Bogdan Dzyubak1, Jiahui Li1, Sudhakar K. Venkatesh1, Kevin J Glaser1, Alina M Allen2, Meng Yin1, and Richard L. Ehman1
1Radiology, Mayo Clinic, Rochester, MN, United States, 2Gastroenterology, Mayo Clinic, Rochester, MN, United States
Synopsis
Evaluation of liver health with quantitative
MRI addresses an issue of major global importance. This work extends ALEC, a previously
validated automated method for defining ROIs and reporting liver stiffness from
MR elastography images, to reporting fat fraction and R2* from multipoint Dixon
images. The tool achieves excellent correlation with an expert reader in 102 clinical
exams and allows multiparametric quantitative liver MRI exams to be analyzed in
a highly reproducible way within 5 minutes.
Introduction
Nonalcoholic fatty
liver disease (NAFLD) is a condition affecting over 20% of the Western
population.1 Its effects manifest as liver fibrosis and increased fat deposition, and
may lead to cirrhosis and loss of liver function. Fibrosis can be accurately
diagnosed with MR elastography (MRE)2 by introducing and imaging acoustic wave propagation in the liver to
calculate liver stiffness. Fat content and R2*, a proxy for iron content which
is another indicator of NAFLD progression, can be accurately quantified using
multipoint Dixon (mDixon) imaging. mDixon acquires multiple echoes with
different TEs and decomposes the acquired data into fat, water, in-phase (water+fat),
out-of-phase (water-fat), fat fraction (PDFF = fat/(water+fat)), and R2*
images. ROIs for MRE and mDixon need to be drawn by experienced readers to
avoid artifacts and calculate the parameters accurately and reproducibly. An
automated MRE analysis method, called ALEC, was previously presented3 and validated within a clinical practice. This study extends ALEC to analyze
mDixon images.Methods
Clinical liver exams
of 102 patients containing 4-slice GRE MRE and mDixon (GE IDEAL) acquisitions were
retrieved with IRB approval and analyzed with the ALEC+mDixon algorithm
summarized in Figure 1. An
experienced reader (JL) drew ROIs to calculate liver stiffness, proton density
fat fraction (PDFF), and R2*. PDFF measurements were also available from a
second reader (SV) in 92 cases and were used to assess inter-reader measurement
variability. The algorithm is summarized in Figure 1.
For MRE, ROIs were
drawn on all 4 acquired slices. For PDFF, the readers located the 8 Couinaud
liver segments, drew an ROI in each, and averaged the values. PDFF within the
liver is considered to be homogeneous, so another common method is to draw ROIs
in a few slices in a large section of the liver, avoiding vessels. The automated
method selected 4 slices which had the most homogeneous tissue at the expected
liver location for the measurement and drew ROIs which avoided vessels,
nonliver tissue, and susceptibility artifacts in those slices.
The intraclass
correlation coefficient (ICC) was used to compare the stiffness, PDFF, and R2*
measurements between ALEC and the primary reader, as well as PDFF between the
two readers. Bland-Altman analysis for % differences was also performed for MRE,
but not PDFF as fat-fraction values below 5%, which are common, are not
considered reliable4 and may be reported
qualitatively in the clinic as having “normal fat content.”Results
ALEC’s MRE agreement
with the reader was excellent and comparable with an earlier study (difference
0 +/- 8.5%, ICC = 0.976 Figure 2) 3. PDFF measurements also
had excellent agreement with the reader (ICC = 0.994 and between readers
(ICC>0.999) (Figure 3). R2* agreement was somewhat lower
but still very high (ICC = 0.914) (Figure
4). Visual inspection did not detect any automated ROIs that included notable
areas outside the liver. An example of the automated MRE+mDixon report is shown
in Figure 5.Discussion
The automated ROI tool
shows excellent agreement with an experienced reader and analyzes clinical data
reliably. mDixon images have much higher contrast and fewer artifacts due to motion,
intensity inhomogeneity, and low-resolution blurring than MRE. Thus, the random-walker
segmentation refinement step was not necessary for their analysis. The simple
local outlier removal method successfully excluded blood vessels (bright lines
in the fat images), inferior lung segments (dark crescents in the water image),
and susceptibility artifacts (bright haze in R2* images near the heart and
lungs). The morphological closing operation ensured that voxels were removed
only if they were part of contiguous areas containing outlying intensities.
Even cases with the largest
algorithm vs. reader PDFF differences did not contain nonliver tissue or
artifacts (Figure 5). So, the
somewhat larger difference between ALEC and the reader, compared to the
difference between readers, is likely attributable to differences in ROI size
and location. R2* differences were a bit larger than for PDFF. Unfortunately,
no inter-reader reference was available for the R2* data.
Using ALEC removes a 10%
inter/intra-reader variability when analyzing MRE images and reduces analysis
time from 10-15 minute. For PDFF, the inter-reader agreement is already
excellent and the analysis takes only about 2 minutes. Nonetheless, the ALEC
extension to mDixon offers a substantial workflow benefit by automating the
analysis for comprehensive multiparametric liver exams. Further, it provides a
natural workflow entry point for tools which predict liver health based on multiparametric
data.
Conclusions
The new automatic MRE+mDixon ROI analysis
algorithm calculates PDFF, and R2* accurately and reliably. The complete ALEC
tool allows multiparametric liver health exams to be analyzed easily and
reproducibly.Acknowledgements
This work was supported by NIH EB07593, NIH EB001981, K23DK115594References
1. Permutt Z, LE T,
Peterson M, Brenner D, Sirlin C, Loomba R. Correlation between liver histology
and novel magnetic resonance imaging in adult patients with non-alcoholic fatty
liver disease - MRI accurately quantifies hepatic steatosis in NAFLD. Aliment Pharmacol
Ther 2012;36:22-29.
2. Muthupillai R,
Ehman RL. Magnetic resonance elastography. Nat Med 1996;2(5):601-603.
3. Dzyubak B,
Venkatesh S, Manduca A, Glaser J, Ehman R. Automated liver elasticity
calculation for MR elastography. J Magn Reson Imag 2016;43(5):1055-1063.
4. Horng D, Hernando
D, Reeder S. Quantification of Liver Fat in the Presence of Iron Overload. J
Magn Reson Imag 2017;45 (2):428–439.