Kai Herz1, Tobias Lindig2, Anagha Deshmane1, Benjamin Bender2, Xavier Golay3, Ulrike Ernemann2, Klaus Scheffler1,4, and Moritz Zaiss1
1High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Department of Diagnostic and Interventional Neuroradiology, Eberhard-Karls University Tuebingen, Tuebingen, Germany, 3Institute of Neurology, University College London, London, United Kingdom, 4Department of Biomedical Magnetic Resonance, Eberhard-Karls University Tuebingen, Tuebingen, Germany
Synopsis
Dynamic glucoCEST at
clinical field strengths is very challenging due to the low effect size. Here,
we present a saturation, imaging and post-processing protocol for minimizing possible
artifacts to detect dynamic CEST effects reliably, and demonstrate the application
in two glioblastoma patients at 3 T.
Introduction
GlucoCEST uptake
has already been shown in human brain tumor patients at UHF MRI1–4, yet few first
attempts at clinical field strengths were presented. The low effect size in
combination with many possible sources of artifacts in a long dynamic
measurement (e.g. head movement, B0 field alterations) makes dynamic
glucose enhanced (DGE) imaging at 3 T very challenging. Herein, we
present an imaging and post-processing protocol for maximizing the CEST effect
and correcting for artifacts, to achieve a stable DGE experiment at 3 T, and
show first results in two brain tumor patients.Methods
T1ρ-based DGE
imaging consisted of a presaturation, using a HSExp5 spin-lock pulse (∆f: 2.5 kHz;
µ: 65; twindow: 3.5ms; B1: 4µT; tadia; TSL: 120ms), which
was optimized for 3 T (Figure 1), followed by a 3D single-shot GRE readout6,7 (TE: 2ms; TR: 4ms; 700Hz/px;
FA: 6°; 12 slices; 2x2x5mm3 voxel size). These imaging parameters
were adapted from the original reference to achieve a sufficient SNR at a
reasonable voxel size. Scanning was performed at a 3 T Siemens PRISMA using a
64-channel head coil. Images were acquired at 160 time points before, during and
after a glucose injection (0.3 mg/kg) with 6.3s temporal resolution (Total:
16:45 min) at five different frequency offsets (-300, 0.6, 0.9, 1.2, and
1.5 ppm; 32 images per offset). Three healthy volunteers were scanned to
optimize the motion correction algorithm. The final measurement and
post-processing protocol, including optimized motion correction and a dynamic B0
correction8, was then applied in two
glioblastoma (IDH wild-type, unmethylated MGMT promoter) patients (1: male,
70y, 2: female, 75y). Z images were generated using the corresponding image at
-300 ppm as S0: Zi(∆ω)
= Si(∆ω)/Si(-300
ppm) with 1≤i≤32.
For 0.6, 0.9, 1.2 and 1.5 ppm, ∆DGE was calculated using the mean of
the first five images per offset as a baseline Zref: ∆DGEi(∆ω) = Zref (∆ω) - Zi(∆ω).Results
Optimal motion
correction was achieved with elastix9 (Figure 2). Highest DGE signal
was detected at 0.6 ppm, in accordance with previous simulations (data not
shown). Therefore, following results and figures correspond to DGE(∆ω = 0.6 ppm). Patient 1 showed
a maximum ∆DGE
of 0.38% in a ROI corresponding to a Gadolinium (Gd) enhancing region,
approximately 5 minutes post-injection (Figure 3). A ROI in normal appearing WM
did not show significant uptake at the same time point (∆DGE = 0.03%). In a necrosis
ROI, ∆DGE was increasing as well, but showed an earlier decrease than the
Gd-enhanced region. 6 minutes post-injection, the patient had a severe head
movement, which could not be retrospectively corrected. Patient 2, with the same
tumor histology and grading, showed almost no gadolinium enhancement. Here,
neither a normal appearing WM ROI (max. ∆DGE = 0.06%), nor a ROI in a
hypointense region in a T1-ce image (max ∆DGE = 0.07%) showed
significant DGE contrast (Figure 4). Figure 5 shows the clinical images and an
overlaid ∆DGE-map
4 minutes post-injection for Patient 1 (a-e) and 2 (f-j). Correlation between
Gd-enhancement and DGE can be seen in Figure 5d.Discussion
As shown in
separately submitted Abstracts, an insufficient Motion correction can generate
artifacts that can be misinterpreted as dynamic CEST effects. Investigating and
correcting for motion is therefore a crucial step for a successful DGE
experiment. Even
with motion correction, altered coil sensitivities or slab selection due to the motion can cause contrast changes. Here, normalizing by
an interleaved M0 scan can help to improve the stability of the contrast. Additionally,
B0 field shifts can influence the CEST contrast, and were therefore corrected in
the post-processing protocol we presented herein. All these correction methods
in combination with an optimized saturation and readout phase led to a stable
DGE contrast. First results indicate that the DGE contrast correlates with
perfusion altered by blood-brain barrier disruption, in accordance with
previous work. These preliminary results must be interpreted carefully, as the
effect size was very small at 3 T, and despite all correction steps head
motion can have a comparable influence on the contrast.Conclusion
We demonstrated
that robust DGE experiments can be accomplished at 3 T by optimizing imaging and
post-processing. A maximum effect size of approx. 0.5% could be detected in a
tumor ROI and successfully distinguished from normal appearing regions in the
brain. First results are promising, but must be validated in further
measurements.Acknowledgements
The financial support of
the Max Planck Society, German Research Foundation (DFG, grant ZA 814/2-1), and
European Union’s Horizon 2020 research and innovation programme (Grant Agreement
No. 667510) is gratefully acknowledged.References
1. Xu X, Yadav NN, Knutsson L, et al. Tomography. 2015;1(2):105-114
2. Schuenke P, Koehler C, Korzowski A, et al. Magn Reson Med. 2017;78(1):215-225
3. Schuenke P, Paech D, Koehler C, et al. Sci Rep. 2017;7:42093
4. Paech D, Schuenke P, Koehler C, et al. Radiology. 2017;285(3):914-922
5. Herz K, Gandhi C, Schuppert M, et al. Magn Reson Med. 2018; Epub ahead
6. Zaiss M, Ehses P, Scheffler K. NMR Biomed. 2018;31(4):e3879
7. Deshmane A, Zaiss M, Lindig T. et al. Magn Reson Med. 2018; In press
8. Windschuh J, Zaiss M, Ehses P, et al. Magn Reson Med 2018; Epub ahead
9. Klein S, Staring M, Murphy K, et al. IEEE Trans Med Imaging. 2010;29(1):196-205
10. Cox RW. Comput Biomed Res. 1996;29(3):162-173
11. Collignon, A., Maes, F., Delaere, D. et al. Information processing in medical imaging;1995:263-274