Xin Huang1, Xi Lin Chen2, and Shiloh Sison1
1Abbott, Sunnyvale, CA, United States, 2Abbott, Sylmar, CA, United States
Synopsis
Huygens
Box is an efficient simulation technique to reduce simulation time and storage
space. This paper uses simulations to investigate the impact of Huygens box on 3T
MRI RF Safety Assessment. The numerical results on ASTM phantom shows the
overall Symmetric Mean Absolute Percentage Error (SMAPE) average on typical MRI
RF simulation is 6.28%.
Introduction
ISO/TS
109741 requires active implantable medical devices (AIMDs) intended
to be used in patients who undergo an MR scan to be designed in such a way to
protect patients from foreseeable hazards. Computational human models loaded
inside 3T MRI RF coils are simulated to determine the exposure E field levels. Huygens Box2 (also known as Total-Field
Scattered-Field (TFSF)3) is an efficient simulation technique to
reduce simulation time and storage space. This approach breaks down the
simulation into two stages: a) Primary simulation of the Excitation Field and
b) Secondary simulation of Huygens source domain. The impact of Huygens box on
E field in MRI RF Simulations is discussed in this paper.
Methods
Sim4Life4, an FDTD-based EM simulation software,
is used for our simulation study. The ASTM phantom 20095 with 10 cm
depth is loaded inside a typical MRI body coil (the bore diameter is 70 cm and
coil length is 60 cm). The properties of the ASTM phantom are set to human
tissue average εr = 78, σ = 0.47 S/m. To simulate a 3T 2-port
multi-channel (MC2) MRI system, both port (denoted as I port and Q port) are
simulated separately.
Two pairs of Sim4Life simulations are simulated:
-
Huygens I and Huygens Q: the empty RF body coil
excited by I or Q port are simulated separately. Then I port and Q port ASTM
phantom simulations are conducted using Huygens box source settings where the
empty I/Q port simulation results are used as excitations.
- non-Huygens I and non-Huygens Q: Use the actual
coil source with I/Q excitation to run I port and Q port ASTM phantom
simulations. This is also known as tuned I and Q.
For each of the body coil simulations, the resonant
frequency simulations are performed to make sure the coil is working at the
correct resonant mode at 3T RF frequency.
For data comparison, the following E field results are
compared for Huygens vs. non-Huygens simulation:
-
CW simulations: the clockwise (CW) circular
polarization simulation with Q to I excitation amplitude ratio 1:1 and phase
different 90 degree.
- CCW simulations: the counter-clockwise (CCW)
circular polarization simulation with Q to I excitation amplitude ratio 1:1 and
phase different 270 degree.
According to ISO/TS 10974, 2 methods of normalization are
applied for each configuration of the E field comparison:
-
B1+avg (B1+ average) scaling: the average B1+
magnitude in the center slice is normalized to the same level (1 µT in this
comparison).
- Whole body SAR (wbSAR) scaling: the whole-body
SAR of the ASTM phantom is normalized to the same level (2 W/kg in this
comparison).
The SMAPE6 (Symmetric Mean Absolute Percentage
Error) values are calculated to measure the overall effect of all E fields (RMS
magnitude) for all 4 configurations mentioned above.
Results
The
data are extracted from Sim4Life simulations. To minimize the effect of large
deviation from small values, the low E field values (lowest 10 percentile of the
overall E field distribution) are omitted.
The
B1+ and wbSAR scaled E field comparison result is shown in Figure 1 for CW simulations and Figure 2 for CCW simulations.
Figure 3 shows
the SMAPE average and standard deviation for all 4 comparison scenarios. The
mean SMAPE average is 6.28%, while the mean SMAPE standard deviation is 5.05%.
Discussion
The
numerical results show a good agreement between Huygens and non-Huygens E
fields at 3T. The assumption of using Huygens box is that the scattered field
does not significantly interfere with the source.Conclusion
The
introduction of Huygens box simulations to 3T MRI RF safety assessment is a
good representation of the actual non-Huygens simulations. The difference
between Huygens and non-Huygens E fields from MRI RF simulations has an
averaged SMAPE of 6.28%.
Acknowledgements
No acknowledgement found.References
-
ISO/TS 10974: 2018 Assessment of the safety of
magnetic resonance imaging for patients with an active implantable medical
device; 2018
- A.
Christ, M. G. Douglas, J. M. Roman, et al, “Evaluation of wireless resonant
power transfer systems with human electromagnetic exposure limits,” IEEE Trans.
Electromagn. Compat., vol. 55, no. 2, pp. 265–274, 2012.
- K.
R. Umashankar and A. Taflove, “A novel method to analyze electromagnetic
scattering of complex objects,” IEEE Trans. Electromagn. Compat., vol. 24, no.
4, pp. 397–405, Nov. 1982.
- Sim4Life. https://zmt.swiss/sim4life/
- ASTM F2182-09, Standard Test Method for
Measurement of Radio Frequency Induced Heating On or Near Passive Implants
During Magnetic Resonance Imaging; 2009
-
Lucano E, Liberti M, Mendoza GG, et al.
Assessing the Electromagnetic Fields Generated By a Radiofrequency MRI Body Coil
at 64 MHz: Defeaturing Versus Accuracy. IEEE Trans Biomed Eng. 2016
Aug;63(8):1591-1601.