Niklas Wehkamp1, Elmar Fischer1, Philipp Rovedo1, Jürgen Hennig1, and Maxim Zaitsev1
1Department of Radiology - Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany
Synopsis
A
new manufacturing method for the creation of magnetic field probes is
presented. The method allows realizing field probes that can be
frequency adjusted during MR acquisition. This opens up new
possibilities for the use of field probes during MR experiments. In
the presented proof-of-concept case, the field probe’s position in
a standard gradient echo experiment was shifted within the field of
view by changing its Larmor frequency using an additional micro-coil.
Purpose:
Magnetic
field probes have great potential to improve MRI measurements.[1]
Hydrogen-based field probes provide the highest signal but can
interfere with the imaging process. Therefore, fluorine, deuterium or
chemically-shifted field probes have been proposed.[2,3]
This work presents newly designed hydrogen-based field probes that
can be frequency adjusted during the measurement. Shifting the
working frequency of a field probe by a known offset can be useful in
many imaging settings. The
Larmor frequency can be changed, thus it is possible to avoid probe excitation
during RF transmission, in addition to the possibility of changing
the apparent position of the probe in frequency encoding direction.
This allows to
use hydrogen-based field probes in cases where the commonly used
fluorine-based field probes are impractical.Methods
The
probes were built in a bottom-up approach. First, a droplet of
UV-curing-glue “Blufixx MGS” from BLUFIXX was cured under a
UV-lamp. Then, the cured droplet was turned upside down and two coils
(receive and B0-modification
coils) were placed on top. The receive coil had an inner diameter of
1 mm whereas the orthogonally oriented outer coil had an inner
diameter of 1.8 mm. Both coils were wound from enameled copper wire
(diameter: 0.2 mm). A second glue droplet was placed on top of the
two coils but not cured. A droplet of water was injected with a
pipette as depicted in Figure 1. The 2.5 – 0.1 µl Eppendorf
pipette was set to deposit 0.65 µl. Note, that due to the viscosity
of the glue, the deposited volume deceeds the set volume.
Subsequently the glue was optically cured. Additional layers of glue
were applied in an iterative process of curing and adding glue until
a spherical shape of the assembly was reached.
The
frequency shift of the new magnetic field probe was measured with
respect to the current flowing through the B0-modification
coil. For each current setting the resonance peak’s maximum was
evaluated in a 1H spectrum.
The
gradient echo (GRE) measurements that were conducted to illustrate
the shift in Larmor frequency of the new field probe were captured
with a FOV 160 mm, FOV phase 100 %, slice thickness 20 mm, TR 100 ms,
TE 6.45 ms and a bandwidth of 100 Hz per pixel.
Results and Discussion
Obtained
frequency shift results are depicted in Figure 2. The coil’s field
shows a linear characteristic in the measured range and exhibits a
slope of 113 Hz/mA.
Figure 3 illustrates the
shifted frequency of the field probe in the GRE measurement. The
numbers in the lower right corner indicate the current that was
flowing through the B0-modification
coil during the measurement. A current of I = 20 mA
shifts the field probe‘s apparent position to the left (in
frequency encoding direction) towards the top of the phantom. If the
current is inverted (I = - 20 mA),
the frequency shift appears to move the probe towards the right of
its actual position that is shown in the I = 0 mA
case. Currently the field probe is implemented as a solenoid
frequency modification coil. In future, Helmholtz or Maxwell designs
will be evaluated to allow for selective dephasing of the probe
signals if required by the target application. This eliminates the
need of using specialized field probes specifically doped on a
per-application basis.Conclusion
The
resonance frequency of our proposed field probe can be changed by
running a current through its B0 - modification
coil. Changing the resonance frequency of magnetic field probes can
be beneficial in manifold ways leading to a more flexible measurement
design and eventually allowing the integration of probes with the RF
coils or the gradient bore. Furthermore, concurrent measurements can
be conducted at a shifted frequency to avoid interference with other
MR measurements.Acknowledgements
The
authors acknowledge Waldemar Schimpf for his excellent tooling
expertise.References
1.
Barmet et al. Spatiotemporal magnetic field monitoring for MR. Magn
Reson Med 2008;60:187–197.
2.
De Zanche et al. NMR Probes for Measuring Magnetic Fields. Magn Reson
Med 2008;60:176-186.
3.
Jorge J et al. Tracking discrete off-resonance markers. Magn Reson
Med. 2018 Jan;79(1):160-171.