radial plane in prior-art stack-of-star
sequences with stepwise phase encodings to reduce the streaking artifacts that
can arise from chemical shifts and system imperfections. The sequence achieved
better image homogeneity with less imaging artifacts when compared to the
prior-art sequence at 1 mm isotropic resolution with golden angle acquisition,
both in phantom and in human brain and abdomen imaging.
Introduction
Stack-of-star (SoS) radial imaging has been
established as a standard technique for T1 weighted imaging in head and abdomen1 for its motion immunity when Golden Angle scheme is used together with its
spoke acquisition2. However, as the readout direction rotates, the k-space
trajectory and phase errors, arising from gradient chain delays, eddy currents
(EC), chemical shifts and B0 field inhomogeneity, would lead to misaligned
radial k-space, and consequently image inhomogeneity and streaking artifacts
after gridding recon. These artifacts are particularly common in imaging areas
such as abdomen and cardiovascular, despite the pre-reconstruction efforts on
the clinical scanners to remove the 0th, 1st and high orders
phase errors. Here we present a cylinder-like acquisition trajectory such that
each 2D radial k-space plane is stepwise phase-encoded, with infinite bandwidth
and hence improved immunity to system imperfections. A similar method has been
used in a volumetric scanning for motion detection and compensation3. In the
present study, we focus on exploring the advantages of cylinder scan in
comparison to SoS. Experiments were performed on phantom and human subjects.
Results indicated that the cylinder methods was helpful in reducing streaking
artifact level and improving image uniformity.Methods
The prior-art SoS 3D k-space and the proposed
alternative are depicted in Figure 1A and 1B, respectively. In prior-art
(Figure 1A), each kxy plane contains multiple readouts, where each
readout passes the 2D k-space center and is rotated by a golden angle from its
adjacent readout, while the 3rd dimension is filled with stepwise
phase encoding. In the proposed cylinder trajectory (Figure 1B), the readout
direction coincides with the 3rd dimension while the 2D radial plane
is stepwise phase-encoded, and each two adjacent Cartesian plane is still
rotated by a golden angle.
The proposed trajectory was implemented into a
spoiled gradient echo sequence on a 1.5 T Multiva system (Philips Healthcare
Suzhou, China) to compare with the prior art trajectory, and experiments were
performed on both phantom and healthy volunteers with informed consent (on both
head and abdomen). The imaging protocol was approved by a local Ethical Review
Body, with 4.2 ms TR, 1.7 ms TE, and 110% angle percentage for both
trajectories to achieve 1 mm isotropic voxel size. The angle percentage
describes the ratio between the number of angles in the radial plane and the
required number of phase encoding lines for the same plane if acquired with
Cartesian k-space. Golden angle was used such that each two adjacent Cartesian
planes were separated by 111.246°. Data acquired from both methods went through
the gridding reconstruction and channel combination implemented in Philips
recon 2.0 (Philips Healthcare Best, the Netherlands).Results
Exemplary phantom images from the SoS
and the cylinder acquisitions are shown in Figure 2A and 2B, respectively. SoS
images turned out to have apparent signal inhomogeneity, streaking artifacts
(short arrow), and blurring of fine structures (long arrow), which are much
less observable in the cylinder images. Figure 2C and 2D demonstrate the
k-space lines acquired from both methods that were used to fill the radial
plane in the middle of the 3rd dimension. Each column in the figures
is a spoke in the radial plane, with phase being color-coded, showing better
alignment of k-space lines in the proposed trajectory. Figure 3A, 3B, and 3C
show the brain images from the SoS without phase correction, the cylinder
acquisition, and the SoS with phase correction, respectively. While phase
correction improved image uniformity for SoS acquisition, the loss of fine
structures sustained in areas like frontal sinus (arrow). These problems were
not observed in the cylinder image (Figure 3B). Figure 3D and 3E display the
images acquired on human abdomen with free breathing, acquired with SoS (phase
corrected) and cylinder respectively. While both methods showed motion
insensitivity, the cylinder image was less contaminated by streaking artifacts.Discussion and Conclusion
The proposed cylinder trajectory reduces
streaking artifacts and improves true image resolution when compared to the
prior-art stack-of-star trajectory, while the motion immunity and the imaging
speed was retained. This technique holds clinical potential particularly for
regions where motion insensitivity is preferred and local B0 field
inhomogeneity presents. Development of trajectory-specific phase correction
method would be favorable for the proposed acquisition scheme in the future to
further improve imaging quality. Acknowledgements
No acknowledgement found.References
1. H. Chandarana, et al. Free-breathing radial 3D fat-suppressed T1-weighted gradient echo sequence: a viable alternative for contrast-enhanced liver imaging in patients unable to suspend respiration. Invest. Radiol. 2011;
46: 648-653.
2. S. Winkelmann, et al. An Optimal Radial Profile Order Based on the
Golden Ratio for Time-Resolved MRI. IEEE Trans Medical Imaging
2007; 26 (1)
3. W. Lin. et al. Fast, Robust and Self-Navigated 3D Cylindrical Imaging: MP-RAGE and FLAIR. ISMRM.