Yunkyoung Ko1, Wenwei Bi1, Jörg Felder1, and N. Jon Shah1,2,3,4
1Institute of Neuroscience and Medicine - 4, Forschungszentrum Juelich, Juelich, Germany, 2Institute of Neuroscience and Medicine 11, INM-11, JARA, Forschungszentrum Jülich, Juelich, Germany, 3JARA - BRAIN - Translational Medicine, Aachen, Germany, 4Department of Neurology, RWTH Aachen University, Aachen, Germany
Synopsis
Here we investigate the
feasibility of using existing high data rate wireless transmission protocols - a
wireless gigabit (WiGig) - suitable for transmitting data from high channel
count array antennas. Moreover, the study aims to use off-the-shelf commercial
components that, with only minor modifications, allow for maximum flexibility
of the implementation. The system presented here overcomes these shortcomings
by increasing the antenna gain and provides a practical solution for the
wireless transmission of receive signals.
Purpose
The last decade has seen the
widespread application of phased array coils. This is largely due to their superior
combined SNR performance, compared to single-channel implementations1,
and due to the fact that their independent elements are sufficiently distinct for spatial encoding, which
addresses parallel imaging applications. However, the use of these coils presents a number of drawbacks, including
the handling of receive antennas with bulky cable bundles dangling from
one side, and safety issues arising from shield currents. One possible way to
address these problems is to transmit the receive signals wirelessly, and
recently, a number of designs which implement wireless receive data
transmission for MRI have been published2,3,4. Here we investigate the
feasibility of using existing high data rate wireless transmission protocols - a
wireless gigabit (WiGig) - suitable for transmitting data from high channel
count array antennas. Moreover, the study aims to use off-the-shelf commercial
components that, with only minor modifications, allow for maximum flexibility
of the implementation. The
system presented here overcomes these shortcomings by increasing the antenna
gain and provides a practical solution for the wireless transmission of receive
signals.Methods
Two WiGig USB dongles (Mr.Loop, Taiwan) were used for data transmission
inside the MR scanner room. The WiGig receive dongle was connected to a laptop
computer located at the outer circumference of the MRI scanner room. An ODROID
XU4 single PCB computer (Hardkernel, Korea) was used as the host PC and
transmitted via the on-board USB 3.0 interface. For shielding, the single PCB
computer was placed inside a custom built box (Fig. 1(a)). Fig. 1(b) shows the
location of the WiGig dongles inside the MR room. To evaluate the achievable
data rate, a large file (840 MB) was transferred via WiGig, and the transmission
duration was timed. The procedure was repeated for different distances (d)
between the transmitter and the receiver in a line of sight configuration. The
effect of the WiGig transmission on the SNR of the MRI scanner was investigated
when operating/not operating the WiGig dongle. All scans were performed on a
Magnetom Tim Trio 3T scanner (Siemens Healthcare, Germany) using
a 32 channel receive array coil. SNR measurements used a 170 mm diameter
spherical water phantom doped with NiSO4 x 6H2O, and a
gradient echo (GRE) sequence. The imaging parameters were: TR = 40 ms,
TE = 3.84 ms, number of slices = 1, slice thickness = 5 mm, matrix size = 128 x
128, BW = 260 Hz/Px, FA = 25 deg. Initial investigations showed that the raw
data rate achieved with the system did not reach the desired data rate. Thus, a
simple solution was sought that enabled transmit capabilities to be increased.
The WiGig dongle used has four antennas (Fig. 2(a)) and the antenna gain with
beamforming is about 10 dBi. This can be increased with the self-shielded method5,
in which two metal plates are located on the top and bottom side of the
beamforming antenna array, respectively (Fig. 2(b)). By changing the gap
between the antenna and the metal plates (Fig. 2(c)), the antenna gain can be
changed. We used CST Microwave Studio (Darmstadt, Germany), to analyse antenna
gain and to optimise shielding distance.Results
The simulated array antenna gain of the unmodified WiGig dongle is approximately
10 dBi. With the shields in place, the antenna gain varies with the distance of
the gap g. With g=4.5 mm the antenna gain is increased to around 14 dBi. Table.
1 shows measured data throughput in the MRI room as a function of
communication distance for the unshielded and shielded antenna. The maximum
transmission rate for a maximum distance of 3 m was above 650 Mbps in both
cases. At a distance greater than 3.5 m, the transmission link broke down in
the unshielded configuration. However, when using the shielded antenna array,
the data rate required to sustain high channel MR receive antennas was
sustained up to a distance of 4 m. Fig. 3 shows the acquired MR image when the
WiGig system was transmitting inside the scanner room. Also, the transmission
of RF pulses did not interfere with the wireless data transfer and no reduction
in data rate during MR operation could be observed.Conclusions
We have successfully implemented and tested an off-the-shelf wireless
transmission system capable of sustaining the data rate required for high
channel-count MRI receive arrays. There were no interferences when operating
either system with the performance of the other. The system presented could be
a 'drop-in solution' to reduce cable handling requirements and potential safety
hazards in standard MRI scanners.Acknowledgements
We would like to thank Claire Rick for proofreading, and Annette Weber for her assistance in 3D printing used in this study.References
[1] P. B. Roemer, W. A. Edelstein, C. E. Hayes, S. P. Souza, and O.
Mueller, "The NMR phased array," Magnetic resonance in medicine, vol.
16, no. 2, pp. 192-225, 1990.
[2] Aggarwal K, Joshi KR, Rajavi Y, Taghivand M, Pauly JM, Poon ASY, et
al. "A Millimeter-Wave Digital Link for Wireless MRI," IEEE T Med Imaging.
2017;36(2):574-83.
[3] Wei J, Liu Z, Chai Z, Yuan J, Lian J, Shen GX, "A realization of
digital wireless transmission for MRI signals based on 802.11 b," Journal of
Magnetic Resonance. 2007;186(2):358-63.
[4] Y. Ko, W. Bi, J. Felder, N. J. Shah, "Wireless Digital Data Transfer
using Millimetre-Waves in MRI system," ESMRMB; 2017.
[5] Alhalabi RA, Chiou Y-C, Rebeiz GM, "Self-shielded high-efficiency
Yagi-Uda antennas for 60 GHz communications," IEEE T Antenn Propag.
2011;59(3):742-50.