Sampada Bhave1, S. Sivaram Kaushik2, Matthew F Koff3, Hollis G Potter3, and Kevin M Koch1
1Medical College of Wisconsin, Milwaukee, WI, United States, 2GE Healthcare, Waukesha, WI, United States, 3Hospital for Special Surgery, New York, NY, United States
Synopsis
The need for examining tissue
integrity near metal implants in post-operative arthroplasty is rapidly increasing.
Quantitative T2 mapping can be beneficial for characterization of
tissues and early diagnosis of tissue pathology. Current T2 mapping
techniques lack the capability to image near metal hardware. A novel 3D fast
spin echo based two-echo multi-spectral T2 mapping technique is
proposed to address this need. This technique is compared to the commercially
available approach for accuracy of relaxometry measurements both in phantom and
human experiments.
Introduction
The
increase of arthroplasty revisions has multiplied the need for imaging near
metal implants to assess post-operative joint integrity [1]. 3D multi spectral
imaging (MSI) techniques like MAVRIC [2], SEMAC [3] and MAVRIC-SL [4] have
successfully mitigated the effects of magnetic susceptibility artifacts due to
metal implants. Quantitative parametric mapping techniques like T2 mapping
are useful in assessing tissue hydration and orientation of collagen, making it
an important biomarker for detecting early-stage osteoarthritis [5]. Several
techniques (Cartigram, Maplt, 3DMAPSS) are available for T2 mapping;
however, they lack the ability to image near metal implants. In this work, a two-echo
multi-spectral 3D fast spin echo (3D-FSE) based T2 mapping pulse
sequence for imaging near metal hardware is described and demonstrated. Methods
The
sequence was developed by modifying the 3D-MSI pulse sequence to serially
acquire images at two echo times within a single scan prescription (Fig 1). In order to acquire all spectral bins for the
second echo time, the number of passes are doubled. The two rows below the RF
waveforms in Fig 1 (rows in blue) and the last two rows (in red) show the
acquisition scheme for shorter echo time (TE1), and longer echo time (TE2)
respectively. The number of skipped echoes are recalculated for TE2. A constant
flip angle (FA) is required to accurately quantify the signal obtained from
both echo times. The refocusing FA of <180o leads to oscillations
in signal intensity from initial echoes, hence we empirically estimate the
number of echoes to skip for TE1 to reach pseudo-steady state. This scheme was
compared to its multi series (5 echoes) sequence and Cartigram on phantom and
knee datasets. The imaging parameters were: FOV=16cm, 128x128 matrix size,
ETL=20, FA=110o, receiver BW = 1kHz/pixel, slice thickness = 4mm,
TEs = (12.54ms, 20.04ms, 25.23ms, 38.16ms, and 46.01ms) for multi-series
variant and (12.54ms, 46.01ms) for the modified two echo sequence.
An in-vivo knee dataset was
acquired using an externally positioned cobalt chrome implant (50 mm diameter
total hip resurfacing acetabular component) to allow for controlled tests with
and without metal artifact at 3T. Imaging parameters: FOV: 16 cm, 192x128
matrix size, 2.5mm slice thickness, and 16 slices. An in-plane phase
acceleration of 2 was used for both cases. 16 spectral bins were collected in
the metal implant setting. Single-acquisition two-echo 3D-MSI T2
maps were also collected on two patients undergoing clinical MRI examinations
of their total hip replacements.Results
Quantitative
comparisons between Cartigram, multi-series 3D-MSI, and the modified two-echo
sequences collected on the T2 phantom are shown in Fig 2. The mean T2
values obtained from the 3 approaches were in good agreement with each other,
with a maximum difference of ~10% demonstrating that the two-echo approach was
sufficient for quantitative assessment. Fig
3 shows the validation of the above techniques on knee with and without metal.
The Cartigram single echo image (3b) shows significant signal void and slice
distortions because of the implant and the trochlear cartilage is lost in the
artifact. On the contrary, the map obtained from the modified two-echo sequence
shows no artifacts in the trochlear cartilage due to the multi-spectral data
acquisition. T2 values in the muscle (shown in green box in (a)) are
comparable to Cartigram, both with and without metal. The modified two-echo sequence
demonstrated on two patients with total hip arthroplasty (THA) is seen in Fig.
4. ROIs 1 and 2 represent a shorter T2 component of the synovium, as
well as extracapsular fat (longer T2 value). ROI 3 identifies posterior short external
rotator tendon, while ROI 4 provides a measurement of extracapsular fat leaked
from the bone marrow into the synovial region.Discussion
In
this work, we have developed an imaging technique to acquire T2 maps
near metal implants by modifying the existing 3D-MSI pulse sequence. Within
this study, it was observed that the conventional sum of squares bin
combination could lead to inflated T2 values because of the
overlapping spectral profiles, especially in regions away from the implants, as
observed in the muscle within the knee experiment. Further investigation on improved methods to
combine the spectral bins will help in improving the accuracy of T2
maps. Conclusion
In
conclusion, with further testing, the proposed T2 mapping technique could
further improve quantitative diagnostic capability of MR to image near metal
implants. Acknowledgements
Research reported in this publication was supported by NIH/NIAMS
R01AR064840. The content is solely the responsibility of the authors and does
not necessarily represent the official views of the NIH.References
[1]
Kurtz,
S., Ong, K., Lau, E., Mowat, F., & Halpern, M. (2007). Projections of
primary and revision hip and knee arthroplasty in the United States from 2005
to 2030. JBJS, 89(4), 780-785.
[2]
Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral three- dimensional
acquisition technique for imaging near metal implants. Magn Reson Med 2009; 61:381–390.
[3]
Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA. SEMAC: slice encoding for
metal artifact correction in MRI. Magn
Reson Med 2009;
[4] Koch, K. M., Brau, A. C., Chen, W., Gold, G.
E., Hargreaves, B. A., Koff, M., & King, K. F. (2011). Imaging near metal with
a MAVRIC‐SEMAC
hybrid. Magn Reson Med, 65(1), 71-82.
[5]
Dunn, T. C., Lu, Y., Jin, H., Ries, M. D., & Majumdar, S. (2004). T2
relaxation time of cartilage at MR imaging: comparison with severity of knee
osteoarthritis. Radiology, 232(2), 592-598.