In a low-field portable MRI system, e.g. a permanent-magnet-based system, $$$B_0$$$ is less homogeneous thus $$$T_2^*$$$ is short. However, the ring-down of the transmit RF energy is slow, which dramatically decreases the strength of acquired signals or limits the minimum echo time for a CPMG type experiment. In this abstract, we present the design of a no-tuning inductive-coupling Q-damping circuit for transmit-coils in a low-field MRI/NMR system. The proposed circuit can effectively and quickly damp RF energy yet simple to be implemented and constructed. The effectiveness is successfully demonstrated in simulations and experimentally. In the experiment, it shows that the ring-down time is reduced by about 45%.
In a low-field MRI system, e.g. a Halbach-array-based portable MRI scanner, a $$$B_0$$$ field with a quadrupole pattern is used for encoding [1-3]. $$$B_0$$$ fields are less homogeneous thus $$$T_2^*$$$ is short. On the other hand, a solenoid is used as a transmit-coil [1-3]. The ring-down time of a solenoid is long due to a high inductance of a solenoid, L, and a low working frequency (< 3 MHz) (the decay time constant, $$$\tau= (2Q)/ \omega_0$$$, and Q is the quality factor, $$$Q = \omega_0 (L/r)$$$, where r is the resistance, $$$\omega_0$$$ is the working frequency of the coil). This dramatically decreases the strength of acquired signals. To deal with this issue, an effective damping circuit is needed to accelerate the damping process.
Available solutions to this problem include those based on a resonant circuit that need to be tuned, e.g. those using quarter wave network with a diode expander [4], and those using coupling transformer [5]. There are tuning-free solutions as well, such as those using a short phase-inverted pulse following the main RF pulse [6], and that using a MOSFET switch with a resistive load [7]. Although tuning is not needed, the former method is complex to implement, and the latter only works up to 2MHz.
RESULTS
The effectiveness of the proposed Q-damping circuit was tested by both simulations and experiments. The response of a single pulse in a solenoid is simulated in ADS [9]. Fig.2 shows the simulation result. As shown, with the proposed circuit, the ring-down time is successfully reduced by about 60% compared to that without the circuit. The proposed circuit was tested in a Halbach-array-based portable MRI system working at 2.83MHz [2-3]. Fig.3 shows the damping-pulse (green) and the signal at the receive-coil with and without the proposed circuit (yellow). As shown, the damping time of the received signal with the proposed circuit is reduced from over 20us to about 11us, about 45% compared to that without the damping circuit. Both the simulation and experimental measurement shows the effectiveness in terms of reducing the ring-down time of a transmit solenoid by using the proposed circuit.DISCUSSION & CONCLUSION
We successfully show the effectiveness of the propose Q-damping circuit in terms of reducing the ring-down time for a low-field MRI/NMR system. Both simulations and experiments were conducted. Experimentally, the ring-down time is effectively reduced by about 45%. The proposed method can be applied to a higher frequency. Compared to other damping techniques [4,5,6], the proposed tuning-free method is much easier to implement. Although the reduction in ring-down time using the proposed method (about twofold reduction) is lower compared to that using techniques in [4,5] (up to sevenfold reduction), the performance of the proposed circuit can further be improved by optimizing the inductance of the damping-coil. This will be investigated next.Fig. 2 The response of the transmitter after a single pulse simulated in ADS. The blue curve is the one without the Q-damping circuit, the red one is with the proposed Q-damping circuit, and green curve is the control signal for the MOSFETs switch. As we can see from the simulation results, the ring-down time is accelerated from over 20 us to about 10us with a reduction of over 50%.