0444

Fast 4D model based eddy current characterization for shim pre-emphasis calibrations
Michael Schwerter1, Markus Zimmermann1, and N. Jon Shah1,2,3,4

1Institute of Neuroscience and Medicine 4, INM-4, Forschungszentrum Jülich, Jülich, Germany, 2Institute of Neuroscience and Medicine 11, INM-11, JARA, Forschungszentrum Jülich, Jülich, Germany, 3JARA - BRAIN - Translational Medicine, Aachen, Germany, 4Department of Neurology, RWTH Aachen University, Aachen, Germany

Synopsis

Eddy current characterizations are needed for pre-emphasis implementations in dynamic shimming applications. However, since a high spatio-temporal sampling of the eddy current fields is required, this a challenging task. Image-based approaches are well-suited for this purpose, but require substantial acquisition times.

This work presents a 2D image-based sampling scheme, which is fast compared to existing 3D techniques and still provides sufficient information for an unambiguous pre-emphasis parameter reconstruction. Moreover, a model-based fit is proposed, which jointly applies the spatial and temporal eddy current model to the acquired data. It is shown, that this approach is well-suited for reducing fitting noise.

Introduction

Spherical harmonic (SH) dynamic shimming1 can improve B0 homogeneity, but requires comprehensive eddy current (EC) characterizations for shim pre-emphasis implementations2. Existing techniques are well-suited for this purpose, but need additional hardware3, undersample the spatial dimension of the ECs4, or require substantial acquisition times5. This work presents a novel EC characterization framework which includes two innovations. 1) Building upon a 3D image-based EC measurement routine5, a novel sampling scheme is proposed, which significantly reduces acquisition times by acquiring only three 2D slices instead of 3D volumes. 2) Using prior knowledge about the temporal EC evolution, a model-based fit is proposed to reduce fitting noise to more robustly reconstruct SH coefficients from acquired EC data.

Methods

The proposed EC characterization is based on measuring phase offsets that are induced in 2D images, which are acquired at different time-points during a shim-pulse induced EC decay. It exploits the fact that an off-center placement of three orthogonal slices introduces sufficient spatial dependence into the data to unambiguously determine their SH components. The sequence and the sampling scheme are illustrated in Fig. 1. An additional reference scan is acquired without the application of a shim pulse and subtracted from the EC data to eliminate phase offsets from sources other than the shim ECs.

Given the acquired EC data, $$$\Delta\text{B}_{0}$$$, and with $$$\text{f}_{n,m}(\cdot)$$$ denoting a SH of order n and degree m and $$$a_{n,m}^{(i)}$$$ and $$$\tau_{n,m}^{(i)}$$$ being the amplitude and time constant of a modulating exponential function with index i, EC fields are modelled at position r and time t following

$$\Delta\text{B}_{0}\left(\mathbf{r},t\right) = \underbrace{\sum_{n=0}^{\infty}\sum_{m=-n}^{n}\text{f}_{n,m}\left(\mathbf{r}\right)}_{\text{Spatial Term}}\cdot\underbrace{\sum_{i=1}^{\infty}a_{n,m}^{\left(i\right)}\cdot\exp\left(-\frac{t}{\tau_{n,m}^{\left(i\right)}}\right)}_{\text{Temporal Evolution}}.$$

Conventional EC processing routines regard the spatial and temporal components independently, by first performing a SH decomposition at each time-point and then fitting the temporal model to the SH coefficients. However, because the SH fit is an ill-posed, inverse problem, this can introduce strong fitting noise.

When analyzing all data points simultaneously, through joint application of the spatio-temporal model, this noise can be reduced. Let $$$\bar{\mathbf{A}}$$$ be a block diagonal matrix of copies of the SH system matrix $$$\mathbf{A}$$$, $$$\mathbf{x}$$$ be a vector of all SH EC amplitudes at each time-point and $$$\mathbf{b}$$$ be a vector of the measured EC field at each time-point. Applying a linear operator, $$$\mathcal{H}(\cdot)$$$, the time-course of the amplitudes of the EC terms can be rewritten as a stack of Hankel matrices. Minimizing the sum of the nuclear norms of these matrices enforces the time-course of the data to be approximated with as few damped exponentials as possible. Data consistency can be established to a desired accuracy, $$$\epsilon$$$, leading to the optimization problem

$$\underset{x}{\text{min}} \left\| \mathcal{H}(\mathbf(x)) \right\|_{*,1} \quad \text{s.t.} \quad \left\| \bar{\mathbf{A}}\mathbf{x}-\mathbf{b} \right\|^{2}_{2} \leq \epsilon^{2}.$$

The optimization was implemented using Bregman iterations with ADMM updates. Simulations were performed and EC data was acquired on a 3T TRIO (Siemens) equipped with a dynamically-driven high-order SH shim insert (RRI). Using a matrix size of 32x32 at 3 mm isotropic resolution and sampling 3000 ms of the EC decay, the total scan time was 9:36 min per shim.

Results

To test the applicability of the proposed sampling scheme, the pre-emphasis module of a C3-shim was deliberately misadjusted to generate artificial ECs with a known time-course. Fig. 2 shows the acquired data and illustrates the slice-based sampling of the high-order SH ECs.

The performance of the proposed model-based EC fit was compared in simulations to the conventional approach using realistic EC parameters6. Fig. 3 illustrates simulated data for a low-amplitude EC decay generated by a C2-shim, confounded by strong additive Gaussian noise (σ=2.5 Hz). Based on this data, Fig. 4 compares the SH decomposition performed with the proposed model-based approach and the conventional approach. The model-based approach is suitable to reconstruct the input parameters with an average accuracy of 99% as compared to 84% for the conventional approach.

Fig. 5 shows an experimentally acquired EC decay induced by the Z3-shim and demonstrates that the individual EC components are robustly reconstructed


Discussion and conclusions

Image-based EC measurements need no additional hardware, but data acquisition is time-consuming. Compared to a 3D approach, the presented sampling scheme substantially reduces acquisition times while still providing sufficient information for a pre-emphasis parameter reconstruction. Considering equal, isotropic resolutions, the acquisition time is reduced by a factor of NPE/3 (NPE ≙ number of phase encoding steps).

The proposed model-based EC processing has shown to substantially reduce fitting noise, thus to be applicable to even recover low-amplitude EC terms in the presence of high noise-levels. Being independent of the data acquisition scheme, it can also be applied to non-image-base EC measurement techniques.


Acknowledgements

No acknowledgement found.

References

1. Blamire, A. M., Rothman, D. L. & Nixon, T. Dynamic shim updating: A new approach towards optimized whole brain shimming. Magn. Reson. Med. 36, 159–165 (1996).

2. Jehenson, P., Westphal, M. & Schuff, N. Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systems. J. Magn. Reson. 90, 264–278 (1990).

3. Vannesjo, S. J. et al. Gradient and shim pre-emphasis by inversion of a linear time-invariant system model. Magn. Reson. Med. 78, 1607–1622 (2017).

4. Terpstra, M., Andersen, P. M. & Gruetter, R. Localized Eddy Current Compensation Using Quantitative Field Mapping. J. Magn. Reson. 131, 139–143 (1998).

5. Bhogal, A. et al. Image-based method to measure and characterize shim-induced eddy current fields. Concepts Magn. Reson. Part A 42, 245–260 (2013).

6. Juchem, C. et al. Dynamic shimming of the human brain at 7 T. Concepts Magn. Reson. Part B Magn. Reson. Eng. 37B, 116–128 (2010).

Figures

(A) After an EC generating shim pulse, the same slice is repeatedly acquired N times. Each time, the same line in k-space is read at a different time-point during the EC decay. This process is repeated for each of M phase encoding steps, providing a full 2D EC image at each sampled time-point. (B) Different SHs can have the same functional form when being analyzed over 2D slices and can have a zero-crossing in iso-centered slices. Therefore, three orthogonal, off-center slices are acquired at each time-point for each shim to be analyzed

(A) An artificial EC decay was generated by adjusting the pre-emphasis module of a C3-shim so as to generate an exponentially decaying shim current corresponding to an amplitude of 0.2 Hz/cm3 and a time constant of 220 ms. The C3-shim was chosen, because it generates negligible ECs itself. The fit revealed, that the entered pre-emphasis parameters could be reconstructed with very high accuracy (0.195 Hz/cm3 and 221 ms). (B) Exemplary 2D EC maps are shown for five time-points (marked in green in the plot above). The image data shows, that the sampling scheme is applicable for resolving high-order SH terms.

(A) Simulated ECs of a C2-shim with Y and B0 cross-term components. To demonstrate the performance of the model-based fit under difficult conditions, the EC amplitudes were set such as if they were induced by a shim step of only 10% of the maximum shim strength. (B) Individual and superimposed EC maps, simulated from the EC curves above. The transversal, sagittal and coronal slices in each column show the EC maps at three time-points, illustrating the decay characteristics of the individual terms. The noisy superimposed maps reveal the noise-level of the data, which is subsequently used in the proposed optimization.

(A) The proposed model-based fit can accurately reconstruct the low-amplitude EC parameters, despite the presence of high noise-levels in the input data. The reconstructed pre-emphasis parameters show minimal deviation from the input values. (B) A corresponding conventional SH decomposition at each time-point provides insufficient data quality for an accurate EC parameter estimation. Especially the noise-level of the Y-term component is too high to accurately reconstruct the low-amplitude EC decay of the input data.

(A) Measured and reconstructed eddy-current decay curves induced by a Z3-shim step with a rapidly decaying self-term and a cross-term to B0. (B) Eddy current maps at two time-points showing transversal, sagittal and coronal slices of the measured EC fields, the fitted B0- and Z3-components, their superposition and the resultant residual after subtracting the fitted EC maps from the measured data.

Proc. Intl. Soc. Mag. Reson. Med. 27 (2019)
0444