Lieke van den Wildenberg1, Quincy van Houtum1, Wybe J. M. van der Kemp1, Catalina S. Arteaga de Castro1, Sahar Nassirpour2, Paul Chang2, and Dennis W. J. Klomp1
1Center for Image Sciences, UMC Utrecht, Utrecht, Netherlands, 2MR Shim GmbH, Reutlingen, Germany
Synopsis
Inhomogeneity of the magnetic field (B0)
in the human body, for instance caused by physiological motion (e.g.
breathing), can lead to artifacts. Static second order magnetic field shimming
even in the absence of breathing cannot provide a uniform magnetic field in
large organs such as the liver, particularly at ultra-high field throughout the
breathing cycle. We show that in conjunction with static second order shimming,
local arrays of shim coils can substantially improve the magnetic field
uniformity in the liver at different breathing states.
Introduction
Variations of the static magnetic field (B0)
in the human body can lead to serious frequency offsets during MR and magnetic
resonance spectroscopy. If not accounted for, this can be reflected in the form
of artifacts, peak shifts, line broadening, and signal loss.1 These
variations are partially caused by physiological motion like breathing, and
other temporal B0 field fluctuations in the MR system. Higher
magnetic field strengths, such as 7T, are more sensitive to the effects of large
dynamic B0 field variations. Applying image based B0 shimming,
while monitoring breathing and/or respiratory triggering, helps in further
reducing the B0 field variations, however, usually at the cost of an
extended scan time. In this study, we
show that variations in B0 can be reduced at different respiratory states
where the static field is kept homogeneous (i.e. B0 shimmed). In
conjunction with static second order shimming, with the addition of a local
array of shim coils that can be steered externally to adjust the residual
magnetic field inhomogeneities, the field homogeneity can be improved
throughout the liver.2Methods
4 healthy subjects were scanned at a 7T MR
scanner (Philips, Cleveland, USA) after giving informed consent. Eight
transceiver fractionated dipole antennas with 16 additional receive loops (MR
Coils BV, Zaltbommel, The Netherlands) interfaced to 8 parallel 2kW channels were positioned symmetrically
around the abdomen.3 To maximize and homogenize the B1+
field in the liver region, RF phase shimming was performed. Four
dual echo B0 maps (GE, 240×304×18mm3
FOV, 2×2×2mm3 voxel size, FA=5$$$^\circ$$$, TR=10ms) during free breathing and three during breath-hold in
expiration and inhalation states (GE, 386×410×180mm3 FOV, 6×6×6mm3
voxel size, FA=4$$$^\circ$$$, TR=6ms)
were acquired to assess the potential effect of the shim coil array on B0
shimming on the liver (MR Shim GmbH, Reutlingen, Germany). The local shim coil array
used for this simulation consisted of 16 circular loops of each 5cm in
diameter. The diameters were optimized using a brute-force algorithm between
the ranges of 1cm to 10cm. The objective function was defined as the standard
deviation of the residual frequency shifts on a sample in vivo dataset. The
coils were arranged in two rows of 8 around the torso, (8 on top and 8 at the
bottom) (Figure 1). The coil centers of the two rows were 11cm apart. Firstly,
shimmed B0 fields were simulated using only static second order
shimming (using the free-breathing field maps as reference maps). Secondly, in
addition to the static second order shims, the local shim array was used to
shim each of the different states of the breathing cycle (using the reference
maps acquired from each state separately). Another two B0 simulations were performed: 1) second order
for each breathing state, and 2) second order + local shim array for each
breathing state. Slice-based analysis was
performed to calculate B0 distributions and standard deviations in
the three orthogonal directions.Results
Figure 2 lists that inhaled state B0
maps shows the most improvement when a local shim array is used, because of the
large inhomogeneity in the unshimmed B0 maps. Notably, exhaled state
and free breathing B0 shims with the local shim coil array show an
almost similar improvement of approximately 20%, as compared to the scanner
shim. However, when driving both the second order and
local shim arrays with per breathing stage optimized shim currents, the exhaled
state gives most improvement. Furthermore, the improvement using the lower
resolution B0 maps as reference maps is similar to when using higher
resolution maps. Figure 3 shows clearly that the local array of shim coils outperforms the
scanner B0 shim. Figure 4 shows an
example of a slice in the middle of the liver of the B0 field of
subject 3 for the four shimming comparisons.Discussion
As
expected, there is a large inhomogeneity found in the fully inhaled state B0
maps, due to the amount of air in the lungs. However, the local array of shim
coils in conjunction with the second order shimming of the MRI system, can
improve the B0 field homogeneity compared to solely using the
scanner shim. The inhaled and exhaled state of the breath-hold are the worst
case scenarios.Conclusion
This study shows that B0 variations
in the liver at 7T can be reduced by an additional 20% after static B0
shimming, even in the presence of respiratory motion by using a local array of
statically driven shim coils. For optimal B0 homogeneity, shim
updating during different phases of respiratory motion is necessary.Acknowledgements
European
H2020-FETOPEN: NICIReferences
1. Truong Trong-Kha, et al. Effects
of static and radiofrequency magnetic field inhomogeneity in ultra-high field
magnetic resonance imaging. Magnetic Resonance Imaging. 2006;24:103-112.
2. Juchem, Christoph, et al.
Dynamic shimming of the human brain at 7 T. Concepts in Magnetic Resonance Part
B: Magnetic Resonance Engineering 2010;37(3):
116-128.
3. Raaijmakers A.J.E., et al. The fractionated
dipole antenna: A new antenna for body imaging at 7 Tesla. Magnetic Resonance
in Medicine. 2016;75:1366–1374.