In this work, a recently developed method called PRO-QUEST (PROgressive saturation for Quantifying Exchange using Saturation Times) is translated to a 3T clinical scanner for assessing pH-sensitive indices in phantoms and a healthy volunteer. Our results demonstrate that quantification of pH sensitive indices using PRO-QUEST is feasible at 3T within clinically acceptable acquisition times. Our initial findings suggest that PRO-QUEST has the potential to provide a new biomarker to study neurological disorders associated with brain tissue acidosis.
The PRO-QUEST sequence was implemented on a 3T Philips Ingenia MRI scanner (Philips Healthcare, Best, the Netherlands) and tested on phantoms consisting of 100mM glutamate in a standard solution of 1x phosphate-buffered saline (PBS) with several pH (6.08, 6.64 and 7.19) and a pure PBS sample (pH 7.14). Phantoms and a healthy volunteer were scanned using a 32 channel head coil. First, a Look-Locker (LL) sequence (Figure 1a) was implemented with 20ms delay times (in lieu of off-resonance saturation pulses displayed in Figure 1b) prior to a multishot turbo field echo planar imaging (TFEPI) readout (EPI factor=7) and n acquisitions (n=128 for phantom; n=143 for volunteer) with the following imaging parameters: imaging pulse=sinc-gaussian, duration=0.67 ms, flip angle=8°\15°, TE=3.8 ms, time between readout pulses=42 ms, acquired resolution=1.88x2.14x5 mm3 (phantom) and 1.96x2.04x5mm3 (volunteer), TR=6s. For the PRO-QUEST scans (Figure 1b), an off-resonance saturation pulse centred at 3.0ppm (glutamate phantom) or 3.5ppm (volunteer) was applied prior to the TFEPI readouts with identical imaging parameters as the LL sequence. Parameters for the off-resonance saturation pulses used in the PRO-QUEST sequence are as follows: off-resonance saturation pulse=sinc-gaussian, bandwidth=300Hz, duration=20 ms, flip angle=400° (equivalent of 1.3μT). For the healthy volunteer scan, single slice acquisitions were obtained with a scan time of 2 min 6 s (3 averages) per sequence. Imaging parameters are summarised in Table 1. Additionally, standard multi-echo turbo spin echo (TSE) sequence (TSE factor=20) consisting of 10 echoes with TE=20-200ms with 20ms of inter-echo spacing was used to quantify T2 (to be used as a input parameter in equation 2) in the same geometry as PRO-QUEST.
Data processing was performed using custom-written scripts in MATLAB (The Mathworks, Natick, MA, USA). The derived Block-McConnell models4 were fitted to magnitude data using maximum likelihood estimation. The following equation was fitted to LL data to estimate the equilibrium magnetization M0, T1 and B1:
Mzd(nτ)={1-[(cosθ)n-1e-(n-1)τR1]}Mzd(τ)/{1-[(cosθ)e-τR1]}+M0(1-e-tdR1)[(cosθ)n-1e-(n-1)τR1 -----[1]
where Mzd(τ) = M0 (1-e-τR1); td is the time between the initial saturation pulse and the first readout pulse; τ is the time between readout pulses with small flip angle θ; R1 = 1/T1; n is number of acquisitions.
Next, the obtained M0,T1, B1 values were used as input parameters for estimating the exchange-dependent relaxation, Rex by fitting the PRO-QUEST data:
Mzsat(nτ)={1-[(cosθ)ne-n(τR1-tsat(R1-R1ρ) )]}Mzsat(τ)/{1-[(cosθ)e-(τR1-tsat(R1-R1ρ))]}+M0(1-e-tdR1)[(cosθ)ne-n(-τR1-tsat(R1-R1ρ))] -----[2]
where Mzsat(τ) = Mss (1-e-(R1ρtsat))(cosθ)e-(τ - tsat)R1 + M0(1-e-(τ-tsat)R1; Mss = (R1cos2φ)/R1ρ; R1ρ = R1cos2φ+(R2+ Rex)sin2φ; φ is the angle between the effective field and the z-axis. Further definition of the equations and parameters are described in literature4.
Similar to the pre-clinical cases4, progressive saturation recovery curves with off-resonance saturation pulses show clear separation among samples with various pH values in glutamate and PBS (Figure 2b) while the ones without off-resonance saturation pulses are nearly indistinguishable (Figure 2a). The estimated Rex significantly correlates with pH in glutamate samples (Figure 2c). In the healthy volunteer, the PRO-QUEST image of signal evolution at the final phase of the amide proton resonance shows clear contrast between white and grey matters (WM/GM) (figure 3b) as contrary to the LL image (without off-resonance saturation pulses) (figure 3a). The origin of contrast between WM and GM needs further investigation. As for prerequisite parameters in estimation of PRO-QUEST indices, calculated T1 values from the LL scan in a healthy volunteer are remarkably consistent with literature values (table 2)5-7. The pH sensitive Rex shows differences between WM and GM.
Due to intrinsic limitations of the specific absorption rate and duty cycle (50%) at clinical field strength, the efficiency of the off-resonance saturation scheme is somewhat compromised. Nonetheless, clinical translation of this technique is very feasible given its easy implementation on standard clinical platforms and the use of existing LL-type of readouts, therefore not requiring pulse programming. Further work is required to achieve full brain coverage within clinically relevant acquisition time.