This paper investigates a non-contact heart and respiration rate monitoring with antennas. In order to examine antenna effect on the non-contact monitoring, differences of antenna resonant frequency or antenna type are investigated by experiment. In the experiment, it is found that the higher the antenna resonant frequency is, the easier the heart rate is identified. In addition, it is shown that by using only one open circuit end type dipole antenna, both the heart and respiration rate in spine posture can be identified.
Figure 2 shows S11 measurement results of a loop antenna 2 mm apart on the back in spine posture. The loop antenna resonant frequency is 130 MHz. The time response is composed of some periodical waves including a wave with a period of 10 seconds. The frequency response has two peaks at 0.1 and 0.37 Hz. It is considered that the peak at 0.1 Hz indicates respiration rate and the other peak at 0.37 Hz is noise occurred by body motion. The frequency response also has other signals higher than 0.5 Hz. These signals make it difficult to identify heart rate in time response.
Figure 3 shows time and frequency response of S21 measured with two loop antennas resonating at 130 MHz. The left side of fig. 3 is supine posture results and the right side is prone posture results. The respiration rate of both results is identified easier than that in fig. 2. The heart rate of spin posture result is not identified because the spin posture frequency response has two peaks around 1 Hz. While in the case of prone posture, the frequency response has a peak around 1 Hz. Therefore, it is possible to identify the heart rate in time response.
The results measured with two loop antennas resonating at 20 MHz are shown in figure 4. These results are similar to those in fig. 3, but there is a difference. The difference is that the signal variation in time response is small. The small signal variation could lead to failure in the identification of heart rate. From these results in fig. 3 and fig. 4, the higher the antenna resonant frequency is, the easier it is to recognize the heart rate.
Figure 5 shows S11 measurement results with a dipole antenna resonating at 300 MHz on the back (left side) and on the chest (right side). In the case of attaching the antenna on the back, the respiration rate is identified but the heart rate is not identified, because the frequency response has a peak at 0.15 Hz but not around 1 Hz. In the other case, the frequency response has two peaks at 0.15 Hz and 1.28 Hz and both the heart and respiration rate are identified. From these results, it is found that respiration and heart rate in spine posture can be identified by only one dipole antenna.
[1] T. Sakamoto et al. Feature-Based Correlation and Topological Similarity for Interbeat Interval Estimation Using Ultrawideband Radar. IEEE Trans. Biomed. Eng. 2016;63(4):747-757.
[2] K. Nakajima et al. Detection of apparent skin motion using optical flow analysis: Blood pulsation signal obtained from optical flow sequence. Rev. Sci. Instrum. 1997;68(2):1331–1336.
[3] Y. An et al. Sensitivity Enhanced Vital Sign Detection Based on Antenna Reflection Coefficient Variation. Biomed. Circuits Syst. 2016;1082):319-327.