Quincy van Houtum1, Catalina Arteaga de Castro1, Dennis Klomp1, and Wybe van der Kemp1
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
We demonstrate increased 31P
metabolite sensitivity by acquiring 31P signals over a large volume
in the liver of a volunteer using a 31P whole body birdcage coil at 7T and minimize
spatial reach of muscle signal leakage plus investigate frequency alignment due
to B0 homogeneities. Sufficient SNR could be obtained in weighted
average spectrum of all liver voxels and of four local liver voxels. Correcting
B0 inhomogeneities by aligning resulted in a two-fold and 30%
increase compared to non-aligned for the liver voxel average and the local
voxel average respectively. The full setup allowed for full liver coverage and
minimized muscle signal leakage.
Introduction
Conventional therapy in liver
metastasis is often palliative care, as therapy effectiveness can only be
monitored after several months. Metabolic imaging with 31P MRSI can potentially
provide insight in tumor staging and therapy effectiveness, thus aiding in
therapy response monitoring before any morphological changes are visible.
However, current 31P applications in the body are still limited as
conventional surface transceivers used with 31P MRSI, are restricted
to small field of views.
Surface coils require high energetic
adiabatic pulses to provide a uniform flip angle distribution, at the cost of increasing
SAR and thus limiting SNR per unit of time. 31P body coils are
proposed to overcome the need for high SAR adiabatic RF pulses as these provide
uniform transmit RF fields over larger field of views1.
Previous studies have shown
muscle signal leakage into liver spectra, therefore discarding numerous voxels2-4.
Pohmann et al. reported on the feasibility of 31P MRSI in the heart,
using conventional surface coils and weighted k-space sampling acquisition using
a Hamming filter5. In combination with small nominal voxel sizes, the
spatial reach of signal leakage from muscles can be minimized. However, small
voxels result in low SNR, which in combination with the low concentrations of
the observed 31P metabolites, make this approach challenging.
Here, we propose to increase
metabolite sensitivity by acquiring 31P signals from a large volume,
and investigate the need for frequency alignment due to B0
inhomogeneities. Using acquisition weighted 3D spectroscopic imaging in
combination with a 31P whole body birdcage coil, we obtained full
liver coverage at 7T, with the ability to exclude contaminated spectra and
align spectra prior to averaging to counteract B0 offsets over the
large field of view in the liver.Methods
31P MRSI was performed
using an in-house designed 31P whole body birdcage coil integrated
in a 7T MR system (Philips Healthcare, Best, Netherlands) in combination with
two 31P receive coils in quadrature mode. The body coil, tuned at
120Mhz is powered by a 25kW amplifier. Two fractionated dipole antennas were
driven in quadrature transceiver mode to allow simultaneous proton MRI for
anatomy localization and image based B0 shimming6. A
volunteer was positioned in a right decubitus position on the 31P receiver
coil with the liver on top, to reduce motion induced B0
inhomogeneity, and the arms stretched upwards.
Phosphorous (31P)
spectra were acquired using a 3D 31P chemical shift imaging protocol
with Hamming weighted acquisition at a 15mm isotropic nominal resolution using
a flip angle of 8° and a TR of 60ms for optimal SNR. Other CSI parameters were
TE, 0.44ms; Bandwidth, 4800Hz; matrix size, 12x8x8; NSA, 80 and 256 acquisition
points for a total scan duration of 21:48min.
All data were processed in Matlab
2017b. 31P MRSI data were averaged using hamming weighting and
filtering. FIDs were apodized using a 24Hz Gaussian filter and zero filled.
Zeroth-order phase corrections were applied by changing the phase to match the
real signal to the magnitude signal of α-ATP of the non-phased spectrum. Liver
voxels with an SNR of PCr above four plus voxels with an SNR of α-ATP
below 2.5 were excluded from processing. Spectra were aligned to the α-ATP
peaks prior to generating the weighted average of the remaining liver volume
and of four local liver voxels, using the SNR of α-ATP as the weighting factor.
First order phase corrections were applied on both the aligned and nonaligned
averages. Spectral
SNR was calculated by dividing the real part of the signal from a metabolite
peak by the absolute standard deviation of the last 50 noise samples of the
spectrum.Results
Spectra from a large field of
view in the liver were acquired as shown in figure 1. Alignment of the spectra
increased the SNR for all labeled metabolites by 100% on average for all
included liver voxels (figure 2) and 30% on average for the four local liver
voxels (figure 3). Practically no PCr is visible in the spectra from the full
liver nor in the four voxel average relatively close to the muscles.Discussion and Conclusion
The body coil setup,
positioning of the volunteer and chosen protocol parameters allowed for full
liver coverage and minimized signal leakage from the muscles. Averaging of the
liver voxels regained metabolite signal without introducing PCr signal.
Alignment of the spectra prior to averaging to minimize B0
inhomogeneity by respiratory motion increases spectral SNR two-fold. As the
current setup only used two receiver coils, further SNR improvements of the
setup by using a full receiver array may still be expected, possibly increasing
metabolite discrimination7.Acknowledgements
No acknowledgement found.References
1. J. Löring et al.,
“Whole-body radiofrequency coil for (31) P MRSI at 7 T,” NMR Biomed,
vol. 29, no. 6, pp. 709–720, Jun. 2016.
2. M. Chmelik et al., “In vivo
31P magnetic resonance spectroscopy of the human liver at 7 T: an initial
experience,” NMR in Biomedicine, vol. 27, no. 4, pp. 478–485, Apr. 2014.
3. Q. van Houtum et al., “Large FOV phosphor MR Spectroscopic
imaging with multi-transmit proton MR imaging in the liver at 7 tesla”, “ISMRM
26 (2018)”, 0623.
4. J. H. Runge et al., “2D AMESING multi-echo 31P-MRSI of the liver at 7T allows
transverse relaxation assessment and T2-weighted averaging for improved SNR,” Magnetic
Resonance Imaging, vol. 34, no. 2, pp. 219–226, Feb. 2016.
5. R. Pohmann et
al., “Accurate phosphorus metabolite images of the human heart by 3D
acquisition-weighted CSI,” Magnetic Resonance in Medicine, vol. 45, no.
5, pp. 817–826, May 2001.
6. A. J. E. Raaijmakers et al., “The
fractionated dipole antenna: A new antenna for body imaging at 7 Tesla,” Magn
Reson Med, vol. 75, no. 3, pp. 1366–1374, Mar. 2016.
7. L. Valkovič et al., “Using
a whole-body 31P birdcage transmit coil and 16-element receive array for human
cardiac metabolic imaging at 7T,” PLoS ONE, vol. 12, no. 10, p.
e0187153, 2017.