Burak Akin1 and Ali Caglar Özen1,2
1Department of Radiology, Medical Physics, Medical Center ‐ University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Cancer Consortium Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
Rx coils for high
field strengths are mostly based on loop coil design and cover relatively large
FOVs, limiting the achievable voxel size due to the SNR and acquisition time
constraints. Microstrip coils have limited sensitivity profile along the B1
direction, thus are useful for resolving cortical activity. In this study, we
introduce flexible microstrip array insert coils (MSAi) that are compatible
with existing volume coils. We demonstrate measurement of BOLD and highly
confined activity wide spread in calcarine sulcus and motor cortex using MSAi
at 3 T
Introduction
There is an increasing
interest on laminar applications at ultra-high field strengths. Visual cortex and
sensory motor systems of the human brain have been predominantly investigated [1,2]. Resolution depends not only on field strength
but also image and temporal snr[3]. Use of dense Rx arrays 2Rx coils designed for imaging a specific region
of brain [4], reduced FOV imaging methods [5]together with the SNR advantage of ultra-high
fields enable sufficient SNR to resolve laminar activation. However, potential
of high field systems (i.e. up to 3 T) have not been studied comprehensively. Existing
Rx coils for high field strengths are mostly based on loop coil design and
cover relatively large FOVs, limiting the achievable voxel size due to the SNR
and acquisition time constraints. Microstrip coils have limited sensitivity
profile along B1 direction [6], therefore useful for resolving cortical
activity. In this study, we introduce flexible microstrip array insert coils
(MSAi) that are compatible with existing coils such that their performance will
not impeded by the volume coils they are inserted in. We demonstrate
measurement of BOLD activity and grey matter activity around the drain vessels
in calcarine sulcus and motor cortex using MSAi at 3 T.
Methods
A four-element microstrip array insert (MSAi) of
5x18cm2 plane size was constructed using nylon formers and a continuous
ground plane (Fig.1). Each microstrip block has dimensions of 1.27x4.2x5.3cm3.
To keep the coil profile low and to prevent volunteer discomfort, remote tuning
and matching was done after an l/4-long coaxial line section. Active detuning
was achieved also by a remote inductance, as well as a serial matching
capacitor for preamplifier decoupling. Each element was tuned to 123 MHz
with an unloaded and loaded Q values of 59.2 and 20.2, respectively. An
intrinsic decoupling of 14 dB between the neighboring strip elements resulted
in a stable resonance even if the MSAi is bent 30° such that S11=-22.9+/-9.3
dB, fc=124.8+/-1.1 MHz, and S21=11.7+/-3.1 dB between the neighboring elements
without preamplifier decoupling (Fig.2-Gif/video). A phantom experiment is
conducted to compare image SNR Data
was acquired on Siemens 3T MAGNETOM Prisma scanner with a commercial EPI sequence.
The sequence parameters as follows: TE=36ms,TR=3s, FA=30o, bandwith
890 Hz/px, 0.8mm isotropic resolution with 38 slices. Field of view(FOV) is
reduced to 144x54mm. Phase encoding direction
is always chosen perpendicular to coil surface to avoid folding over, so anterior
to posterio for occipital lobe and head to foot chosen for imaging the motor cortex. Slice
prescriptions are shown in Fig.4. The visual paradigm was a flickering checkerboard
with 8Hz reversal time displayed for 10seconds and 20s of resting period with a
fixation cross. For motor task, subject was asked to perform finger tapping
during displayed visual cues. A simple
general linear model (GLM) is used to analyze functional data. Results & Discussion
2 to 3 fold SNR
enhancement is achieved (Fig.3). GLM results showed confined activation
patterns in central sulcus and calcarine sulcus (see inFigure5 & Figure6).
Although there are many gray matter overlapping areas, some regions are very
close to draining veins. Further studies needed to change sensitivity profile
more on gray matter regions. In this study, We have demonstrated feasibility of
sub-mm image acquisition in 3T by using commercially available sequences and
only reducing FOV and without any sequence acceleration. Advantages of MSAi
are; there is no need to move the patient-, could be insert inside the cushion
and could easily be selected through the console and compatibly works with
existing coils. MSAi can also be designed for ultra-high fields, which will
improve the SNR in cortex even further and BOLD sensitivity will be more
focused on capillaries instead of draining veins. Optimization of sensitivity
profile might further improve necessary SNR. Integration of further MSAi
elements also along the PE direction should be studied for better performance. Acknowledgements
No acknowledgement found.References
1. Huber, L. et al.
High-Resolution CBV-fMRI Allows Mapping of Laminar Activity and Connectivity of
Cortical Input and Output in Human M1. Neuron 96, 1253-1263.e7
(2017).
2.Koopmans,
P. J., Barth, M. & Norris, D. G. Layer-specific BOLD activation in human
V1. Hum. Brain Mapp. 31, 1297–1304 (2010).
3.Triantafyllou,
C. et al. Comparison of physiological noise at 1.5 T, 3 T and 7 T and
optimization of fMRI acquisition parameters. NeuroImage 26,
243–250 (2005).
4. Farivar,
R., Grigorov, F., van der Kouwe, A. J., Wald, L. L. & Keil, B. Dense,
shape-optimized posterior 32-channel coil for submillimeter functional imaging
of visual cortex at 3T. Magn. Reson. Med. 76, 321–328 (2016).
5.Yuan,
J., Zhao, T.-C., Tang, Y. & Panych, L. P. Reduced field-of-view single-shot
fast spin echo imaging using two-dimensional spatially selective radiofrequency
pulses. J. Magn. Reson. Imaging JMRI 32, 242–248 (2010).
6.Kumar,
A. & Bottomley, P. A. Optimizing the Intrinsic Signal-to-Noise Ratio of MRI
Strip Detectors. 166, 157–166 (2006).