Feng Wang1, Daniel C. Colvin1, Suwan Wang2, Hua Li1, Raymond C. Harris2, Ming-Zhi Zhang2, and John C. Gore1
1Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN, United States, 2Division of Nephrology and Hypertension, Vanderbilt University Medical Center, Nashville, TN, United States
Synopsis
Currently there are no reliable non-invasive
means for assessing the severity and progression of fibrosis in kidneys. Here
we evaluate spin-lock MR imaging with different locking fields for detecting progressive
renal fibrosis in an hHB-EGFTg/Tg
mouse model. We fit
the dispersion of spin-lock relaxation rates R1ρ at different locking fields
(frequencies) to a model of exchanging water pools and assessed the value of the derived quantities for detecting
tubulointerstitial fibrosis in kidney, including the transverse relaxation rate
(R2), the asymptotic value of R1ρ at high locking frequency (R1ρinf), an exchange rate-weighted parameter Sρ, and the inflection frequencies
ωinfl in the dispersion.
Purpose
Renal fibrosis is a
hallmark of chronic kidney disease, which drives further kidney injury and
leads to renal failure. Excessive tissue scarring or fibrosis is a critical
contributor to chronic kidney disease; however, current clinical tests are
insufficient for assessing renal fibrosis non-invasively. At high fields, the spin-lock
relaxation rate R1ρ varies
with the locking field amplitude (frequency) and analysis of this dispersion
provides information related to exchange
processes between water pools in tissues, such as free water and labile side
groups on macromolecules, and their characteristic mobilities and relaxation
rates. In this study, we evaluated
the ability of parameters derived from spin-lock relaxation dispersion imaging
for detecting tubulointerstitial fibrosis in a murine renal fibrosis model with
selective activation of epidermal growth factor receptor in renal proximal
tubule.1Methods
The hHB-EGFTg/Tg mice, a well-established
model of progressive fibrosis1, at ages 9-12 and 30-33 weeks old, and normal
wild type (WT) mice of similar age, were imaged at 7T. Spin-lock images were acquired in a transverse plane (Fig. 1) using a fast spin echo sequence
preceded by a preparatory spin-lock cluster. Imaging parameters were TR/TE =
3000/24 msec, RARE factor = 8, resolution = 0.25 x 0.25 x 1 mm3,
number of excitations NEX = 12. Sets of images were acquired with different spin-lock
amplitudes ω1 (200,
400, 600, 800, 1000, 1500, 2000, 2500 and 3000 Hz). Spin-lock times were varied
as 1, 5, 15, 25, 35, 55, 75 msec (Fig. 1).
R1ρ values
were obtained for each locking field by fitting signals to a single exponential
decay with locking time. The dispersion of R1ρ with ω1 was
fit to a model proposed by Chopra et al.2 as in previous
studies3.
The fits provided transverse relaxation rate R2,
R1ρ at infinite spin-lock frequency (R1ρinf), and an exchange rate-weighted
parameter Sρ. In
realistic cases, Sρ2 ≈ ksw2 + Δωs2, where ksw is the chemical exchange rate
of protons with resonance frequency offset from water Δωs. The inflection frequencies ωinfl
of the
dispersion curves were also calculated4. The significances of differences in derived
parameters between groups were evaluated using Student’s t-tests. Paraffin
tissue sections were stained with Masson trichrome using standard procedures
for histological confirmation.Results
MR images differentiated cortex from medulla with good
contrast (Fig. 1). To simplify
comparisons between groups, we integrated values over ROIs corresponding to the
cortex and outer medulla (Fig. 2). At
7T all kidneys showed considerable
dispersion in R1ρ with
locking field, deceasing about 25% over the range 400 – 3000 Hz (Fig. 3). Figure 4 summarizes the values
of the derived parameters for WT and diseased animals at different ages. R2
values clearly dropped slightly as fibrosis progressed. There was less
variation in the asymptotic value of R1ρinf between
the groups. These small changes suggest that R1 and R2
relaxation rates are not changed much during fibrosis as total protein and
water content are not changed drastically.
However, the exchange parameter Sρ and the inflection frequencies changed by
much larger factors. To good approximation, both Sρ and
the inflection
frequency depend primarily on the exchange rate between water and other
chemically shifted resonances such as hydroxyls and amides2. This average exchange rate is sensitive to changes
in pH and other physico-chemical aspects of the tissue microenvironment such as
increases in fractions of proteins with slower exchanging or smaller chemically
shifted protons. Histology detected
extensive fibrosis in the hHB-EGFTg/Tg mice (Fig. 5), in which collagen deposition and capillary density
reduction were observed in the fibrotic regions of kidneys.Conclusion
Renal tubulointerstitial
fibrosis in kidneys can be assessed by spin-locking MRI and measures from R1ρ dispersion. This technique provides exchange rate
information in addition to relaxation rates, which may be used as a novel
imaging biomarker to assess chronic renal diseases.Acknowledgements
We thank Mr. Fuxue
Xin, Mr. Ken Wilkens, Mr. Jarrod True, and Dr. Mark D. Does in the Center for
Small Animal Imaging at Vanderbilt University Institute of Imaging Science.
This work was supported by National Institutes of Health grants EB024525 and
DK114809. This work was also supported by grant
1S10OD019993-01 for the Advance III HD Console, housed in the Vanderbilt
Center for Small Animal Imaging.References
1. Overstreet JM, Wang YQ, Wang X, et al. Selective activation of
epidermal growth factor receptor in renal proximal tubule induces
tubulointerstitial fibrosis. Faseb J. 2017;31:4407-4421.
2. Chopra S,
Mcclung RED, Jordan RB. Rotating-Frame Relaxation Rates of Solvent Molecules in
Solutions of Paramagnetic-Ions Undergoing Solvent Exchange. J Magn Reson.
1984;59:361-372.
3. Spear JT, Gore
JC. Effects of diffusion in magnetically inhomogeneous media on rotating frame
spin-lattice relaxation. J Magn Reson. 2014;249: 80-87.
4. Spear JT, Zu ZL,
Gore JC. Dispersion of Relaxation Rates in the Rotating Frame Under the Action
of Spin-Locking Pulses and Diffusion in Inhomogeneous Magnetic Fields. Magn Reson
Med. 2014;71:1906-1911.