Synopsis
Phase
only B1 shimming is a cost-effective and simple approach to improving RF field homogeneity
for high field MRI. Without a multiple
channel transmitter, this is implemented
by switching transmission lines or possibly lumped element phase shift
networks. This abstract investigates
the minimum phase shift required in a potential multi-bit electronic phase
shifter. For practical shimming
solutions, those not requiring significant increases in power to achieve a 90
degree tip angle as compared to ‘birdcage’ or conjugate phase currents, it is
concluded that 45 degrees may be a sufficient resolution for such a phase
shifter.
Introduction
Many
approaches to the mitigation of B1 inhomogeneity at high fields have been
investigated, including the use Transmit SENSE1-4, B1 shimming using
either amplitude and phase or just phase control5,6, dielectric
inserts7, and time sequencing of coils8,9. Phase only
shimming can be very effective, but is not trivial to implement at high powers
without multiple transmitters. We are implementing an electronically controlled
phase shifter10 to potentially enable dynamic B1 shimming on systems
without multiple channel transmitters. Here we investigate the minimum phase control
bit required for a switchable phase shifter to be used for B1 shimming.Methods
Multi-bit
phase shifters can be implemented by switching transmission lines or lumped-element
phase delay networks. As each ‘bit’ of phase control adds insertion loss and
complexity, it is desirable to use as few bits as possible. We simulated the effectiveness of B1
shimming solutions as a function of the minimum phase control bit employed, φmin. Two models were simulated using in-house
developed full-wave FDTD software and optimization programs. One, a 37x26 cm
uniform phantom inside a 12 rung volume array, was used for comparison and validation6.
The other was a human model exported from a commercial FDTD program. Reference
shim solutions in each case used ‘conjugate phase’, where the phases of each rung
are adjusted to be in-phase at the center of the optimization region. Phase only optimizations were performed for
tip angle homogeneity, with a 90 degree tip at the center of the region of
interest (ROI), over different ROIs, with and without constraints on maximum
local SAR. For each optimization case
examined, the standard deviation (STD) of the tip angle over the ROI, maximum
local SAR, and “power factor”, defined as the ratio of the power required to
generate a 90 degree tip in the homogeneity region to the power required for
the reference conjugate phase currents. Finally, for each case the sensitivity
to minimum phase control bit was studied as follows. For each value of minimum phase control bit φmin, a
different random phase of +/- φmin/2 was introduced to each rung in the array. 10000
simulations were rung for each value of φmin, and the worst case of tip homogeneity
(standard deviation), maximum SAR and power factor were recorded.Results
Figure
1 compares the tip angle map from the reference conjugate phase determined currents
to those obtained from a phase only optimization for homogeneity only, and
another solution placing equal “weight” on constraining maximum local SAR. Both
provide significant improvement over the reference phases, but removing the
constraint on SAR increases the power to achieve a 90 degree pulse by a factor
of nearly 30. Adding a constraint on SAR improves the situation dramatically,
as seen in Fig. 1c. Figure 2 compares the results of adding 10000 random phase
variations for each value of minimum phase control.. As can be seen, the
constrained optimization is relatively insensitive to the phase error, while
the unconstrained optimization is highly sensitive. Figures 3 and 4 repeat this
analysis for another case, optimizing the B1 shim over a 6 x 6 cm circular ROI
including the prostate. Figure 3a shows the tip angle map from the reference
solution, Fig. 3b the map for a phase only optimization with no constraint on
SAR, and Fig. 3c the corresponding map with a weight on SAR minimization. Note
that the scale on Fig. 3b has been chosen to highlight the ROI. Figures 4a and
4b repeat the random phase error analysis of Fig. 2 for this model. The dashed
lines indicate the tip angle homogeneity (STD) and max local SAR from the reference
solution. Again, the unconstrained optimization, aside from being unacceptable
due to high power requirements, is highly sensitive to phase error, while the
constrained optimization is much less sensitive. Additionally the constrained optimization
improves both max local SAR and homogeneity over the region of interest while
requiring only 64% more power.Conclusion
Simulations
indicate that a three bit phase shifter with phase bits of 45, 90 and 180
degrees, providing a maximum phase error of +/- 22.5 degrees on any rung, should be sufficient for practical B1 shimming solutions. This may be insufficient for unconstrained optimizations, which would require more accurate phase control. However, unconstrained
solutions are likely not practical for high field body MR, the target of this
investigation, due to very high power multipliers that result.Acknowledgements
Support from the Cancer Prevention and Research Institute of Texas through research grant RP160847 is gratefully acknowledged.References
1. Snyder,
C.J., et al., Comparison between eight‐
and sixteen‐channel TEM transceive arrays for body imaging at 7 T. Magnetic
Resonance in Medicine, 2012. 67(4):
p. 954-964.
2. Hollingsworth, N., et al. An eight channel transmit system for
Transmit SENSE at 3T. in Biomedical
Imaging: From Nano to Macro, 2011 IEEE International Symposium on. 2011.
3. Zhang, Z., et al., Reduction of transmitter B1 inhomogeneity
with transmit SENSE slice‐select pulses. Magnetic Resonance in Medicine: An
Official Journal of the International Society for Magnetic Resonance in
Medicine, 2007. 57(5): p. 842-847.
4. Katscher, U., et al., Transmit sense. Magnetic Resonance in
Medicine: An Official Journal of the International Society for Magnetic
Resonance in Medicine, 2003. 49(1):
p. 144-150.
5. Metzger, G., et al. Local B1 shimming for imaging the prostate
at 7 Tesla. in Proc 15th Annual
Meeting ISMRM. 2007.
6. van den Berg, B., et al., 7 T body MRI: B 1 shimming with simultaneous
SAR reduction. Physics in Medicine and Biology, 2007. 52(17): p. 5429.
7. Haines, K., N. Smith, and A. Webb, New high dielectric constant materials for
tailoring the B1+ distribution at high magnetic fields. Journal of magnetic
resonance, 2010. 203(2): p. 323-327.
8. Orzada, S., et al., RF excitation using time interleaved
acquisition of modes (TIAMO) to address B1 inhomogeneity in high-field MRI.
Magnetic Resonance in Medicine, 2010. 64(2):
p. 327-333.
9. Thulborn, K.R., et al., SERIAL transmit–parallel receive (STxPRx) MR
imaging produces acceptable proton image uniformity without compromising field
of view or SAR guidelines for human neuroimaging at 9.4 Tesla. Journal of
Magnetic Resonance, 2018. 293: p.
145-153.
10. Sun, C., et al., Digitally Controlled High-Power Phase Shifter for B1 Shimming at 7T. Proc. Intl. Soc. Mag. Reson. Med.,
2019, submitted.