Yunkyoung Ko1, Chang-Hoon Choi1, N. Jon Shah1,2,3,4, and Jörg Felder1
1Institute of Neuroscience and Medicine - 4, Forschungszentrum Juelich, Juelich, Germany, 2Institute of Neuroscience and Medicine 11, INM-11, JARA, Forschungszentrum Jülich, Juelich, Germany, 3JARA - BRAIN - Translational Medicine, Aachen, Germany, 4Department of Neurology, RWTH Aachen University, Aachen, Germany
Synopsis
The crossbar-type matrix creates open-stubs of varying length, depending
on the switch configuration, which potentially degrades the MR image quality. However,
this issue can be overcome efficiently with the use of a compensation circuit. But
as the number of receive coils increases, a larger number of RF switches is
required. In this study, we propose a miniaturised crossbar switch matrix which
employs two independent, stacked boards.
Introduction
In recent years, the use of multi-channel phased array coils has become
common practice in MRI. This is largely due to the outstanding improvements
they offer in terms of signal-to-noise ratio (SNR) and in the feasibility of accelerating
acquisitions1. Receive coils are frequently connected via
crossbar-type switch matrixes. These use RF switches to route the receive
signals to appropriate receive paths2, which reduces the number of
hardware circuits required. This crossbar-type matrix, however, creates
open-stubs of varying length, depending on the switch configuration, which
potentially degrades the MR image quality. However, this issue can be overcome efficiently
with the use of a compensation circuit3. As the number of receive coils
increases, a larger number of RF switches is required. Thus, the size of the switch
matrix board gets larger, causing elevated losses on the long transmission
lines of the PCB. For example, a 128 (input) x 64 (output) matrix requires 8192
switches, and the size of the board can be as big as 2560 mm x 1280 mm (assuming
a spacing of 2 cm between neighbouring transmission lines - which is realistic
to avoid cross talk between long parallel lines). In this study, we propose a miniaturised
crossbar switch matrix which employs two independent, stacked boards.Methods
Fig. 1(a) shows the
proposed crossbar switch matrix configuration which was modified from the
conventional, dual layer switch matrix. In the novel design, two independent matrix
boards were stacked together via an insulator between the boards and two switch
blocks were used (Fig. 1(b)). The switch blocks #1 and #2 allow two independent
boards to be connected, in which the switch block 1 is located on the input port,
enabling one to select between the input, open, and compensation capacitor.
Switch block #2 is located between the two boards, enabling the selection of
open, compensation capacitor, and connecting switches A and B. Fig. 2 shows a 16
x 8 (f = 400MHz) switch matrix created by stacking two 8 x 4 switch matrices
built from double-layer FR4 PCBs and implemented for validation purposes. In
order to evaluate the quality of the proposed design, the insertion losses of
the implemented switch matrix in all channels were measured using a network
analyser (ZVR, Rhode & Schwarz, Germany). Results
Fig. 1(b) shows an example of the 8 x 4 switch matrix operation. Table 1
gives the configuration of switch block #1 and #2 for some exemplary
input-output configurations. Note that with the proposed switch topology, it is
possible to connect any input port with any output port.
Fig. 3 shows the measured insertion loss of the prototype switch matrix.
The insertion losses could be maintained below 2 dB when using the proposed compensation
method. Fig. 4 illustrates the space savings for the proposed matrix
implementation in comparison with the conventional design. This result shows
that, even if the number of channels increases, the crossbar-type switch matrix
can be applied and it maintains its flexibility.Discussion
A previous study3 has demonstrated that crossbar switch
matrixes can be implemented in ultra-high field MRI when proper compensation is
applied. However, as the number of channels increases, the number of switches
and board size increase, making the conventional implementation costly and
bulky. This miniaturisation and simplification method shows the number of
switches can be reduced by a factor of 2 and the size of the board can be
decreased to ¼.Acknowledgements
We would like to thank Claire Rick for proofreading, and Annette Weber for her assistance in preparing the PCBs used in this study.References
[1] P. B. Roemer, W. A. Edelstein, C. E. Hayes, S. P. Souza, and O.
Mueller, "The NMR phased array," Magnetic resonance in medicine, vol.
16, no. 2, pp. 192-225, 1990.
[2] Kröckel H. "Switching matrix with two control
inputs at each switching element." U.S. Patent 11 283 274, Mar. 23, 2010.
[3] Y. Ko, C.-H. Choi, N. J. Shah, and J. Felder, "Signal Loss
Compensation of RF Crossbar Switch Matrix System in Ultra-High Field MRI",
IEEE transactions on biomedical circuits and systems, 2018.