To perform Chemical Exchange Saturation Transfer MRI of the whole brain a homogeneous saturation and fast readout are required. To achieve a fast and robust 3D acquisition a spiral-centric-reordered GRE readout was used. In addition, a Multiple Interleaved Mode Saturation scheme was applied to mitigate B1+-inhomogeneity effects of the CEST saturation. Combination of these two methods allows acquiring a homogenous CEST contrast in a volume of approximately 220x220x45mm3 within an acquisition time of 7 min 26s.
Pulse sequences were implemented on a 7-Tesla whole-body MR system (MAGNETOM Terra, Siemens Healthineers, Erlangen, Germany). An 8Tx/32Rx head coil (Nova Medical, Wilmington, USA) was used for the measurements. Phantom measurements were performed with an 18 cm diameter sphere filled with egg-white. In vivo measurements were performed on a healthy female volunteer. The images were acquired using a pulsed CEST spiral-centric-reordered -GRE sequence[1, 8].
For the MIMOSA acquisition, two sets of complimentary transmitter phases with constant amplitudes (‘modes’) were used for the saturation pulse train as described previously[7]. The mode A had a 45° phase difference between two adjacent transmitter coil channels (“Circular Polarization - CP”). In mode B a 90° phase increment between the channels was chosen (also called B2+)[9]. Each mode has a distinct effective relative B1+ value distribution. Through interleaving of the two modes during saturation, (c.f. Figure 1) a more homogenous saturation can be achieved. During the GRE multi-slice acquisition the excitation pulses are employed in the CP mode. CEST images were also acquired using the CP mode during both acquisition and saturation. MTRRex(NOE) contrast was calculated for both standard CEST acquisition and MIMOSA acquisition based on the work by Windschuh et al.[4]. B1+ maps of both CP and B2+ mode were acquired using a multi flip angle (FA) sequence. A two-point B1 correction was applied on the CP mode acquisition as presented by Windschuh et al.[4]. The B1+ distribution of MIMOSA was calculated based on the B1 continuous wave power equivalent (CWPE) as described by Zu et al[10]:
$$ B_{1,CWPE} (r,t)=\sqrt{\frac{1}{(\tau_p+\tau_d)}\int_0^{\tau_p} \! B^2_P(r,t) \, \mathrm{d}t} $$
Sequence parameters: CEST Saturation parameters: τp=46.08ms, τD=30 ms, n=50, Trec=1s, B1,Nominal=0.6µT and 1.0µT in CP mode, Gaussian pulse; GRE acquisition parameters: FA=6°, TR=4.5ms, TE=2.3ms, Matrix Size=128x128x16, FoV=220x220x80 mm3, GRAPPA 3, Elongation Factor=0.6,
Results and Discussion
In comparison to the standard saturation using CP mode, MIMOSA reduces the intra slice variation of the B1+-inhomogeneity in all of the acquired slices (c.f. phantom measurement, Figure 2). An influence of the higher rB1 value of the CP mode can be observed in the central slices of the phantoms B1,CWPE maps. Figure 3 shows that homogeneity of the MTRRex(NOE) contrast is comparable to B1 corrected images in most of the slices with MIMOSA saturation. The fitting routine fails in the outermost slices which might be caused by imperfect slab selection as well as the lower B1+ or B0 homogeneity.
In vivo results show similarities to the phantom measurements. MIMOSA reduces the intra slice variation of the B1+-inhomogeneity in all of the acquired slices (c.f. Figure 4). In cranial slices, similarly to the outermost slices of the phantom, a decrease of the rB1 value in the MIMOSA B1,CWPE maps can be observed.
In Figure 5, MTRRex(NOE) values of a healthy human brain derived both without and with the use of MIMOSA are presented. In CP mode without B1 correction, a correlation between the loss of MTRRex(NOE) magnitude and the B1 map can be observed (c.f. Figure 4A). The application of MIMOSA resulted in a more homogeneous image in central slices of the brain (c.f. Figure 5C), which is comparable to B1 corrected images (c.f. Figure 5B). Cranial slices acquired with MIMOSA show reduced MTRRex(NOE) values and slightly reduced homogeneity in comparison to images acquired in CP mode both with and without B1 correction. This is caused by low CWPE B1+ values in these slices (c.f. Figure 4C).
1. Zaiss, M., P. Ehses, and K. Scheffler, Snapshot-CEST: Optimizing spiral-centric-reordered gradient echo acquisition for fast and robust 3D CEST MRI at 9.4 T. NMR Biomed, 2018. 31(4): p. e3879.
2. Akbey, S., et al., Single-shot whole-brain CEST imaging using centric-reordered 3D-EPI. Proc. Intl. Soc. Mag. Reson. Med. , 2018. 26.
3. Khlebnikov, V., et al., On the transmit field inhomogeneity correction of relaxation-compensated amide and NOE CEST effects at 7 T. NMR Biomed, 2017. 30(5).
4. Windschuh, J., et al., Correction of B1-inhomogeneities for relaxation-compensated CEST imaging at 7 T. NMR Biomed, 2015. 28(5): p. 529-37.
5. Schuenke, P., et al., Simultaneous mapping of water shift and B1 (WASABI)-Application to field-Inhomogeneity correction of CEST MRI data. Magn Reson Med, 2016.
6. Tse, D.H., et al., B1+ inhomogeneity mitigation in CEST using parallel transmission. Magn Reson Med, 2017.
7. Liebert, A., et al., Multiple Interleaved Mode Saturation (MIMOSA) for B1+ inhomogeneity mitigation in chemical exchange saturation transfer. Proc. Intl. Soc. Mag. Reson. Med. , 2018. 26.
8. Schmitt, B., et al., Optimization of pulse train presaturation for CEST imaging in clinical scanners. Magn Reson Med, 2011. 65(6): p. 1620-9.
9. Orzada, S., et al., Mitigation of B1(+) inhomogeneity on single-channel transmit systems with TIAMO. Magn Reson Med, 2013. 70(1): p. 290-4.
10. Zu, Z., et al., Optimizing pulsed-chemical exchange saturation transfer imaging sequences. Magn Reson Med, 2011. 66(4): p. 1100-8.