Philipp Backhaus1,2,3, Florian Büther1, Lydia Wachsmuth3, Lynn Frohwein2, Klaus Schäfers2, Sven Hermann2, Michael Schäfers1,2, and Cornelius Faber3
1Department of Nuclear Medicine, University of Münster, Münster, Germany, 2European Institute for Molecular Imaging - EIMI, University of Münster, Münster, Germany, 3Translational Research Imaging Center - TRIC, University of Münster, Münster, Germany
Synopsis
Quantification of the arterial input
function (AIF) in small animals is challenging in both dynamic
contrast-enhanced (DCE)-MRI and in radiotracer studies. We present a novel
extracorporeal circulation mouse model for DCE-measurements of the AIF in mice.
The approach allows parallel measurement of tissue contrast dynamics as well as
of radiotracer AIF using a β-microprobe integrated into the extracorporeal circulation.
Introduction
Quantification of the dynamic
arterial input function (AIF) in small animals is challenging in dynamic contrast-enhanced (DCE-)MRI
and in radiotracer studies. Only few examples of direct AIF-measurements of MR
contrast agent (CA) in mice are published in the literature1-4, each
featuring significant limitations.Methods
Nine Intracranial tumor-bearing mice (female
NMRI Nude, 10 -12 w.o; U87, PSMA-expressing) were anaesthetized (isoflurane,
fentanyl & midazolam) and received an extracorporeal shunt from the femoral
artery to the tail vein. MRI scanning was performed using a 9.4 T MRI (Bruker
BioSpec) with a CryoProbe. The extracorporeal line featured two reservoirs (glass
capillaries, 0.94 mm inner diameter) positioned on the skull in the MRI field
of view (Figure 1). A MR-compatible measuring chamber featuring a β-microprobe
(biospace lab) for β-emitting radiotracers was included in the extracorporeal shunt. Dynamic
MRI scanning of the head was performed for 15 minutes using a 3D FLASH sequence
(80 x 80 x 8, TR 5.019 ms, TE 1.961 ms, flipangle 15°) with a spatial
resolution of 0.175 x 0.175 x 1 mm3 and a temporal resolution of
4.015 s. A 100 µl solution containing CA (Gadovist, 35 µmol/ml) and (for mice with
simultaneous injection) 10-20 MBq F-18-PSMA-1007 was injected into the tail
vein at 1 ml/min for 6 s. Dispersion correction for MRI CA was performed based
on the recorded distinct dispersion effect at the two interspaced reservoirs. In
vitro calibration measurements were performed with defined Gadovist concentrations
in human blood, which was circulated with defined flow velocities using an
injection pump. Calculation of the gadolinium concentration (cGd)
was performed by the formula cGd = (Spost-Spre)
/ (Spre x T10 x r1)5, with Spre and Spost denoting the pre- and
post-contrast signal intensities, respectively, T10 denoting the T1
of pre-contrast blood (2.4 s at 37 °C and 9.4 T, adapted from Dobre et al.6) and r1 the estimated relaxivity at 37 °C and 9.4 T (4.1
s⁻¹·mM⁻¹, extrapolated from Shen et al.7). PET scanning was performed
using the same animal bed as for MRI scanning in a quadHIDAC small animal PET
scanner in list-mode for 35 minutes.Typical transfer
delay between DCE-MRI and PET measurement was 5-10 minutes. Blood was immediately withdrawn after the end
of DCE-scanning for quantification of cGd using mass spectrometry,
for calibration of the β-Microprobe measurements by gamma counting and hematocrite
measurement.Results
The CA AIFs of nine recorded mice show
little noise and typical AIF curve shapes after dispersion correction (Figure
2, A & C). Eight of nine mice show a close range of peak concentrations (0.55
– 0.85 µmol/ml) and
shunt flow velocities (34-58 µl/min) (Figure 3). β-emitting radioactive tracer AIFs can be simultaneously recorded
using a MR-compatible β-Microprobe (Figure 2 A) and mice were transferred into the
PET-scanner immediately after DCE-MRI (Figure 2 B). Significant
inverse correlation between AIF maxima and the delays between the
CA-influx into the two reservoirs was observed (r =
-0.84) (Figure 3 A). The time constant τ for monoexponential
deconvolution was significantly positively correlated with the delay (r = 0.98) (Figure 3 B). The results of mass spectrometry validation show a systematic and
consistent underestimation of the image-derived concentrations (4 mice, ratio mass spectrometry-derived
/ image-derived: 1.57-1.80) (Figure 2 C). However, the MR-based quantification
shows good agreement with circulated human blood with defined CA concentrations
in the range of expected concentrations and flow velocities (Figure 4).Discussion
Our method allows for parallel measurements
of the AIF of MR CA and PET radiotracer as well as CA dynamics in tissue. AIF curves
generated with our method appear to be robust and show low level of noise. The
approach should be well transferable to any tissue/region of interest and makes
measurements of the AIF independent of specific local constraints. A major
limitation of our approach is the need for surgical arterial catheterization
which impedes longitudinal studies. The quantitative precision of the MRI-based
CA quantification needs further evaluation as discrepancies exist between
calibration measurements (Figure 4) and validation by mass spectrometry (Figure 2 C).Conclusions
We present a novel approach for DCE-measurements
of the AIF in mice with conceivable potential compared to so far published
methods. Moreover, we present the first dual recordings of AIFs of a MR CA and a
PET tracer in mice. This supports evaluation approaches to deduce the CA/PET
tracer AIF from one another. Further, it might provide the basis for simultaneous
and integrated modeling of PET tracer and CA kinetics in mice, which is of high
interest in integrated, simultaneous small animal PET/MRI.Acknowledgements
References
- Loveless ME, Halliday J, Liess C, et
al. A quantitative comparison of the influence of individual versus
population-derived vascular input functions on dynamic contrast enhanced-MRI in
small animals. Magn Reson Med. 2012;67:226–236.
- Moroz J, Wong CL, Yung AC, et al. Rapid measurement of
arterial input function in mouse tail from projection phases. Magn Reson
Med. 2014;71:238–245.
- Kim J-H, Im GH, Yang J, et al. Quantitative dynamic
contrast-enhanced MRI for mouse models using automatic detection of the
arterial input function. NMR Biomed. 2012;25:674–684.
- Barnes SL, Whisenant JG, Loveless ME, et al. Practical
Dynamic Contrast Enhanced MRI in Small Animal Models of Cancer: Data
Acquisition, Data Analysis, and Interpretation. Pharmaceutics.
2012;4:442–478.
- Vlachos F, Tung Y-S, Konofagou EE. Permeability assessment
of the focused ultrasound-induced blood–brain barrier opening using dynamic
contrast-enhanced MRI. Phys Med Biol. 2010;55:5451–5466.
- Dobre MC, Uğurbil K, Marjanska M. Determination of blood
longitudinal relaxation time (T1) at high magnetic field strengths. Magn Reson
Imaging. 2007;25:733–735.
- Shen Y, Goerner FL, Snyder C, et al. T1 Relaxivities of
Gadolinium-Based Magnetic Resonance Contrast Agents in Human Whole Blood at
1.5, 3, and 7 T. Invest Radiol. 2015;50:330–338.