Shimul Chandra Saha1,2, Roberto Pricci1,2, Maria Koutsoupidou1,3, Ditjoni Katana1, Srinivas Rana3, Helena Cano-Garcia1,2, Panagiotis Kosmas1,3, George Palikaras1,2, and Efthymios Kallos1,2
1Medical Wireless Sensing Ltd, London, United Kingdom, 2Metamaterial Technologies Inc., Dartmouth, NS, Canada, 3Department of Informatics, King's College London, London, United Kingdom
Synopsis
We present a switch matrix for automatic tuning and
detuning of a metasurface-resonator employed to enhance MRI scanning
performance. A digital circuit with an inductor was used to pick up the
magnetic field and generate a clock for switching which does not require any
wire connection to the MRI scanner. The clock was used to activate an array of
MOSFET switches, each of them connected to an adjacent pair of parallel wires
of the resonator. The circuit was tested using a solenoid and the metasurface resonance frequency was successfully tuned and
detuned (cloaked) when the magnetic field was off-on respectively.
Introduction
Magnetic Resonance
Imaging (MRI) is a prominent clinical method for acquiring structural and
functional information of biological tissues1-2. The authors have reported a novel metasurface
(MS) that can enhance the imaging efficiency of MRI3, by boosting local
field intensity up to 300% 4. Currently the device can operate optimally
during MRI scanning at a small flip angle for avoiding tissue over-heating [4]. However, instead of lowering the power level
of the transmitted RF field, it would be more beneficial to amplify the echo
signal from the body during receive-mode and turn off the metasurface during
transmission mode. While MRI dedicated
coils use active switching traditionally5, our metasurface is a standalone wireless device
which does not require any connection with the MRI control mechanism, therefore
an automatic switching is more practical. An effective way of turning off, i.e.
cloaking the effects of resonator device during transmission-mode is by
shorting the metasurface wires for detuning
its resonating frequency. Here a novel switching matrix is presented using
MOSFET as the switching element that can turn off the metasurface in
transmitting mode and tune to Larmor frequency in receiving mode automatically.
The proposed setup has been successfully tested in laboratory conditions using
a solenoid.
Methods
The
metasurface consists of 2x14 wires rods (l=372mm and d=2mm) embedded in deionised
water (Fig. 1)4. Each pair of adjacent wires was shorted both
horizontally and vertically (Fig. 1) and the two setups were simulated. Results
showed shorting horizontally demonstrated the desired resonance frequency shift.
To validate the simulation, we have used a set of push button switches (Fig. 2)
and measured the resonance frequency by turning them on and off simultaneously.
For electronic tuning, MOSFETs operating at 10s of MHz were used. Although
diodes are mainly used as switching elements, the significant advantage of
MOSFETs is its reduced power consumption6. Inductors
were used between gate and source of each MOSFET for biasing (Fig. 3 (a)). A
digital electronic circuit converts a small (tens of mV) RF signal picked by the
inductor to an R-R square wave by a comparator and then down-converts it by a
12-bit asynchronous counter (Fig. 3(b)). A monostable multivibrator down
converts the output to a fixed frequency of ~10 kHz (~RC time-constant) with a
duty cycle of more than 90%. A solenoid fed by an RF generator and an amplifier
has been built to generate the desired magnetic field distribution (Fig 3(c)).
The resonance frequency of the metasurface was measured by a loop antenna using
a vector network analyser while turning on-off the RF source manually. Results and discussions
The simulated resonance frequency
for shorting the metasurface resonator wire rods both horizontally and
vertically clearly demonstrates that vertical
shorting does not shift the resonance (Fig. 4 (a)). When shorted horizontally,
the resonance shifts by approximately 5 MHz. The measured resonance frequency
shift for the on-off state of the duel inline package (DIP) switches connected horizontally
match the simulation results (Fig. 4(b)). For electronic tuning, 14 MOSFETs
with inductors (3.3 µH) were used as the switching mechanism. The inductance
produces a very small impedance (<1 Ohm) at DC and tens of kHz and thus, a
very small voltage drop during actuation. The same inductance generates high-impedance
(>300 Ohm) at 64 MHz, which would act like an open circuit and isolate the
switches from each other. The gates were connected to a common bias line. There
was no contact between the bias line and parallel wires, thus does not affect
the resonance. Initially, a 3.3V DC was used to activate the MOSFETs directly
and to detune the resonance frequency of the metasurface with voltage on-off (Fig.
5). Then, the digital circuit is used with an inductor to detect the magnetic
field generated by a solenoid and an RF source. The digital circuits converted the small AC
voltage picked by the inductor to an R-R wave of 3.3V at 10 KHz, which was used
to activate the MOSFET circuits. Figure 5 shows the resonance of the
metasurface resonator when the solenoid is turned on and off by switching the
RF generator. It can be clearly seen for both for DC and RF generator that the
resonance shifts by almost 5 MHz matching both simulation and DIP switches
results. The switch matrix would enable the device to operate more efficiently
by automatically detuning it at transmission, effectively cloaking its
operation to avoid body overheating, and automatically tuning it back at
receive mode, leading to significant enhancement of SNR and imaging quality of 1.5T
MRI scans.
Acknowledgements
This
work was financially supported in part by Innovate UK under Projects 710771
& 102788.References
1. Khoo V S et al. Magnetic resonance
imaging (MRI): considerations and applications in radiotherapy treatment
planning, Radiotherapy and Oncology. 1997; 42(1):1-15.
2. S Lord S J et al. A systematic review
of the effectiveness of magnetic resonance imaging (MRI) as an addition to
mammography and ultrasound in screening young women at high risk of breast
cancer, European journal of Cancer.2007; 43(13):1905-1917.
3.
ASlobozhanyuk A P et al, Enhancement of Magnetic Resonance Imaging with
Metasurfaces. Advanced Materials.
2016; 28(9):1832-1838.
4.
Saha S C et al.
Evaluation of metasurface resonator for in-vivo imaging at 1.5T. In proceeding
of ISMRM 2017, HI, USA.
5.
Burl M, Zou M X.
Transmit mode coil detuning for MRI systems, Patent number: US6850067 B1, May
2003.
6.
Reykowski A, Housen R. FET switches as detune circuits for
MRI RF coils, Paten pub no: US 2014/0070808 A1, 2014.